Directory Details

Dubai Electricity and Water Authority PJSC

Initial Accreditation Date:Aug 10 , 2017

Contacts Info

Address

Dubai, United Arab Emirates

PO box / Zip Code

564

Email

mohammad.alshamsi@dewa.gov.ae

Website

dewa.gov.ae

Main Field

Calibration Laboratories

Valid Since:Oct 26 , 2023

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Electrical
  • DC Voltage
  • Multi Product Calibrator (Fluke 5522A) Calibration Procedure # DP-DAM-DAM-W&AR-Manual-001.A12C U: Measured Voltage value
  • Up to 329.9999 mV
16×10ˉ⁶ U + 0.79 µVDEWA Metrology LaboratoryActive
  • < 330 mV to 3.3 V
8.4×10ˉ⁶ U + 1.9 µVDEWA Metrology LaboratoryActive
  • < 3.3 V to 33 V
9.2×10ˉ⁶ U + 19 µVDEWA Metrology LaboratoryActive
  • < 33 V to 330 V
14×10ˉ⁶ U + 0.13 mVDEWA Metrology LaboratoryActive
  • < 330 V to 1000 V
14×10ˉ⁶ U + 1.4 mVDEWA Metrology LaboratoryActive
  • AC Voltage
  • (Up to 33 mV) 45 Hz to 10 kHz
0.11×10ˉ³ U + 4.8 µVDEWA Metrology LaboratoryActive
  • (Up to 33 mV) < 10 kHz to 20 kHz
0.15x10ˉ³ U + 4.8 µVDEWA Metrology LaboratoryActive
  • (Up to 33 mV) < 20 kHz to 50 kHz
0.77x10ˉ³ U + 4.8 µVDEWA Metrology LaboratoryActive
  • (< 33 mV to 330 mV) 45 Hz to 10 kHz
0.11x10ˉ³ U + 6.3 µVDEWA Metrology LaboratoryActive
  • (< 33 mV to 330 mV) < 10 kHz to 20 kHz
0.12x10ˉ³ U + 6.3 µVDEWA Metrology LaboratoryActive
  • (< 33 mV to 330 mV) < 20 kHz to 50 kHz
0.27x10ˉ³ U + 6.3 µVDEWA Metrology LaboratoryActive
  • (<0.33 V to 3.3 V) 45 Hz to 10 kHz
0.12x10ˉ³ U + 47 µVDEWA Metrology LaboratoryActive
  • (<0.33 V to 3.3 V) < 10 kHz to 20 kHz
0.15x10ˉ³ U + 47 µVDEWA Metrology LaboratoryActive
  • (<0.33 V to 3.3 V) < 20 kHz to 50 kHz
0.23x10ˉ³ U + 39 µVDEWA Metrology LaboratoryActive
  • (< 3.3 V to 33 V) 45 Hz to 10 kHz
0.12x10ˉ³ U + 0.47 mVDEWA Metrology LaboratoryActive
  • (< 3.3 V to 33 V) < 10 kHz to 20 kHz
0.19x10ˉ³ U + 0.47 mVDEWA Metrology LaboratoryActive
  • (< 3.3 V to 33 V) < 20 kHz to 50 kHz
0.27x10ˉ³ U + 0.47 mVDEWA Metrology LaboratoryActive
  • (< 33 V to 330 V) 45 Hz to 10 kHz
0.16x10ˉ³ U + 4.7 mVDEWA Metrology LaboratoryActive
  • (< 33 V to 330 V) < 10 kHz to 20 kHz
0.19x10ˉ³ U + 4.7 mVDEWA Metrology LaboratoryActive
  • (< 33 V to 330 V) < 20 kHz to 50 kHz
0.23x10ˉ³ U + 4.7 mVDEWA Metrology LaboratoryActive
  • (< 330 V to 1020 V) 45 Hz to 1 kHz
0.23x10ˉ³ U + 8.8 mVDEWA Metrology LaboratoryActive
  • (< 330 V to 1020 V) < 1 kHz to 5 kHz
0.19x10ˉ³ U + 9.0 mVDEWA Metrology LaboratoryActive
  • (< 330 V to 1020 V) < 5 kHz to 10 kHz
0.23x10ˉ³ U + 8.8 mVDEWA Metrology LaboratoryActive
  • DC Current
  • Multi Product Calibrator (Fluke 5522A ) Calibration Procedure # DP-DAM-DAM-W&AR-Manual-001.A12C I: Measured Current value
  • Up to 330 µA
0.12×10ˉ³ I + 0.02 µADEWA Metrology LaboratoryActive
  • < 0.33 mA to 3.3 mA
1.2×10ˉ⁶ I + 12 µADEWA Metrology LaboratoryActive
  • < 3.3 mA to 33 mA
1.1×10ˉ⁶ I + 0.12 mADEWA Metrology LaboratoryActive
  • < 33 mA to 330 mA
1.1×10ˉ⁶ I + 1.2 mADEWA Metrology LaboratoryActive
  • < 330 mA to 1.1 A
1.9×10ˉ⁶ I + 12 mADEWA Metrology LaboratoryActive
  • < 1.1 A to 3 A
16×10ˉ⁶ I + 12 mADEWA Metrology LaboratoryActive
  • < 3 A to 11 A
10×10ˉ⁶ I + 0.12 ADEWA Metrology LaboratoryActive
  • < 11 A to 20 A
85×10ˉ⁶ I + 0.12 ADEWA Metrology LaboratoryActive
  • AC Current
  • Multi Product Calibrator (Fluke 5522A) Calibration Procedure # DP-DAM-DAM-W&AR-Manual-001.A12C I: Measured Current value
  • (29 µA to 330 µA) 45 Hz to 1 kHz
0.97×10ˉ³ I + 0.08 µADEWA Metrology LaboratoryActive
  • (29 µA to 330 µA) < 1 kHz to 5 kHz
2.3×10ˉ³ I + 0.12 µADEWA Metrology LaboratoryActive
  • (29 µA to 330 µA) < 5 kHz to 10 kHz
6.2×10ˉ³ I + 0.16 µADEWA Metrology LaboratoryActive
  • (< 0.33 mA to 3.3 mA) 45 Hz to 1 kHz
0.78×10ˉ³ I + 0.12 µADEWA Metrology LaboratoryActive
  • (< 0.33 mA to 3.3 mA) < 1 kHz to 5 kHz
1.6×10ˉ³ I + 0.16 µADEWA Metrology LaboratoryActive
  • (< 0.33 mA to 3.3 mA) < 5 kHz to 10 kHz
3.9×10ˉ³ I + 0.23 µADEWA Metrology LaboratoryActive
  • (< 3.3 mA to 33 mA) 45 Hz to 1 kHz
0.31×10ˉ³ I + 1.6 µADEWA Metrology LaboratoryActive
  • (< 3.3 mA to 33 mA) < 1 kHz to 5 kHz
0.62×10ˉ³ I + 1.6 µADEWA Metrology LaboratoryActive
  • (< 3.3 mA to 33 mA) < 5 kHz to 10 kHz
1.6×10ˉ³ I + 2.3 µADEWA Metrology LaboratoryActive
  • (< 33 mA to 330 mA) 45 Hz to 1 kHz
0.31×10ˉ³ I + 16 µADEWA Metrology LaboratoryActive
  • (< 33 mA to 330 mA) < 1 kHz to 5 kHz
0.78×10ˉ³ I + 39 µADEWA Metrology LaboratoryActive
  • (< 33 mA to 330 mA) < 5 kHz to 10 kHz
1.6×10ˉ³ I + 78 µADEWA Metrology LaboratoryActive
  • (< 0.33 A to 1.1 A) 45 Hz to 1kHz
0.39×10ˉ³ I + 78 µADEWA Metrology LaboratoryActive
  • (< 0.33 A to 1.1 A) < 1 kHz to 5 kHz
4.7×10ˉ³ I + 0.78 mADEWA Metrology LaboratoryActive
  • (< 0.33 A to 1.1 A) < 5 kHz to 10 kHz
19×10ˉ³ I + 3.9 mADEWA Metrology LaboratoryActive
  • (< 1.1 A to 3 A) 45 Hz to 1 kHz
0.47×10ˉ³ I + 78 µADEWA Metrology LaboratoryActive
  • (< 1.1 A to 3 A) < 1 kHz to 5 kHz
4.7×10ˉ³ I + 0.78 mADEWA Metrology LaboratoryActive
  • (< 1.1 A to 3 A) < 5 kHz to 10 kHz
19×10ˉ³ I + 3.9 mADEWA Metrology LaboratoryActive
  • (< 3 A to 11 A) 45 Hz to 100 Hz
0.47×10ˉ³ I + 1.6 mADEWA Metrology LaboratoryActive
  • (< 3 A to 11 A) < 100 Hz to 1 kHz
0.78×10ˉ³ I + 1.6 mADEWA Metrology LaboratoryActive
  • (< 3 A to 11 A) < 1 kHz to 5 kHz
23×10ˉ³ I + 1.6 mADEWA Metrology LaboratoryActive
  • (< 11 A to 20.5 A) 45 Hz to 100 Hz
0.93×10ˉ³ I + 3.9 mADEWA Metrology LaboratoryActive
  • (< 11 A to 20.5 A) < 100 Hz to 1 kHz
1.2×10ˉ³ I + 3.9 mADEWA Metrology LaboratoryActive
  • (< 11 A to 20.5 A) < 1 kHz to 5 kHz
23×10ˉ³ I + 3.9 mADEWA Metrology LaboratoryActive
  • Temperature Simulation
  • 1- Multi Product Calibrator (Fluke 5522A, TC-K type) Calibration Procedure # DP-DAM-DAM-W&AR-Manual-001.A12C
  • ­200 °C to ­100 °C
0.26 °CDEWA Metrology LaboratoryActive
  • < -100 °C to -25 °C
0.14 °CDEWA Metrology LaboratoryActive
  • < -25 °C to 120 °C
0.12 °CDEWA Metrology LaboratoryActive
  • < 120 °C to 1000 °C
0.20 °CDEWA Metrology LaboratoryActive
  • < 1000 °C to 1372 °C
0.31 °CDEWA Metrology LaboratoryActive
  • 2- Multi Product Calibrator (Fluke 5522A, TC-J type) Calibration Procedure # DP-DAM-DAM-W&AR-Manual-001.A12C
  • ­200 °C to ­100 °C
0.21 °CDEWA Metrology LaboratoryActive
  • < ­100 °C to ­25 °C
0.12 °CDEWA Metrology LaboratoryActive
  • < ­25 °C to 120 °C
0.21 °CDEWA Metrology LaboratoryActive
  • < 120 °C to 1000 °C
0.13 °CDEWA Metrology LaboratoryActive
  • < 1000 °C to 1372 °C
0.18 °CDEWA Metrology LaboratoryActive
  • Capacitance
  • Multi Product Calibrator (Fluke-5520A) Calibration Procedure # DP-DAM-DAM-W&AR-Manual-001.A12C C: Measured Capacitance value
  • 1.1 nF to 3.3 nF
3.9×10ˉ³ C + 7.8 pFDEWA Metrology LaboratoryActive
  • < 3.3 nF to 11 nF
1.9×10ˉ³ C + 78 pFDEWA Metrology LaboratoryActive
  • < 11 nF to 33 nF
1.9×10ˉ³ C + 78 pFDEWA Metrology LaboratoryActive
  • < 33 nF to 110 nF
1.9×10ˉ³ C + 78 pFDEWA Metrology LaboratoryActive
  • < 110 nF to 330 nF
1.9×10ˉ³ C + 0.23 nFDEWA Metrology LaboratoryActive
  • < 0.33 µF to 1.1 µF
1.9×10ˉ³ C + 0.78 nFDEWA Metrology LaboratoryActive
  • < 1.1 µF to 3.3 µF
1.9×10ˉ³ C + 2.3 nFDEWA Metrology LaboratoryActive
  • < 3.3 µF to 11 µF
1.9×10ˉ³ C + 7.8 nFDEWA Metrology LaboratoryActive
  • < 11 µF to 110 µF
3.5×10ˉ³ C + 78 nFDEWA Metrology LaboratoryActive
  • < 110 µF to 330 µF
3.5×10ˉ³ C + 0.23 µFDEWA Metrology LaboratoryActive
  • < 0.33 mF to 1.1 mF
3.5×10ˉ³ C + 0.78 µFDEWA Metrology LaboratoryActive
  • < 1.1 mF to 3.3 mF
3.5×10ˉ³ C + 2.3 µFDEWA Metrology LaboratoryActive
  • < 3.3 mF to 11 mF
3.5×10ˉ³ C +7.8 µFDEWA Metrology LaboratoryActive
  • < 11 mF to 33 mF
5.8×10ˉ³ C + 23 µFDEWA Metrology LaboratoryActive
  • < 33 mF to 110 mF
8.5×10ˉ³ C + 78 µFDEWA Metrology LaboratoryActive
  • Resistance
  • Multi Product Calibrator (Fluke-5520A) Calibration Procedure # DP-DAM-DAM-W&AR-Manual-001.A12C R: Measured Resistance value
  • 0 to 11 Ω
31×10ˉ⁶ R + 7.8 mΩDEWA Metrology LaboratoryActive
  • < 11 Ω to 33 Ω
23×10ˉ⁶ R + 12 mΩDEWA Metrology LaboratoryActive
  • < 33 Ω to 110 Ω
22×10ˉ⁶ R + 12 mΩDEWA Metrology LaboratoryActive
  • < 110 Ω to 330 Ω
22×10ˉ⁶ R + 16 mΩDEWA Metrology LaboratoryActive
  • < 330 Ω to 1.1 kΩ
22×10ˉ⁶ R + 16 mΩDEWA Metrology LaboratoryActive
  • < 1.1 kΩ to 3.3 kΩ
22×10ˉ⁶ R + 0.16 ΩDEWA Metrology LaboratoryActive
  • < 3.3 kΩ to 11 kΩ
22×10ˉ⁶ R + 78 mΩDEWA Metrology LaboratoryActive
  • < 11 kΩ to 33 kΩ
22×10ˉ⁶ R + 0.78 ΩDEWA Metrology LaboratoryActive
  • < 33 kΩ to 110 kΩ
22×10ˉ⁶ R + 0.78 ΩDEWA Metrology LaboratoryActive
  • < 110 kΩ to 330 kΩ
25×10ˉ⁶ R + 7.8 ΩDEWA Metrology LaboratoryActive
  • < 330 kΩ to 1.1 MΩ
25×10ˉ⁶ R + 7.8 ΩDEWA Metrology LaboratoryActive
  • < 1.1 MΩ to 3.3 MΩ
47×10ˉ⁶ R + 0.12 kΩDEWA Metrology LaboratoryActive
  • < 3.3 MΩ to 11 MΩ
0.10×10ˉ³ R + 0.19 kΩDEWA Metrology LaboratoryActive
  • < 11 MΩ to 33 MΩ
0.19×10ˉ³ R + 1.9 kΩDEWA Metrology LaboratoryActive
  • < 33 MΩ to 110 MΩ
0.39×10ˉ³ R + 2.3 kΩDEWA Metrology LaboratoryActive
  • < 110 MΩ to 330 MΩ
2.3×10ˉ³ R + 78 kΩDEWA Metrology LaboratoryActive
  • < 330 MΩ to 1100 MΩ
9.5×10ˉ³ R + 63 kΩDEWA Metrology LaboratoryActive
  • Frequency
  • Multi Product Calibrator (Fluke-5520A) Calibration Procedure # DP-DAM-DAM-W&AR-Manual-001.A12C f: Measured Frequency value
  • 0.01 Hz to 120 Hz
0.02×10ˉ⁶ f + 12 mHzDEWA Metrology LaboratoryActive
  • < 120.0 Hz to 1.2 kHz
0.02×10ˉ⁶ f + 0.12 HzDEWA Metrology LaboratoryActive
  • < 1.2 kHz to 12 kHz
0.02×10ˉ⁶ f + 1.2 HzDEWA Metrology LaboratoryActive
  • < 12 kHz to 120 kHz
0.02×10ˉ⁶ f + 12 HzDEWA Metrology LaboratoryActive
  • < 120 kHz to 1200 kHz
0.02×10ˉ⁶ f + 0.12 kHzDEWA Metrology LaboratoryActive
  • < 1.2 MHz to 2 MHz
5.2×10ˉ⁹ f + 1.2 kHzDEWA Metrology LaboratoryActive
  • DC Current
  • Multi Product Calibrator Fluke-5522A & Current Coil Fluke-5500A /Coil Calibration Procedure # DP-DAM-DAM-W&AR-Manual-001.A12D I: Measured Current value
  • 10 A to 16.5 A
5.0×10ˉ³ I + 20 mADEWA Metrology LaboratoryActive
  • < 16.5 A to 55 A
5.0×10ˉ³ I + 0.14 ADEWA Metrology LaboratoryActive
  • < 55 A to 150 A
5.0×10ˉ³ I + 0.14 ADEWA Metrology LaboratoryActive
  • < 150 A to 550 A
5.0×10ˉ³ I + 0.50 ADEWA Metrology LaboratoryActive
  • < 550 A to 1000 A
5.0×10ˉ³ I + 0.50 ADEWA Metrology LaboratoryActive
  • AC Current
  • Multi Product Calibrator Fluke-5522A & Current Coil Fluke-5500A/Coil Calibration Procedure # DP-DAM-DAM-W&AR-Manual-001.A12D I = Measured Current value
  • 10 A to 16.5 A
5.6×10ˉ³ I + 30 mADEWA Metrology LaboratoryActive
  • 45 Hz to 65 Hz
5.6×10ˉ³ I + 30 mADEWA Metrology LaboratoryActive
  • < 16.5 A to 150 A
5.6×10ˉ³ I + 0.25 ADEWA Metrology LaboratoryActive
  • 45 Hz to 65 Hz
5.6×10ˉ³ I + 0.25 ADEWA Metrology LaboratoryActive
  • < 150 A to 1000 A
5.6×10ˉ³ I + 0.9 ADEWA Metrology LaboratoryActive
  • 45 Hz to 65 Hz
5.6×10ˉ³ I + 0.9 ADEWA Metrology LaboratoryActive
  • Insulation Resistance (50V to 5000V)
  • Insulation Calibration Box- CB 101. DP-DAM-DAM-W&AR- Manual-001.A12E
  • 10 MΩ
0.12 MΩDEWA Metrology LaboratoryActive
  • 100 MΩ
1.2 MΩDEWA Metrology LaboratoryActive
  • 1 GΩ
12 MΩDEWA Metrology LaboratoryActive
  • 10 GΩ
0.12 GΩDEWA Metrology LaboratoryActive
  • Insulation Tester/ Output Voltage Measurement
  • Output Voltage Measurement Using DMM. DP-DAM-DAM-W&AR- Manual-001.A12E
  • 50 V
23 mVDEWA Metrology LaboratoryActive
  • 100 V
38 mVDEWA Metrology LaboratoryActive
  • 250 V
88 mVDEWA Metrology LaboratoryActive
  • 500 V
0.22 VDEWA Metrology LaboratoryActive
  • 1000 V
0.38 VDEWA Metrology LaboratoryActive
  • 2500 V
58 VDEWA Metrology LaboratoryActive
  • 5000 V
0.12 kVDEWA Metrology LaboratoryActive