Master Systems Standardization Laboratory

Initial Accreditation Date:Feb 25 , 2019

Contacts Info

Address

Dubai, United Arab Emirates

PO box / Zip Code

127421

Email

dubai@mastersystems-intl.com

Website

www.mastersystems.org

Main Field

Calibration Laboratories

Valid Since:Jan 17 , 2023

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Temperature
  • Temperature Indicator with Sensor RTD
  • MS-DMC-WP-043 as per DKD-R 5-1 guideline
  • ­30 °C to 150 °C
0.35 °CLaboratoryActive
  • > 150 °C to 600 °C
0.88 °CLaboratoryActive
Electrical
  • Temperature Indicator with Sensor RTD
  • Direct Method as per EURAMET Cg-15:2015 by using Multifunction Calibrator, Model No.: 5025E. U = measured voltage
  • 0 mV to 200 mV
47 X 10-6 U + 5.1 µVLaboratoryActive
  • >0.2 V to 2.0 V
46 X 10-6 U + 30 µVLaboratoryActive
  • >2 V to 20 V
46 X 10-6 U + 0.29 mVLaboratoryActive
  • >20 V to 200 V
47 X 10-6 U + 3.5 mVLaboratoryActive
  • >200 V to 1000 V
71 X 10-6 U + 35 mVLaboratoryActive
  • DC Current
  • Direct Method as per EURAMET Cg-15:2015 by using Multifunction Calibrator, Model No.: 5025E. I = measured current
  • 0 µA to 200 µA
0.18 X 10-3 I + 30 nALaboratoryActive
  • >0.2 mA to 2 mA
0.15 X 10-3 I + 64 nALaboratoryActive
  • >2 mA to 20 mA
0.15 X 10-3 I + 0.24 µALaboratoryActive
  • >20 mA to 200 mA
0.15 X 10-3 I + 0.58 µALaboratoryActive
  • >0.2 A to 2 A
0.47 X 10-3 I + 93 µALaboratoryActive
  • >2 A to 20 A
0.70 X 10-3 I + 2.3 mALaboratoryActive
  • AC Voltage
  • Direct Method as per EURAMET Cg-15:2015 by using Multifunction Calibrator, Model No.: 5025E. U = measured voltage
  • (@50Hz) 0 mV to 200 mV
0.58 X 10-3 U + 0.12 mVLaboratoryActive
  • (@50Hz) >0.2 V to 2 V
0.47 X 10-3 U + 0.23 mVLaboratoryActive
  • (@50Hz) >2 V to 20 V
0.70 X 10-3 U + 2.3 mVLaboratoryActive
  • (@50Hz) >20 V to 200 V
0.70 X 10-3 U + 29 mVLaboratoryActive
  • (@50Hz) >200 V to 1000 V
1.0 X 10-3 U + 0.12 VLaboratoryActive
  • (@1Hz) 0 mV to 200 mV
0.58 X 10-3 U + 0.12 mVLaboratoryActive
  • (@1Hz) >0.2 V to 2 V
0.47 X 10-3 U + 0.23 mVLaboratoryActive
  • (@1Hz) >2 V to 20 V
0.70 X 10-3 U + 2.3 mVLaboratoryActive
  • (@1Hz) >20 V to 200 V
0.70 X 10-3 U + 29 mVLaboratoryActive
  • (@1Hz) >200 V to 1000 V
1.0 X 10-3 U + 0.12 VLaboratoryActive
  • (@10Hz) 0 mV to 200 mV
1.2 X 10-3 U + 0.17 mVLaboratoryActive
  • (@10Hz) >0.2 V to 2 V
0.70 X 10-3 U + 0.29 mVLaboratoryActive
  • (@10Hz) >2 V to 20 V
0.81 X 10-3 U + 3.5 mVLaboratoryActive
  • AC Current
  • Direct Method as per EURAMET Cg-15:2015 by using Multifunction Calibrator, Model No.: 5025E. I = measured current
  • (@50Hz) 20 µA to 200 µA
0.86 X 10-3 I + 0.35 µALaboratoryActive
  • (@50Hz) >0.2 mA to 2 mA
0.86 X 10-3 I + 0.42 µALaboratoryActive
  • (@50Hz) >2 mA to 20 mA
0.86 X 10-3 I + 0.23 mALaboratoryActive
  • (@50Hz) >20 mA to 200 mA
0.97 X 10-3 I + 0.23 mALaboratoryActive
  • (@50Hz) >0.2 A to 2 A
1.5 X 10-3 I + 2.4 mALaboratoryActive
  • (@50Hz) >2 A to 22 A
2.5 X 10-3 I + 7.3 mALaboratoryActive
  • (@400Hz) 20 µA to 200 µA
0.86 X 10-3 I + 0.35 µALaboratoryActive
  • (@400Hz) >0.2 mA to 2 mA
0.86 X 10-3 I + 0.35 µALaboratoryActive
  • (@400Hz) >2 mA to 20mA
0.86 X 10-3 I + 3.5 µALaboratoryActive
  • (@400Hz) >20 mA to 200 mA
0.97 X 10-3 I + 35 µALaboratoryActive
  • (@400Hz) >0.2 A to 2 A
1.5 X 10-3 I + 0.58 mALaboratoryActive
  • (@400Hz) >2 A to 22 A
2.5 X 10-3 I + 6.9 mALaboratoryActive
  • (@1Hz) 20 µA to 200 µA
0.86 X 10-3 I + 0.35 µALaboratoryActive
  • (@1Hz) >0.2 mA to 2 mA
0.86 X 10-3 I + 0.42 µALaboratoryActive
  • (@1Hz) >2 mA to 20 mA
0.86 X 10-3 I + 3.5 µALaboratoryActive
  • (@1Hz) >20 mA to 200mA
0.97 X 10-3 I + 35 µALaboratoryActive
  • 2 Wire Resistance
  • Direct Method as per EURAMET Cg-15:2015 by using Multifunction Calibrator, Model No.: 5025E. R = measured resistance
  • 10 Ω
0.59 X 10-3 R + 5.8 mΩLaboratoryActive
  • 100 Ω
0.10 X 10-3 R + 5.8 mΩLaboratoryActive
  • 1 kΩ
70 X 10-6 R + 5.8 mΩLaboratoryActive
  • 10 kΩ
70 X 10-6 R + 8.2 mΩLaboratoryActive
  • 100 kΩ
70 X 10-6 R + 58 mΩLaboratoryActive
  • 1 MΩ
0.17 X 10-3 R + 0.58 ΩLaboratoryActive
  • 10 MΩ
1.2 X 10-3 R + 5.8 ΩLaboratoryActive
  • 100 MΩ
13 X 10-3 R + 58 ΩLaboratoryActive
  • Frequency
  • Direct Method as per EURAMET Cg-15:2015 by using Multifunction Calibrator, Model No.: 5025E. F = measured frequency
  • 10 Hz
23 X 10-6 F + 5.8 µHzLaboratoryActive
  • 50 Hz
23 X 10-6 F + 58 µHzLaboratoryActive
  • 1 kHz
23 X 10-6 F + 0.58 mHzLaboratoryActive
  • 100 kHz
23 X 10-6 F + 58 mHzLaboratoryActive
  • Capacitance
  • Direct Method as per EURAMET Cg-15:2015 by using Multifunction Calibrator, Model No.: 5025E. C = measured Capacitance
  • 1 nF
2.4 X 10-3 C + 1.3 pFLaboratoryActive
  • 10 nF
2.4 X 10-3 C + 5.9 pFLaboratoryActive
  • 20 nF
2.4 X 10-3 C + 58 pFLaboratoryActive
  • 50 nF
2.4 X 10-3 C + 58 pFLaboratoryActive
  • 100 nF
2.4 X 10-3 C + 58 pFLaboratoryActive
  • 1 µF
2.4 X 10-3 C + 0.58 nFLaboratoryActive
Pressure
  • Analog and Digital Pressure Gauges Medium: Liquid
  • Direct Comparison method as per DKD-R 6-1:2014 by using Pressure Comparator
  • 0.01 MPa to 10 MPa
58 kPaLaboratoryActive
  • > 10 MPa to 100 MPa
70 kPaLaboratoryActive
  • > 100 MPa to 280 MPa
0.32 MPaLaboratoryActive
Dimensions
  • External Micrometer
  • MS-DMC-WP-057, based on BS EN ISO 3611 For determining error of indicated size, Comparison with reference gauge blocks
  • 0 mm to 25 mm
4.0 µmLaboratoryActive
  • Vernier Calipers
  • MS-DMC-WP-056, based on BS EN ISO 13385-1 For determining error of indicated size, Comparison with gauge blocks or calliper checker
  • 0 mm to 300 mm
30.0 µmLaboratoryActive
  • Height Gauges
  • MS-DMC-WP-058, based on ISO 13225 For determining error of indicated vertical size, Comparison with calibrated gauge blocks or calliper checker
  • 0 mm to 600 mm
35.0 µmLaboratoryActive
Mass and Balance
  • OIML Masses Class F1 and lower
  • As per OIML R-111-1:2004
  • 0.001 g
0.10 mgLaboratoryActive
  • 0.002 g
0.11 mgLaboratoryActive
  • 0.005 g
0.11 mgLaboratoryActive
  • 0.01 g
0.12 mgLaboratoryActive
  • 0.02 g
0.14 mgLaboratoryActive
  • 0.05 g
0.13 mgLaboratoryActive
  • 0.10 g
0.12 mgLaboratoryActive
  • 0.20 g
0.14 mgLaboratoryActive
  • 0.50 g
0.14 mgLaboratoryActive
  • 1.0 g
0.12 mgLaboratoryActive
  • 2 g
0.13 mgLaboratoryActive
  • 5 g
0.15 mgLaboratoryActive
  • 10 g
0.14 mgLaboratoryActive
  • 20 g
0.10 mgLaboratoryActive
  • 50 g
0.19 mgLaboratoryActive
  • 100 g
0.11 mgLaboratoryActive
  • 200 g
0.19 mgLaboratoryActive
  • Top Loading Direct reading weighing balance
  • As per OIML R-76-1:2006 & EURAMET cg-18:2015
  • 0 g to 200 g
0.90 mgClient PremisesActive
  • ˃ 200 g to 10 kg
0.60 gClient PremisesActive
  • ˃ 10 kg to 20 kg
0.95 gClient PremisesActive
  • ˃ 20 kg to 30 kg
1.4 gClient PremisesActive