PROMPT CALIBRATION AND INSTRUMENTATION EQUIPMENT TRADING LLC

Initial Accreditation Date:Dec 06 , 2017

Contacts Info

Address

AL Sarab Commercial Center, Musaffah-14, Plot No 70,, Abu Dhabi, United Arab Emirates

PO box / Zip Code

9580

Email

cal@promptuae.com sophia@promptuae.com

Website

www.promptuae.com

Phone No.

+971 02 558 5055

Mobile No.

+971 02 558 5910

Main Field

Calibration Laboratories

Valid Since:Mar 13 , 2024

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Electrical
  • DC Voltage Calibration of Measuring Instruments
  • P/CM/E/01 Based on Euramet cg-15 (Fluke 5502A MPC)
  • 0 V to 329.9999 mV
59 µVPrompt Laboratory/ Client PremisesActive
  • 0.33 V to 3.299999 V
0.58 mVPrompt Laboratory/ Client PremisesActive
  • 3.3 V to 32.99999 V
1.4 mVPrompt Laboratory/ Client PremisesActive
  • 33 V to 329.9999 V
58 mVPrompt Laboratory/ Client PremisesActive
  • 330 V to 1020 V
0.57 VPrompt Laboratory/ Client PremisesActive
  • AC Voltage Calibration of Measuring Instruments
  • P/CM/E/02 Based on Euramet cg-15 (Fluke 5502A MPC)
  • (50 Hz) 1.0 mV to 32.999 mV
70 µVPrompt Laboratory/ Client PremisesActive
  • (50 Hz) 33 mV to 329.999 mV
0.11 mVPrompt Laboratory/ Client PremisesActive
  • (50 Hz) 0.33 V to 3.29999 V
0.98 mVPrompt Laboratory/ Client PremisesActive
  • (50 Hz) 3.3 V to 32.9999 V
9.8 mVPrompt Laboratory/ Client PremisesActive
  • (50 Hz) 33 V to 329.999 V
0.14 VPrompt Laboratory/ Client PremisesActive
  • (50 Hz) 330 V to 1020 V
0.70 VPrompt Laboratory/ Client PremisesActive
  • (1 kHz) 1.0 mV to 32.999 mV
70 µVPrompt Laboratory/ Client PremisesActive
  • (1 kHz) 33 mV to 329.999 mV
0.11 mVPrompt Laboratory/ Client PremisesActive
  • (1 kHz) 0.33 V to 3.29999 V
0.98 mVPrompt Laboratory/ Client PremisesActive
  • (1 kHz) 3.3 V to 32.9999 V
9.8 mVPrompt Laboratory/ Client PremisesActive
  • (1 kHz) 33 V to 329.999 V
0.22 VPrompt Laboratory/ Client PremisesActive
  • (1 kHz) 330 V to 1020 V
0.85 VPrompt Laboratory/ Client PremisesActive
  • (10 kHz) 1.0 mV to 32.999 mV
70 µVPrompt Laboratory/ Client PremisesActive
  • (10 kHz) 33 mV to 329.999 mV
0.11 mVPrompt Laboratory/ Client PremisesActive
  • (10 kHz) 0.33 V to 3.29999 V
0.98 mVPrompt Laboratory/ Client PremisesActive
  • (10 kHz) 3.3 V to 32.9999 V
9.8 mVPrompt Laboratory/ Client PremisesActive
  • (10 kHz) 33 V to 329.999 V
0.22 VPrompt Laboratory/ Client PremisesActive
  • (10 kHz) 330 V to 1020 V
0.92 VPrompt Laboratory/ Client PremisesActive
  • DC Current Calibration of Measuring Instruments
  • P/CM/E/03 Based on Euramet cg-15 (Fluke 5502A MPC)
  • 0 A to 32.9999 mA
6.3 µAPrompt Laboratory/ Client PremisesActive
  • 33 mA to 329.999 mA
63 µAPrompt Laboratory/ Client PremisesActive
  • 330 mA to 1.09999 A
0.67 mAPrompt Laboratory/ Client PremisesActive
  • 1.1 A to 2.99999 A
1.1 mAPrompt Laboratory/ Client PremisesActive
  • 3 A to 10.9999 A
7.9 mAPrompt Laboratory/ Client PremisesActive
  • 11 A to 20 A
59 mAPrompt Laboratory/ Client PremisesActive
  • > 20 A to 1000 A
4.6 APrompt Laboratory/ Client PremisesActive
  • AC Current Calibration of measuring Instruments
  • P/CM/E/04 based on Euramet cg-15 (Fluke 5502A MPC)
  • (50 Hz) 0.029 mA to 32.9999 mA
13 µAPrompt Laboratory/ Client PremisesActive
  • (50 Hz) 33 mA to 329.999 mA
0.13 mAPrompt Laboratory/ Client PremisesActive
  • (50 Hz) 0.33 A to 1.09999 A
0.81 mAPrompt Laboratory/ Client PremisesActive
  • (50 Hz) 1.1 A to 2.99999 A
1.7 mAPrompt Laboratory/ Client PremisesActive
  • (50 Hz) 3 A to 10.9999 A
9.7 mAPrompt Laboratory/ Client PremisesActive
  • (50 Hz) 11 A to 20 A
62 mAPrompt Laboratory/ Client PremisesActive
  • (50 Hz) > 20 A to 1000 A
8.5 APrompt Laboratory/ Client PremisesActive
  • (1 kHz) 0.029 mA to 32.9999 mA
22 µAPrompt Laboratory/ Client PremisesActive
  • (1 kHz) 33 mA to 329.999 mA
0.30 mAPrompt Laboratory/ Client PremisesActive
  • (1 kHz) 0.33 A to 1.09999 A
5.8 mAPrompt Laboratory/ Client PremisesActive
  • (1 kHz) 1.1 A to 2.99999 A
14 mAPrompt Laboratory/ Client PremisesActive
  • (1 kHz) 3 A to 10.9999 A
0.25 APrompt Laboratory/ Client PremisesActive
  • (1 kHz) 11 A to 20 A
0.46 APrompt Laboratory/ Client PremisesActive
  • (10 kHz) 0.029 mA to 32.9999 mA
53 µAPrompt Laboratory/ Client PremisesActive
  • (10 kHz) 33 mA to 329.999 mA
0.59 mAPrompt Laboratory/ Client PremisesActive
  • (10 kHz) 0.33 A to 1.09999 A
25 mAPrompt Laboratory/ Client PremisesActive
  • (10 kHz) 1.1 A to 2.99999 A
61 mAPrompt Laboratory/ Client PremisesActive
  • DC Resistance calibration of Measuring Instruments
  • P/CM/E/05 based on Euramet cg-15 (Fluke 5502A MPC)
  • 0 Ω to 10.999 Ω
57 mΩPrompt Laboratory/ Client PremisesActive
  • 11 Ω to 32.999 Ω
57 mΩPrompt Laboratory/ Client PremisesActive
  • 33 Ω to 109.999 Ω
57 mΩPrompt Laboratory/ Client PremisesActive
  • 110 Ω to 329.999 Ω
62 mΩPrompt Laboratory/ Client PremisesActive
  • 330 Ω to 1.09999 kΩ
0.57 ΩPrompt Laboratory/ Client PremisesActive
  • 1.1 kΩ to 3.29999 kΩ
0.62 ΩPrompt Laboratory/ Client PremisesActive
  • 3.3 kΩ to 10.9999 kΩ
5.7 ΩPrompt Laboratory/ Client PremisesActive
  • 11 kΩ to 32.9999 kΩ
6.2 ΩPrompt Laboratory/ Client PremisesActive
  • 33 kΩ to 109.999 kΩ
57 ΩPrompt Laboratory/ Client PremisesActive
  • 110 kΩ to 329.999 kΩ
65 ΩPrompt Laboratory/ Client PremisesActive
  • 0.33 MΩ to 1.09999 MΩ
0.58 kΩPrompt Laboratory/ Client PremisesActive
  • 1.1 MΩ to 3.29999 MΩ
0.70 kΩPrompt Laboratory/ Client PremisesActive
  • 3.3 MΩ to 10.9999 MΩ
7.6 kΩPrompt Laboratory/ Client PremisesActive
  • 11 MΩ to 32.9999 MΩ
63 kΩPrompt Laboratory/ Client PremisesActive
  • 33 MΩ to 109.999 MΩ
0.42 MΩPrompt Laboratory/ Client PremisesActive
  • 110 MΩ to 329.999 MΩ
1.3 MΩPrompt Laboratory/ Client PremisesActive
  • 330 MΩ to 1100.00 MΩ
13 MΩPrompt Laboratory/ Client PremisesActive
  • CAPACITANCE Calibration of Measuring Instruments
  • P/CM/E/06 based on Euramet cg-15 (Fluke 5502A MPC)
  • 0.22 nF to 0.3999 nF
12 pFPrompt Laboratory/ Client PremisesActive
  • 0.4 nF to 1.099 nF
14 pFPrompt Laboratory/ Client PremisesActive
  • 1.1 nF to 3.2999 nF
21 pFPrompt Laboratory/ Client PremisesActive
  • 3.3 nF to 10.999 nF
30 pFPrompt Laboratory/ Client PremisesActive
  • 11 nF to 32.999 nF
0.15 nFPrompt Laboratory/ Client PremisesActive
  • 33 to 109.99 nF
0.64 nFPrompt Laboratory/ Client PremisesActive
  • 110 nF to 329.99 nF
1.0 nFPrompt Laboratory/ Client PremisesActive
  • 0.33 μF to 1.0999 μF
2.9 nFPrompt Laboratory/ Client PremisesActive
  • 1.1 μF to 3.2999 μF
10 nFPrompt Laboratory/ Client PremisesActive
  • 3.3 μF to 10.999 μF
30 nFPrompt Laboratory/ Client PremisesActive
  • 11 μF to 32.999 μF
0.14 µFPrompt Laboratory/ Client PremisesActive
  • 33 μF to 109.99 μF
0.74 µFPrompt Laboratory/ Client PremisesActive
  • Thermocouple Simulation (Type K) calibration of Measuring Instruments
  • P/CM/E/08 based on Euramet cg-15 (Fluke 5502A MPC)
  • -200 °C to -100 °C
0.28 °CPrompt Laboratory/ Client PremisesActive
  • -100 °C to -25 °C
0.12 °CPrompt Laboratory/ Client PremisesActive
  • -25 °C to 120 °C
0.18 °CPrompt Laboratory/ Client PremisesActive
  • 120 °C to 1000 °C
2.0 °CPrompt Laboratory/ Client PremisesActive
  • 1000 °C to 1372 °C
4.2 °CPrompt Laboratory/ Client PremisesActive
  • DC Voltage Calibration of sources
  • P/CM/E/10 (Fluke 8846A 6½ Digit Precision Multimeter)
  • 0 V to 100 mV
13 µVPrompt Laboratory/ Client PremisesActive
  • > 0.100 V to 1 V
61 µVPrompt Laboratory/ Client PremisesActive
  • > 1 V to 10 V
0.57 mVPrompt Laboratory/ Client PremisesActive
  • > 10 V to 100 V
11 mVPrompt Laboratory/ Client PremisesActive
  • > 100 V to 1000 V
64 mVPrompt Laboratory/ Client PremisesActive
  • AC voltage Calibration of sources
  • P/CM/E/11 (Fluke 8846A 6½ Digit Precision Multimeter)
  • (50 Hz) > 0.100 V to 1 V
1.6 mVPrompt Laboratory/ Client PremisesActive
  • (50 Hz) > 1 V to 10 V
14 mVPrompt Laboratory/ Client PremisesActive
  • (50 Hz) > 10 V to 100 V
0.12 VPrompt Laboratory/ Client PremisesActive
  • (50 Hz) > 100 V to 1000 V
1.1 VPrompt Laboratory/ Client PremisesActive
  • (10 kHz) > 0.100 V to 1 V
1.3 mVPrompt Laboratory/ Client PremisesActive
  • (10 kHz) > 1 V to 10 V
14 mVPrompt Laboratory/ Client PremisesActive
  • (10 kHz) > 10 V to 100 V
0.14 VPrompt Laboratory/ Client PremisesActive
  • (10 kHz) > 100 V to 1000 V
1.1 VPrompt Laboratory/ Client PremisesActive
  • DC Current Calibration of sources
  • P/CM/E/12 (Fluke 8846A 6½ Digit Precision Multimeter)
  • 0.100 mA to 10 mA
8.9 µAPrompt Laboratory/ Client PremisesActive
  • > 10 mA to 100 mA
74 µAPrompt Laboratory/ Client PremisesActive
  • > 0.100 A to 1 A
0.86 mAPrompt Laboratory/ Client PremisesActive
  • > 1 A to 3 A
4.3 mAPrompt Laboratory/ Client PremisesActive
  • > 3 A to 10 A
19 mAPrompt Laboratory/ Client PremisesActive
  • AC Current calibration of sources
  • P/CM/E/13 (Fluke 8846A 6½ Digit Precision Multimeter)
  • (50 Hz) 0.100 mA to 10 mA
70 µAPrompt Laboratory/ Client PremisesActive
  • (50 Hz) > 10 mA to 100 mA
0.71 mAPrompt Laboratory/ Client PremisesActive
  • (50 Hz) > 100 mA to 1 A
3.0 mAPrompt Laboratory/ Client PremisesActive
  • (50 Hz) > 1 A to 3 A
12 mAPrompt Laboratory/ Client PremisesActive
  • (50 Hz) > 3 A to 10 A
26 mAPrompt Laboratory/ Client PremisesActive
  • (1 kHz) 0.100 mA to 10 mA
59 µAPrompt Laboratory/ Client PremisesActive
  • (1 kHz) > 10 mA to 100 mA
0.67 mAPrompt Laboratory/ Client PremisesActive
  • (1 kHz) > 100 mA to 1 A
15 mAPrompt Laboratory/ Client PremisesActive
  • (1 kHz) > 1 A to 3 A
41 mAPrompt Laboratory/ Client PremisesActive
  • DC Resistance calibration of sources
  • P/CM/E/14 (Fluke 8846A 6½ Digit Precision Multimeter)
  • 1 Ω to 100 Ω
25 mΩPrompt Laboratory/ Client PremisesActive
  • > 0.100 kΩ to 1 kΩ
0.14 ΩPrompt Laboratory/ Client PremisesActive
  • > 1 kΩ to 100 kΩ
13 ΩPrompt Laboratory/ Client PremisesActive
  • > 0.100 MΩ to 1 MΩ
0.13 kΩPrompt Laboratory/ Client PremisesActive
  • > 1 MΩ to 10 MΩ
5.2 kΩPrompt Laboratory/ Client PremisesActive
  • > 10 MΩ to 100 MΩ
0.92 MΩPrompt Laboratory/ Client PremisesActive
  • > 0.100 GΩ to 1 GΩ
29 MΩPrompt Laboratory/ Client PremisesActive
  • Capacitance calibration of sources
  • P/CM/E/15 (Fluke 8846A 6½ Digit Precision Multimeter)
  • 0.1 nF to 1 nF
55 pFPrompt Laboratory/ Client PremisesActive
  • > 1 nF to 10 nF
0.19 nFPrompt Laboratory/ Client PremisesActive
  • > 10 nF to 100 nF
1.7 nFPrompt Laboratory/ Client PremisesActive
  • > 100 nF to 1 µF
18 nFPrompt Laboratory/ Client PremisesActive
  • > 1 µF to 10 µF
0.17 µFPrompt Laboratory/ Client PremisesActive
  • > 10 µF to 100 µF
1.8 µFPrompt Laboratory/ Client PremisesActive
  • > 0.100 mF to 1 mF
20 µFPrompt Laboratory/ Client PremisesActive
  • > 1 mF to 10 mF
0.18 mFPrompt Laboratory/ Client PremisesActive
  • > 10 mF to 100 mF
5.0 mFPrompt Laboratory/ Client PremisesActive
  • Thermocouple Simulation (Type K) calibration of Measuring Instruments
  • P/CM/E/18 based on Euramet cg-11 (Fluke 5502A MPC)
  • -200 °C to -100 °C
0.27 °CPrompt Laboratory/ Client PremisesActive
  • -100 °C to -25 °C
0.1 °CPrompt Laboratory/ Client PremisesActive
  • -25 °C to 120 °C
0.18 °CPrompt Laboratory/ Client PremisesActive
  • 120 °C to 1000 °C
2.0 °CPrompt Laboratory/ Client PremisesActive
  • 1000 °C to 1372 °C
4.2 °CPrompt Laboratory/ Client PremisesActive
  • High Resistance Generate
  • P/CM/E/19 Using Fluke 5320 Electrical Tester Calibrator
  • 10 kΩ to 39.99 kΩ
0.083 kΩPrompt LaboratoryActive
  • 40 kΩ to 99.99 kΩ
0.16 kΩPrompt LaboratoryActive
  • 100 kΩ to 199.99 kΩ
0.31 kΩPrompt LaboratoryActive
  • 200 kΩ to 999.9 kΩ
1.5 kΩPrompt LaboratoryActive
  • 1 MΩ to 9.999 MΩ
0.061 MΩPrompt LaboratoryActive
  • 10 MΩ to 999.9 MΩ
3.8 MΩPrompt LaboratoryActive
  • 1 GΩ to 10 GΩ
0.096 GΩPrompt LaboratoryActive
  • 10 GΩ to 100 GΩ
0.97 GΩPrompt LaboratoryActive
  • Using High Resistance Source
  • P/CM/E/19 Using Fluke 5320A HV Adapter/R Multiplier
  • 350 MΩ to 99.99 GΩ
0.78 GΩPrompt LaboratoryActive
  • 100 GΩ to 999.9 GΩ
15 GΩPrompt LaboratoryActive
  • 1 TΩ to 10 TΩ
0.24 TΩPrompt LaboratoryActive
  • Low Resistance (4 Wire)
  • P/CM/E/19 Using Fluke 5320 Electrical Tester Calibrator
  • 0.1 Ω to 4.99 Ω
0.020 ΩPrompt LaboratoryActive
  • 5 Ω to 29.9 Ω
0.078 ΩPrompt LaboratoryActive
  • 30 Ω to 199.9 Ω
0.32 ΩPrompt LaboratoryActive
  • 200 Ω to 499 Ω
0.95 ΩPrompt LaboratoryActive
  • 500 Ω to 1.999 kΩ
0.0031 kΩPrompt LaboratoryActive
  • 2 kΩ to 4.99 kΩ
0.057 kΩPrompt LaboratoryActive
  • 5 kΩ to 10 kΩ
0.059 kΩPrompt LaboratoryActive
  • Leakage current
  • P/CM/E/19 Using Fluke 5320 Electrical Tester Calibrator
  • 300 µA
1.0 µAPrompt LaboratoryActive
  • 3 mA
0.0083 mAPrompt LaboratoryActive
  • 30 mA
0.065 mAPrompt LaboratoryActive
  • RCD Trip Current
  • P/CM/E/19 Using Fluke 5320 Electrical Tester Calibrator
0.15 mAPrompt LaboratoryActive
  • 300 mA
1.4 mAPrompt LaboratoryActive
  • 3000 mA
13 mAPrompt LaboratoryActive
Pressure
  • Pneumatic Pressure (Gauge Pressure)
  • P-CM-M-09 based on BS EN 837-1 (1998) DKD-R 6-1 (03/2014)
  • -0.08 MPa to 0.2 MPa
0.31 kPaLaboratoryActive
  • > 0.2 MPa to 7 MPa
5.5 kPaLaboratoryActive
  • Hydraulic Pressure (gauge pressure)
  • 0 to 20 MPa
9.7 kPaLaboratoryActive
  • 20 MPa to 70 MPa
28 kPaLaboratoryActive
Temperature & Humidity
  • Digital Thermometers (Temperature Indicator with Sensor: RTD, Thermocouple), Transmitters, Temperature Switches, Dial Thermometers, Temperature Gauges, Temperature Controllers
  • P/CM/T/01 & P/CM/T/04
  • -20 °C to 0 °C
0.82 °CLaboratory & Customer PremisesActive
  • 0°C to 200°C
± 0.75°CLaboratory & Customer PremisesActive
  • >200°C to 660°C
± 1.1°CLaboratory & Customer PremisesActive
  • Dry Block Calibrators
  • P/CAL/T/05 based on EURAMET cg-13
  • -20 °C to 0 °C
0.82 °CLaboratoryActive
  • 0°C to 200°C
± 0.59°CLaboratoryActive
  • > 200°C to 700°C
± 1.2°CLaboratoryActive
  • IR Thermometer
  • P/CM/T/03
  • -30°C to 150°C
± 0.9°CLaboratoryActive
  • 150°C to 500°C
± 1.7°CLaboratoryActive
  • Temperature Chambers: Freezer/Chiller/ Refrigerator/Oven/Incubator/ hot cabinet, Autoclave (temperature only)
  • P/CM/T/07 & DKD R5-7
  • -30 °C to 250 °C
0.38 °CCustomer Premises/ LaboratoryActive
  • 250 °C to 300 °C
0.46 °CCustomer Premises/ LaboratoryActive
  • 300°C to 350 °C
0.70 °CCustomer Premises/ LaboratoryActive
  • Relative Humidity Meters, Thermo-Hygrometers, Data Loggers, Recorders calibrated for relative humidity
  • P/CM/T/06 Using Portable Temperature & Humidity Generator
  • 20% RH to 90 %RH @ 10 °C to 50 °C
3.9 % RHLaboratoryActive
  • Relative Humidity Meters, Thermo-Hygrometers, Data Loggers, Recorders calibrated for air temperature
  • 10 °C to 50 °C @ 20% RH to 90 %RH
0.9 °C to 1.1 °CLaboratoryActive