| Measured Quantity / Calibration Instrument |
Calibration Method |
Range and Specification |
Expanded Measurement |
Location |
Status |
| Balance & Mass Calibration |
| - Metromac Calibration Procedure MQS 03 C 75 based on ASTM898:20.
| | 1.4 mg | Customer Premises | Active
|
- OIML Masses Class F1 and Lower
| - Using E2 class W.I. MQS 03 C89 based on OIML R 111:2004
| | 0.01 mg | Metromac Premises | Active
|
- OIML Masses Class F2 and Lower
| - Using F1 & M1 class W.I. MQS 03 C89 based on OIML R 111:2004
| | 0.85 mg | Metromac Premises | Active
|
| Dimensional Calibration |
- Calipers Dial & Analogue & Digital
| - MQS 03 C 54 based on BS EN ISO 13385-12019 For determining error of indicated size Comparison with gauge blocks or calliper checker
| | 26 µm | Metromac Premises | Active
|
| - MQS 03 C 113 For determining error of indicated displacement, Comparison with reference measuring tape
| | 0.4 mm | Metromac Premises / Customer Premises | Active
|
- Ruler (Engineer or machinst scale)
| - MQS 03 C 112 Measurement of line spacing using length measuring machine equipped with optical magnification unit.
| 0.82 mm | Metromac Premises | Active
|
| - MQS 03 C 112 Measurement of line spacing using length measuring machine equipped with optical magnification unit. Stitching is carried out after length of 1 m.
| | (0.7 mm +0.12 x L) mm L : m | Metromac Premises | Active
|
- Dial gauges (Digital & Analogue)
| - "MQS 03 C 111 based on JIS B 7503 for Mechanicals and based on VDI/VDE/DGQ/DKD 2618 Part 11.4 for Digital dial gauges For determining error of indicated displacement, Comparison with dial gauge calibrator"
| - Up to 25 mm (Digital / 0.001 mm)
| 2.7 µm | Metromac Premises | Active
|
- Internal two point (bore) micrometer
| - MQS 03 C 110 For determining error of indicated size Comparison with gauge blocks and scale of the universal length measuring machine (ULM).
| | 10 µm | Metromac Premises | Active
|
- External Micrometer (Digital & Analogue)
| - MQS 03 C 109 based on BS EN ISO 3611-2023 For determining error of indicated size Comparison with reference gauge blocks
| | 3 µm | Metromac Premises | Active
|
| - Metromac calibration procedure MQS 03 C 164 based on JIS B 7503 2017 For determining error of indicated size Comparison with scale of the universal length measuring machine (ULM).
| | 6.4 µm | Metromac Premises | Active
|
- Thickness Gauge (Digital/Analogue)
| - Metromac calibration procedure MQS 03 C 197 For determining error of indicated size Comparison with calibrated gauge blocks
| | 2.5 µm | Metromac Premises | Active
|
| - Metromac calibration procedure MQS 03 C 162 based on EURAMET Guide -EURAMET cg-10. For determining Simple Pitch Diameter using ULM (three wire method) with nominal pitch: 0.5 mm to 6 mm.
| | 3.5 µm | Metromac Premises | Active
|
- Diamater Standards: Setting / Plain Plug Gauges
| - Metromac calibration Procedure MQS 03 C 176 based on EURAMET Calibration Guide No.6 (V 3.0) - without measuring form For determining diameter Comparison with ULM scale / gauge blocks
| | (1 + 0.010* D) µm, ( D: mm) | Metromac Premises | Active
|
- Diamater Standards: Setting / Plain Ring Gauge
| - Metromac calibration Procedure MQS 03 C 165 based on EURAMET Calibration Guide No.6 (V 3.0) - without measuring form For determining diamater Comparison with reference ring gauge using ULM
| | (1.7 + 0.04* D) µm, ( D: mm) | Metromac Premises | Active
|
- Parallel Thread Ring Gauge
| - Metromac calibration Procedure MQS 03 C 163 based on EURAMET Guide -EURAMET cg-10. For determining Simple Pitch Diameter using ULM (two sphere method) with nominal pitch: 1 mm to 6 mm.
| | (3.6 + 0.040* D) µm, ( D: mm) | Metromac Premises | Active
|
| Volume Calibration |
- Macro Pipette, Micro Pipette, Dispenser
| - MQS 03 C 70, For determining error by Gravimetric Method based on ISO 8655 and uncertainty based on Euramet cg 19.
| | 0.04 µl | Metromac Premises | Active
|
| - METROMAC CALIBRATION PROCEDURE MQS 03 C 121 based on ISO 4787:2021.
| | 0.32 ml | Metromac premises | Active
|
| | 0.031 ml | Metromac premises | Active
|
| | 0.4 ml | Metromac premises | Active
|
| | 0.21 ml | Metromac premises | Active
|
| Speed Calibration |
- Speed (Centrifuge / Variable Rotation Speed) eg. : Centrifuges, Mixers, Orbit & Plate shakers, MST Apparatus, Autoclaves
| | | 1.0 rpm | Metromac Premises & Customer Premises | Active
|
- Timer (Centrifuge / Variable Rotation Speed) eg. : Centrifuges, Mixers, Orbit & Plate shakers, MST Apparatus, Autoclaves
| | 0.81 s | Metromac Premises & Customer Premises | Active
|
| - Metromac calibration Procedure:MQS 03 C 144 Comparison with reference Stopwatch
| - upto 8 hr (Resolution : 1 s )
| 0.81 s | Metromac Premises & Customer Premises | Active
|
| | | 0.01 rpm | Metromac Premises | Active
|
| Electrical-Measuring |
| - MetroMac Calibration Procedure MQS 03 C 60 (U=Measured Voltage Value)
| - ( 1.0 mV to 32.999 mV ) 10 Hz to 45 Hz
| 0.48 x 10-3 U + 3.9 µV | Metromac Premises | Active
|
| - MetroMac
Calibration Procedure /
MQS 03 C 60
(U=Measured Voltage Value)
| | 11 x 10-6 U + 1.4 µV | Metromac Premises | Active
|
| - MetroMac Calibration Procedure MQS 03 C 60 (R=Measured Resistance Value)
| | 22 x 10-6 R + 5.4 mΩ | Metromac Premises | Active
|
| - MetroMac Calibration Procedure / MQS 03 C 60 (I=Measured Current Value)
| | 82 x 10-6 I + 11 nA | Metromac Premises | Active
|
| - MetroMac Calibration Procedure MQS03 C60 (I=Measured Current Value)
| - (29.00 µA to 329.99 µA) >1 kHz to 5 kHz
| 1.6 x 10-3 I + 82 nA | Metromac Premises | Active
|
| - MetroMac Calibration Procedure MQS 03 C 60 (C=Measured Capacitance Value)
| - (150 A to 1025 A) 0.19 to 0.3999 nF
| 3.1 x 10-3 C + 5.5 pF | Metromac Premises | Active
|
| - MetroMac Calibration Procedure MQS 03 C 60 (f=Measured Frequency Value)
| | 0.87 x 10-6 f + 0.08 mHz | Metromac Premises | Active
|
| - MetroMac Calibration Procedure MQS 03 C 60
| | 0.055 °C | Metromac Premises | Active
|
| - (J-Type) -210 °C to -100 °C
| 0.25 °C | Metromac Premises | Active
|
| Electrical-Source |
| - MetroMac Calibration Procedur MQS 03 C 124 (U= Source Voltage Value)
| - (10 mV to 100 mV ) >10 Hz to 20 kHz
| 0.34 x 10-3 U + 22 µV | Metromac Premises | Active
|
| - MetroMac Calibration Procedure MQS 03 C 124 (U= Source Voltage Value)
| - (>100 V to 1000 V) 0 mV to 100 mV
| 20 x 10-6 U + 2.2 µV | Metromac Premises | Active
|
| - MetroMac Calibration Procedure MQS 03 C 124 (R= Source Resistance Value)
| | 56 x 10-6 R + 1.6 mΩ | Metromac Premises | Active
|
| - MetroMac Calibration Procedure MQS 03 C 124 (I= Source Current Value)
| | 0.28 x 10-3 I + 14 nA | Metromac Premises | Active
|
| - MetroMac Calibration Procedure MQS 03 C 124 (I = Source Current Value)
| - (Upto 100 µA) 10 Hz to 5 kHz
| 0.55 x 10-3 I + 26 nA | Metromac Premises | Active
|
| - MetroMac Calibration Procedure MQS 03 C 124 (C= Source Capacitance Value)
| | 15 x 10-3 C + 14 pF | Metromac Premises | Active
|
| - MetroMac Calibration Procedure MQS 03 C 124 (f =Source Frequency Value)
| | 0.26 x 10-3 f + 0.21 mHz | Metromac Premises | Active
|
- RTD Source Pt 100 (-200 to 600)°C
| - MetroMac Calibration Procedure MQS 03 C 124
| | 0.10 °C | Metromac Premises | Active
|
- Thermocouple - Simulation method
| - (J type) -210 °C to -100 °C
| 0.24 °C | Metromac Premises | Active
|
| Electrical |
| - MetroMac Calibration Procedure MQS 03 C 73
| | 0.65 % rdg | Metromac Premises | Active
|
| | 0.1 % rdg | Metromac Premises | Active
|
| - MetroMac Calibration Procedure MQS 03 C 74
| | 1.2 % rdg | Metromac Premises | Active
|
| | 1.2 % rdg | Metromac Premises | Active
|
| | 1.2 % rdg | Metromac Premises | Active
|
| | 0.62 % rdg | Metromac Premises | Active
|
| | 0.59 ms | Metromac Premises | Active
|
| | 0.84 % rdg + 36mΩ | Metromac Premises | Active
|
| Temperaure Calibration |
- Air temperature & Relative Humidity Sensors
| - METROMAC calibration procedure MQS 03 C 122
| | 0.6 °C | Metromac Premises | Active
|
- Liquid in Glass Thermometer
| - METROMAC calibration procedure MQS 03 C 55
| | 0.05 °C | Metromac Premises | Active
|
| - METROMAC calibration procedure MQS 03 C 59
| 0.05 °C | Metromac Premises | Active
|
| - METROMAC calibration procedure MQS 03 C 107
| | 1.6 °C | Metromac Premises | Active
|
- Temperature Dry Block Calibrator
| - METROMAC calibration procedure MQS 03 C 24
| | 0.20 °C | Metromac Premises | Active
|
| - METROMAC calibration procedure MQS 03 C 76 (5 points)
| | 0.60 °C | Customer Premises | Active
|
| - METROMAC calibration procedure MQS 03 C 65
| | 0.45 °C | Customer Premises | Active
|
- Refrigerators/ Freezers (9 points)
| - METROMAC calibration procedure MQS 03 C 77
| | 0.55 °C | Customer Premises | Active
|
| - METROMAC calibration procedure MQS 03 C 40 (9 points)
| | 0.75 °C | Customer Premises | Active
|
| - METROMAC calibration procedure MQS 03 C 90
| | 3.0 °C | Customer Premises | Active
|
| - METROMAC calibration procedure MQS 03 C 96
| | 0.8 °C | Customer Premises | Active
|
| Pressure Calibration |
- Pneumatic Pressure (Gauge) Analog and digital indicating and electrical output devices
| - Comparison with pressure transducer using W.I. MQS 03 C 61 & MQS 03 C 117 Based on: DKD R 6-1 : 2014 & cg17: 2017.
| | 5 Pa + 0.02 % of reading | Metromac Premises and Customer Premises | Active
|
- Hydraulic Pressure (Gauge) Analog and digital indicating and electrical output devices
| - Comparison with DWT using W.I:MQS 03 C 56 & MQS 03 C 117 Based on DKD R 6 - 1: 2014 & cg 17: 2017
| | 0.025 % of reading | Metromac Premises | Active
|
| Torque Calibration |
| - Direct comparison with reference torque transducer using Work Instruction METROMAC calibration procedure MQS 03 C 115 Based on: ISO 6789-1:2017 and ISO 6789-2:2017
| - 0.04 N•m to 2 N•m Torque Transducer No:124712
| 0.015 | Metromac Premises | Active
|