Directory Details

Medflex Technical Services LLC

Initial Accreditation Date:Mar 05 , 2019

Contacts Info

Address

Ajman, United Arab Emirates

PO box / Zip Code

3336 Sharjah

Email

info@medflex.ae; a.ziad@medflex.ae

Website

www.medflex.ae

Main Field

Calibration Laboratories

Valid Since:Mar 05 , 2025

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Temperature
  • Thermometers with indicator and temperature data loggers with resistance sensor
  • Procedure CSOP-04, Sec. 5.1.1 based on DKD-R 5-1
  • -90 °C to -45 °C
0.5 °CLaboratory/ Client PremisesActive
  • -45 °C to 140 °C
0.2 °CLaboratory/ Client PremisesActive
  • Thermometers with indicator and temperature data loggers with TC sensor
  • Procedure CSOP-04, Sec. 5.1.1
  • -90 °C to -45 °C
2.0 °CLaboratory/ Client PremisesActive
  • -45 °C to 140 °C
0.6 °CLaboratory/ Client PremisesActive
  • Incubators, ovens, heat climatic chambers, furnace
  • Procedure CSOP-04, Sec. 5.1.2 based on DKD-R 5-7
  • 0 °C to 200 °C
0.4 °CLaboratory/ Client PremisesActive
  • 200 °C to 300 °C
0.7 °CLaboratory/ Client PremisesActive
  • Freezers, chiller, refrigerator, cold climatic chamber, thawing cabinets, refrigerated centrifuge
  • -90 °C to -40 °C
0.6 °CLaboratory/ Client PremisesActive
  • -40 °C to 37 °C
0.4 °CLaboratory/ Client PremisesActive
  • Liquid baths (using 5 sensors)
  • Procedure CSOP-04, Sec. 5.1.4 based on DKD-R 5-1 and 5-7
  • -40 °C to 200 °C
0.25 °CLaboratory/ Client PremisesActive
  • PCR system, Thermal Cycler, Reaction station (using 5 sensors)
  • Procedure CSOP-04, Sec. 5.1.5 based on DKD-R 5-4 and 5-7
  • 0 °C to 140 °C
0.4 °CLaboratory/ Client PremisesActive
  • Autoclave, sterilizer (using 5 sensors)
  • Internal Procedure CSOP-04, Sec. 5.1.6
  • 100 °C to 140 °C
0.4 °CLaboratory/ Client PremisesActive
  • Temperature block Calibrator
  • Procedure CSOP-04, Sec. 5.1.7 based on DKD-R 5-4
  • -45 °C to 140 °C
0.3 °CLaboratoryActive
  • Thermocouple sensors
  • Internal Procedure CSOP-04, Sec. 5.1.8
  • -90 °C to -45 °C
2.0 °CLaboratoryActive
  • -45 °C to 140 °C
0.6 °CLaboratoryActive
  • Platinum Resistance Thermometer, RTD
  • Procedure CSOP-04, Sec. 5.1.8 based on DKD-R 5-1
  • -90 °C to -45 °C
0.5 °CLaboratoryActive
  • -45 °C to 140 °C
0.2 °CLaboratoryActive
  • Thermo-hygrometer
  • Procedure CSOP-04 by comparison with humidity generator
  • -5 °C up to 50 °C (for temperature)
0.3 °CLaboratoryActive
  • 10 to 60 %RH
2.0 %RHLaboratoryActive
  • 60 to 90 %RH
2.8 %RHLaboratoryActive
  • Infrared Thermometer including medical Pyrometer
  • Procedure CSOP-04-B by direct measurement on Black Body
  • -20 °C to 0 °C
1.0 °CLaboratory/ Client PremisesActive
  • 0 °C to 25 °C
0.3 °CLaboratory/ Client PremisesActive
  • 25 °C to 50 °C
0.6 °CLaboratory/ Client PremisesActive
  • Medical Pyrometer: 35 °C to 42 °C
0.2 °CLaboratory/ Client PremisesActive
  • Thermo-hygrometer
  • Procedure CSOP-04-A by comparison with humidity generator
  • Fixed-Points: 21.61 to 23.38 %RH (10°C to 30°C) Potassium Acetate
1 %RHLaboratory/ Client PremisesActive
  • Fixed-Points: 29.93 to 33.47 %RH (10°C to 55°C) Magnesium Cloride
1 %RHLaboratory/ Client PremisesActive
  • Fixed-Points: 43.14 to 43.17 %RH (10°C to 30°C) Potassium Carbonate
1 %RHLaboratory/ Client PremisesActive
  • Fixed-Points: 45.44 to 57.36 %RH (10°C to 50°C) Magnesium Nitrate
2.0 %RHLaboratory/ Client PremisesActive
  • Procedure CSOP-04 using humidity fixed point (HFP) based on ASTM E104
  • Fixed-Points: 50.15 to 62.15 %RH (10°C to 55°C)Sodium Bromide
2.0 %RHLaboratory/ Client PremisesActive
  • Fixed-Points: 80.70 to 86.77 %RH (10°C to 55°C) Potassium Chloride
2.5 %RHLaboratory/ Client PremisesActive
  • Fixed-Points: 85.06 to 90.55 %RH (10°C to 25°C) Strontium Nitrate
2.5 %RHLaboratory/ Client PremisesActive
Dimensions
  • Length Calibration For microscope reticle without camera
  • By comparison with calibrated stage micrometer. In accordance to ASTM 1951 and Internal procedure CSOP-01 L: length in mm
  • 0.00 < L < 5.00
3.1 mmLaboratory/ Client PremisesActive
  • 5.00 < L < 20.0
3.2 mmLaboratory/ Client PremisesActive
  • Length Calibration For digital microscope with camera and software, digital endoscope, digital pathology and digital camera
  • 0.00 < L < 5.00
1.1 mmLaboratory/ Client PremisesActive
  • 5.00 < L < 20.0
1.3 mmLaboratory/ Client PremisesActive
  • Field of View for optical systems without camera
  • 0.00 < L < 5.00
3.1 mmLaboratory/ Client PremisesActive
  • 5.00 < L < 20.0
3.2 mmLaboratory/ Client PremisesActive
  • Field of View for optical systems with camera
  • 0.00 < L < 5.00
1.1 mmLaboratory/ Client PremisesActive
  • 5.00 < L < 20.0
1.3 mmLaboratory/ Client PremisesActive
  • Microscope Stage micrometer
  • 0.00 < L < 5.00
1.1 mmLaboratory/ Client PremisesActive
  • 5.00 < L < 20.0
1.3 mmLaboratory/ Client PremisesActive
  • Microscopic Diameter
  • 0.00 < L < 5.00
1.1 mmLaboratory/ Client PremisesActive
  • 5.00 < L < 20.0
1.3 mmLaboratory/ Client PremisesActive
  • Aspect Ratio of camera sensor and camera view screen image
  • By comparison with calibrated stage micrometer. In accordance to ASTM 1951 and Internal procedure CSOP-01
  • Width: length From 1:1 to 32:9
Width: length 0.1:1Laboratory/ Client PremisesActive
  • Magnification Calibration for microscope, digital microscope with camera and software, digital endoscope, digital pathology and digital camera
  • By comparison with calibrated stage micrometer. In accordance to ASTM 1951 and Internal procedure CSOP-01 M: X times magnification
  • 0.4 < M < 10
(2 x 10-5) XLaboratory/ Client PremisesActive
  • 10 < M < 50
0.01 XLaboratory/ Client PremisesActive
  • 50 < M < 170
0.09 XLaboratory/ Client PremisesActive
  • 170 < M < 981
2.0 XLaboratory/ Client PremisesActive
  • 981 < M < 4900
50.0 XLaboratory/ Client PremisesActive
  • 4900 < M < 12500
310 XLaboratory/ Client PremisesActive
  • Microscopic Area
  • By comparison with calibrated stage micrometer. In accordance to ASTM 1951 and Internal procedure CSOP-01 A: area in mm2
  • 8.2x10-5 < A < 7x10-4
3.2 x 10-4 mm2Laboratory/ Client PremisesActive
  • 7x10-4 < A < 0.2
0.002 mm2Laboratory/ Client PremisesActive
  • 0.2 < A < 20
0.016 mm2Laboratory/ Client PremisesActive
  • 20 < A < 491
0.096 mm2Laboratory/ Client PremisesActive
  • Microscope and endoscope resolution
  • By comparison with calibrated stage micrometer In accordance to ASTM 1951 and Internal procedure CSOP-01 R: Resolution in cycle/mm 1 cycle/mm = 0.5 * width of line
  • 0.25 < R < 100
3.5 mmLaboratory/ Client PremisesActive
  • 100 < R < 228
2.9 mmLaboratory/ Client PremisesActive
  • 228 < R < 500
2.9 mmLaboratory/ Client PremisesActive
  • 500 < R < 1000
2.9 mmLaboratory/ Client PremisesActive
Balance
  • Balance
  • EURAMET Calibration Guide 18, using calibrated masses class OIML: E2, F1, M1 and M2
  • 1 mg to 20 g
0.08 mgCustomer PremisesActive
  • >20 g to 50 kg
1.3 gCustomer PremisesActive
  • >50 to 300 kg
6.0 gCustomer PremisesActive
Frequency & Speed
  • Rotational Speed (Centrifuge)
  • CSOP-03 Direct comparison with Tachometer
  • 6 rpm to 999.9 rpm
0.47 rpmPermanent laboratory/ Customer PremisesActive
  • 1000 rpm to 12500 rpm
4.4 rpmPermanent laboratory/ Customer PremisesActive
  • 12501 rpm to 25000 rpm
7.7 rpmPermanent laboratory/ Customer PremisesActive
  • Rotational Speed (Orbital Mixer, Sample Mixer, Vortex)
  • CSOP-03 Direct comparison with Function Generator
  • 6 to 499.9 rpm
0.37 rpmPermanent laboratory/ Customer PremisesActive
  • 500 to 4000 rpm
2.9 rpmPermanent laboratory/ Customer PremisesActive
  • Tachometer (Non Contact)
  • 6 to 59.9 rpm
0.058 rpmPermanent laboratory/ Customer PremisesActive
  • 60 to 999.9 rpm
0.25 rpmPermanent laboratory/ Customer PremisesActive
  • 1000 to 12500 rpm
0.66 rpmPermanent laboratory/ Customer PremisesActive
  • 12501 to 25000 rpm
0.84 rpmPermanent laboratory/ Customer PremisesActive
  • 25001 to 50000 rpm
1.4 rpmPermanent laboratory/ Customer PremisesActive
  • 50001 to 99999 rpm
2.5 rpmPermanent laboratory/ Customer PremisesActive
Volume
  • Piston-operated Volumetric apparatus
  • Gravimetric – internal procedure CSOP-02 based on ISO 8655-6 :2022 & ISO/TR 20461:2023
  • ( 1.0 £ V < 5.0 ) μL
0.17 μLLaboratoryActive
  • ( 5.0 £ V < 10.0 ) μL
0.27 μLLaboratoryActive
  • (10.0 < V <100) μL
0.37 μLLaboratoryActive
  • (100 < V < 200 ) μL
0.52 μLLaboratoryActive
  • ( 200 £ V < 500 ) μL
0.57 μLLaboratoryActive
  • ( 0.5 £ V < 1 ) mL
1.2 μLLaboratoryActive
  • ( 1 £ V < 2 ) mL
4.0 μLLaboratoryActive
  • ( 2 £ V < 5 ) mL
12 μLLaboratoryActive
  • ( 5 £ V < 10 ) mL
29 μLLaboratoryActive
  • ( 10 £ V £ 20 ) mL
81 μLLaboratoryActive