| Measured Quantity / Calibration Instrument |
Calibration Method |
Range and Specification |
Expanded Measurement |
Location |
Status |
| Temperature |
- RTD sensor with/ without Indicator/Controller
| - GTS-WP-AD-04 (based on DKD-R 5-1)
| | 0.7 °C | Laboratory Permanent Premises | Active
|
- Thermocouple with/ without Indicator/Controller
| - GTS-WP-AD-03 (based on Euramet cg-8 Guidelines)
| | 0.7 °C | Laboratory Permanent Premises | Active
|
- Direct reading Thermometer, Dial Thermometer, Digital Thermometer with probe
| - GTS-WP-AD-04 (based on DKD-R 5-1)
| | 0.7 °C | Laboratory Permanent Premises | Active
|
- Temperature bath (at one single point)
| - GTS-WP-AD-05 (based on DKD-R 5-7 & Euramet cg-13)
| 0.7 °C | Laboratory Permanent Premises/ Customer premises | Active
|
| Pressure |
- Pressure (Pressure Indicating device -Digital and Analog Pressure Gauge, Pressure Sensor, Pressure Transducer) – Hydraulic Pressure
| - GTS-WP-AD-01 (based on DKD-R 6-1 Calibration of Pressure Gauges)
| | 0.16 bar | Customer Premises & Laboratory Permanent Premises | Active
|
| Electrical |
| - GTS-WP-AD-10E Comparison Method Using MultiFunction Calibrator (Based on EURAMET cg-15) C = Measured Capacitance Value
| | 25 pF | Laboratory Permanent Premises | Active
|
| - GTS-WP-AD-10E Comparison Method Using MultiFunction Calibrator (Based on EURAMET cg-15)
| | 0.25 A | Laboratory Permanent Premises | Active
|
| - 0.2 A to 2 A, x 1 turns upto 20 Hz
| 0.24 A | Laboratory Permanent Premises | Active
|
| - GTS-WP-AD-10E Comparison Method Using MultiFunction Calibrator (Based on EURAMET cg-15) U = Measured Voltage Value
| | 80 x 10-6 U + 1.7 μV | Laboratory Permanent Premises | Active
|
| - GTS-WP-AD-10E Comparison Method Using MultiFunction Calibrator (Based on EURAMET cg-15) I = Measured Current Value
| - > 10 μA to 200 μA: upto 20 Hz
| 2.6 x 10-3 I + 0.1 μA | Laboratory Permanent Premises | Active
|
| - GTS-WP-AD-10E Comparison Method Using MultiFunction Calibrator (Based on EURAMET cg-15) R = Measured Resistance Value
| | 63 x 10-6 R + 3.0 mΩ | Laboratory Permanent Premises | Active
|
| - GTS-WP-AD-12G Direct Method Using Insulation Resistance Box Upto 5 kV (SANAS TG 02-02)
| | 0.12 MΩ | Laboratory Permanent Premises | Active
|
| - GTS-WP-AD-10E Comparison Method Using MultiFunction Calibrator (Based on EURAMET cg-15) U = Measured Voltage Value
| - > 2 mV to 20 mV: up to 20 Hz
| 0.11 x 10-3 U + 12 μV | Laboratory Permanent Premises | Active
|
| - GTS-WP-AD-10E Comparison Method Using MultiFunction Calibrator (Based on EURAMET cg-15) I = Measured Current Value
| | 0.36 x 10-3 I + 58 nA | Laboratory Permanent Premises | Active
|
| - GTS-WP-AD-10E Comparison Method Using MultiFunction Calibrator (Based on EURAMET cg-15) F = Measured Frequency Value
| | 25 x 10-6 F | Laboratory Permanent Premises | Active
|
- Earth Resistance/Earth Bond Resistance/DC Resistance
| - GTS-WP-AD-11F Direct Method Using Precision Decade Resistance Box R = Measured Resistance Value
| | 11 x 10-3R + 74 mΩ | Laboratory Permanent Premises | Active
|