Directory Details

Geo Chem Middle East

Initial Accreditation Date:Feb 17 , 2020

Contacts Info

Address

Dubai, United Arab Emirates

PO box / Zip Code

2209

Email

sindhu.seelan@geochemglobal.com

Website

www.geochemglobal.com

Main Field

Calibration Laboratories

Valid Since:May 10 , 2023

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Temperature
  • Digital Thermometers with resistance sensor
  • In house method GC I&C-WI-010, based on DKD-R 5-1
  • 1 °C to 250 °C
0.3 °CLaboratoryActive
  • Digital Thermometers with thermocouple sensor
  • 1 °C to 80 °C
0.4 °CLaboratoryActive
  • In house method GC I&C-WI-010
  • 80 °C to 140 °C
0.6 °CLaboratoryActive
  • In house method GC I&C-WI-010, based on DKD-R 5-1
  • 140 °C to 250 °C
0.8 °CLaboratoryActive
  • Temperature controlled chambers (Chillers, Freezers, Incubators, Ovens)
  • In house method GC I&C-WI-013, based on DKD-R 5-7
  • ­30 °C to 100 °C
0.6 °CLaboratory/ Customer PremisesActive
  • Temperature liquid baths
  • 25 °C to 100 °C
0.3 °CLaboratory/ Customer PremisesActive
  • Muffel Frnace (one sensors)
  • GC I&C-WI-025 based on ASTM A 991/A991M-10
  • 500 °C to 1000 °C
5.1 °CCustomers PremisesActive
  • Liquid in Glass Thermometer
  • GC I&C-WI-019 based on BS 1041-2-1985
  • From -30 °C to 50 °C
0.2 °CLaboratoryActive
  • > 50 °C to 100 °C
0.3 °CLaboratoryActive
  • >100 °C to 250 °C
0.6 °CLaboratoryActive
Balance
  • Weighing Balance
  • EURAMET Calibration Guide No. 18 Version 4.0 (10/2015)
  • 0-5 g
0.2 mgCustomer PremisesActive
  • > 5 g-20 g
0.32 mgCustomer PremisesActive
  • > 20 g-50 g
0.4 mgCustomer PremisesActive
  • > 50 g-100 g
0.7 mgCustomer PremisesActive
  • > 100 g-150 g
1.0 mgCustomer PremisesActive
  • > 150 g-220 g
1.7 mgCustomer PremisesActive
  • > 200 g to 30 kg
3.0 gCustomer PremisesActive
  • > 30 kg to 200 kg
0.13 kgCustomer PremisesActive
Dimensions
  • Steel Rules
  • Scale & Tap WI 008 based on IS 1269 Part I & II, IS 1481, Measurement of line spacing using length measuring machine equipped with optical magnification unit. Stitching is carried out after length of 1 m
  • Up to 1 m
o.6 mmLaboratoryActive
  • Steel Tapes
o.6 mmLaboratoryActive
  • >1 m Up to 5 m
0.6 mmLaboratoryActive
Pressure
  • Pneumatic Pressure Gauge
  • DKD R-6-1
  • ­95 kPa to 4 Mpa
0.08% of Full ScaleLaboratory/ Customer PremisesActive
  • Hydraulic Pressure Gauge
  • 50 kPa to140 Mpa
0.05% of Full ScaleLaboratoryActive