Directory Details

Delta Wye Calibration and Testing Company L.L.C

Initial Accreditation Date:Jan 02 , 2017

Contacts Info

Address

Dubai, United Arab Emirates

PO box / Zip Code

PO BOX: 26704

Email

qc@deltawye.ae

Website

NA

Main Field

Calibration Laboratories

Valid Since:Apr 03 , 2023

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Electrical
  • DC Voltage
  • Euramet/cg-15/v.03
  • Up to 200 mV
15 µVLaboratoryActive
  • > 0.2 V up to 2.02 V
78 µVLaboratoryActive
  • > 2.02 V up to 20.2 V
0.81 mVLaboratoryActive
  • > 20.2 V up to 202 V
7.9 mVLaboratoryActive
  • > 202 V up to 1000 V
42 mVLaboratoryActive
  • AC Voltage
  • Up to 200 mV @ 40 Hz
0.52 mVLaboratoryActive
  • Up to 202 mV @ 1 kHz
0.25 mVLaboratoryActive
  • Up to 202 mV @ 20 kHz
0.80 mVLaboratoryActive
  • > 0.202 V up to 2.02 V @ 40 Hz
5.0 mVLaboratoryActive
  • > 0.202 V up to 2.02 V @ 1 kHz
2.3 mVLaboratoryActive
  • > 0.202 V up to 2.02 V @ 20 kHz
8.4 mVLaboratoryActive
  • > 2.02 V up to 20.2 V @ 40 Hz
50 mVLaboratoryActive
  • > 2.02 V up to 20.2 V @ 1 kHz
19 mVLaboratoryActive
  • > 2.02 V up to 20.2 V @ 20 kHz
91 mVLaboratoryActive
  • > 20.2 V up to 202 V @ 40 Hz
0.20 VLaboratoryActive
  • > 20.2 V up to 202 V @ 1 kHz
0.28 VLaboratoryActive
  • > 20.2 V up to 202 V @ 20 kHz
0.39 VLaboratoryActive
  • > 202 V up to 700 V @ 40 Hz
0.80 VLaboratoryActive
  • > 202 V up to 700 V @ 1 kHz
1.7 VLaboratoryActive
  • DC Current
  • Up to 202 µA
58 nALaboratoryActive
  • > 0.202 mA up to 2.02 mA
40 µALaboratoryActive
  • > 2.02 mA up to 20.2 mA
40 µALaboratoryActive
  • > 20.2 mA up to 202 mA
45 µALaboratoryActive
  • > 0.202 A up to 2.02 A
0.95 mALaboratoryActive
  • > 2.02 A up to 30 A
15 mALaboratoryActive
  • AC Current
  • 20 µA up to 202 µA @ 40 Hz
0.76 µALaboratoryActive
  • 20 µA up to 202 µA @ 1 kHz
2.1 µALaboratoryActive
  • > 0.202 mA up to 2.02 mA @ 40 Hz
5.2 µALaboratoryActive
  • > 0.202 mA up to 2.02 mA @ 1 kHz
17 µALaboratoryActive
  • > 2.02 mA up to 20.2 mA @ 40 Hz
52 µALaboratoryActive
  • > 2.02 mA up to 20.2 mA @ 1 kHz
0.12 mALaboratoryActive
  • > 20.2 mA up to 202 mA @ 40 Hz
0.52 mALaboratoryActive
  • > 20.2 mA up to 202 mA @ 1 kHz
1.5 mALaboratoryActive
  • > 0.202 A up to 2.02 A @ 40 Hz
5.5 mALaboratoryActive
  • > 0.202 A up to 2.02 A @ 1 kHz
15 mALaboratoryActive
  • > 2.02 A up to 20 A @ 40 Hz
53 mALaboratoryActive
  • > 2.02 A up to 20 A @ 1 kHz
74 mALaboratoryActive
  • Resistance (2 Wire)
  • 10 Ω
67 mΩLaboratoryActive
  • 100 Ω
70 mΩLaboratoryActive
  • 1 kΩ
0.15 ΩLaboratoryActive
  • 10 kΩ
0.97 ΩLaboratoryActive
  • 100 kΩ
9.3 ΩLaboratoryActive
  • 1 MΩ
0.16 kΩLaboratoryActive
  • 10 MΩ
4.5 kΩLaboratoryActive
  • 100 MΩ
0.58 MΩLaboratoryActive
  • Resistance (4 Wire)
  • 100 mΩ
5.8 mΩLaboratoryActive
  • 1 Ω
9.8 mΩLaboratoryActive
  • 10 Ω
46 mΩLaboratoryActive
  • 100 Ω
12 mΩLaboratoryActive
  • 1 kΩ
93 mΩLaboratoryActive
  • 10 kΩ
0.93 ΩLaboratoryActive
  • 100 kΩ
9.3 ΩLaboratoryActive
  • Capacitance
  • 10 nF
0.10 nFLaboratoryActive
  • 100 nF
0.45 nFLaboratoryActive
  • 1 µF
6.5 nFLaboratoryActive
  • 10 µF
0.11 µFLaboratoryActive
  • Frequency
  • 1 Hz up to 100 Hz
3.3 mHzLaboratoryActive
  • > 0.1 kHz up to 1 kHz
33 mHzLaboratoryActive
  • > 1 kHz up to 10 kHz
0.33 HzLaboratoryActive
  • > 10 kHz up to 100 kHz
3.3 HzLaboratoryActive
  • > 0.1 MHz to 1 MHz
33 HzLaboratoryActive
  • High Current DC
  • 10 A up to 100 A
1.1 ALaboratoryActive
  • > 100 A up to 1500 A
8.3 ALaboratoryActive
  • High Current AC
  • 10 A up to 100 A @ 40 Hz
0.41 ALaboratoryActive
  • > 100 A up to 1000 A @ 40 Hz
4.5 ALaboratoryActive
Pressure
  • Hygraulic pressure (gauge)
  • Based on requirement of DKD-R 6-1 pressure gauge calibration procedure DW-TR-PRO-01(2)Rev 05.
  • 0 MPa to 70MPa
42 kPaLaboratoryActive
  • Pneumatic pressure (gauge)
  • Based on requirement of DKD-R 6-1 pressure gauge calibration procedure DW-TR-PRO-01(2)Rev 05
  • 0 MPa to 2 MPa
1.2 kPaLaboratoryActive
Balance
  • Non-automatic Weighing Instruments
  • Based on the requirements of EURAMET cg-18 v4: 2015Procedure for the Calibration of Weighing ScaleDW-TR-PRO-01 (28) REV 04
  • 0 g - 1 g
0.40 mgCustomer PremisesActive
  • > 1 g - 2 g
0.51 mgCustomer PremisesActive
  • >2 g - 5 g
0.62 mgCustomer PremisesActive
  • >5 g - 10 g
0.74 mgCustomer PremisesActive
  • >10 g -20 g
0.97 mgCustomer PremisesActive
  • >20 g -50 g
1.2 mgCustomer PremisesActive
  • >50 g - 100 g
2.0 mgCustomer PremisesActive
  • > 100 g - 200 g
3.7 mgCustomer PremisesActive
  • >200 g - 500 g
9.8 mgCustomer PremisesActive
  • >500 g - 1000 g
19 mgCustomer PremisesActive
  • >1 kg - 2 kg
0.89 gCustomer PremisesActive
  • >2 kg - 5 kg
1.3 gCustomer PremisesActive
  • >5 kg - 10 kg
2.1 gCustomer PremisesActive
  • >10 kg - 20 kg
3.7 gCustomer PremisesActive
  • Euramet/cg-18/v 4.0
  • >20 kg - 40 kg
22 gCustomer PremisesActive
  • >40 kg - 100 kg
39 gCustomer PremisesActive
  • >100 kg - 150 kg
55 gCustomer PremisesActive
Temperature
  • Temperature chamber
  • "DW/TR/PRO/01(38) Rev01" Based on DKD-R 5-7
  • >-22 °C to 0 °C
0.91 °CCustomer PremisesActive
  • >0 °C to 4 °C
0.69 °CCustomer PremisesActive
  • >4 °C to 50 °C
1.19 °CCustomer PremisesActive
  • >50 °C to 100 °C
1.69 °CCustomer PremisesActive
  • >100 °C to 200 °C
2.7 °CCustomer PremisesActive