| Measured Quantity / Calibration Instrument |
Calibration Method |
Range and Specification |
Expanded Measurement |
Location |
Status |
| Balance & Scale |
- Top loading weighing scale
| | | | | Active
|
| Electrical |
- DC Voltage Calibration of meters
| - CI/01/M1 Based on Euramet cg-15 U: Voltage Measured value
| | 57 X 10-6U + 5.0 µV | Laboratory | Active
|
| 57 X 10-6U + 41 µV | Laboratory | Active
|
| 57 X 10-6U + 0.36 mV | Laboratory | Active
|
| 57 X 10-6U + 3.6 mV | Laboratory | Active
|
| 57 X 10-6U + 23 mV | Laboratory | Active
|
- AC Voltage Calibration of meters
| - CI/01/M1 Based on Euramet cg-15 U: Measured Voltage value
| - (0.001 V to 0.2 V) 45 Hz up to 1.999 kHz
| 0.33 X 10-3U + 52 µV | Laboratory | Active
|
- (0.001 V to 0.2 V) 2 kHz up to 19.999 kHz
| 1.2 X 10-3U + 0.23 mV | Laboratory | Active
|
- (> 0.2 V up to 2 V) 45 Hz up to 1.999 kHz
| 0.38 X 10-3U + 0.34 mV | Laboratory | Active
|
- (> 0.2 V up to 2 V) 2 kHz up to 19.999 kHz
| 0.65 X 10-3U + 0.59 mV | Laboratory | Active
|
- (> 2 V up to 20 V) 45 Hz up to 1.999 kHz
| 0.38 X 10-3U +3.4 mV | Laboratory | Active
|
- (> 2 V up to 20 V) > 2 kHz up to 19.999 kHz
| 0.77 X 10-3U + 5.8 mV | Laboratory | Active
|
- (> 20 V up to 200 V) 40 Hz up to 1.999 kHz
| 0.49 X 10-3U + 33 mV | Laboratory | Active
|
- (> 20 V up to 200 V) 2 kHz up to 20 kHz
| 1.1 X 10-3U + 50 mV | Laboratory | Active
|
- (> 200 V up to 1025 V) 40 Hz up to 1.999 kHz
| 0.49 X 10-3I + 0.19 V | Laboratory | Active
|
- (> 200 V up to 1025 V) > 2 kHz up to 10 kHz
| 1.1 X 10-3I + 0.38 V | Laboratory | Active
|
- DC Current Calibration of meters
| - CI/01/M1 Based on Euramet cg-15 I : Measured Current value
| | 0.14 X 10-6I+ 23 nA | Laboratory | Active
|
| 0.12 X 10-6I + 55 nA | Laboratory | Active
|
| 0.12 X 10-6I + 1.0 µA | Laboratory | Active
|
| 0.14 X 10-3I + 9.3 µA | Laboratory | Active
|
| 0.57 X 10-3I + 0.13 mA | Laboratory | Active
|
| 0.57 X 10-3I + 1.2 mA | Laboratory | Active
|
- DC Current Calibration of Meters (Current coil method 10-turn & 50-turn)
| - CI/01/C6 Based on Euramet cg-15 I : Measured Current value
| | 16 X 10-3I | Laboratory | Active
|
| 11 X 10-3I | Laboratory | Active
|
| 6.6 X 10-3I | Laboratory | Active
|
- AC Current Calibration of meters
| - CI/01/M1 Based on Euramet cg-15 I: Measured Current value
| - (2.5 µA up to 200 µA) 45 Hz up to 1.999 kHz
| 0.81 X 10-3I + 0.35 µA | Laboratory | Active
|
- (> 0.2 mA up to 2 mA) 45 Hz up to 1.999 kHz
| 0.77 X 10-3I + 1.0 µA | Laboratory | Active
|
- (> 2 mA up to 20 mA) 45 Hz up to 1.999 kHz
| 0.77 X 10-3I + 5.0 µA | Laboratory | Active
|
- (> 20 mA up to 200 mA) 45 Hz up to 1.999 kHz
| 0.77 X 10-3I + 50 µA | Laboratory | Active
|
- (> 0.2 A up to 2 A) 45 Hz up to 2 kHz
| 0.67 X 10-3I + 1.8 mA | Laboratory | Active
|
- (> 2 A up to 20 A) 45 Hz up to 200 Hz
| 1.7 X 10-3I + 8.2 mA | Laboratory | Active
|
- AC Current Calibration of Meters (Current coil method 10-turn & 50-turn)
| - CI/01/C6 Based on Euramet cg-15 I: Measured Current value
| - (20 A up to 100 A) 30 Hz up to 60 Hz
| 16 X 10-3I | Laboratory | Active
|
- (> 100 A up to 500 A) 30 Hz up to 60 Hz
| 13 X 10-3I | Laboratory | Active
|
- (> 500 A to 1000 A) 30 Hz to 60 Hz
| 6.6 X 10-3I | Laboratory | Active
|
- Capacitance Calibration of Meters
| - CI/01/M1 Based on Euramet cg-15
| | 58 pF | Laboratory | Active
|
| 0.10 nF | Laboratory | Active
|
| 0.24 nF | Laboratory | Active
|
| 0.48 nF | Laboratory | Active
|
| 10 nF | Laboratory | Active
|
| - CI/01/M1 Based on Euramet cg-15 f :Measured Frequency Value
| | 2.0 X 10-6f+ 18 µHz | Laboratory | Active
|
| 2.0 X 10-6f+ 90 µHz | Laboratory | Active
|
| 2.0 X 10-6f+ 0.90 mHz | Laboratory | Active
|
- 10.000 kHz up to 19.999 kHz
| 2.0 X 10-6f+ 0.59 mHz | Laboratory | Active
|
- 20.000 kHz up to 49.999 kHz
| 2.0 X 10-6f+ 1.8 mHz | Laboratory | Active
|
- 50000 kHz up to 99.999 kHz
| 2.0 X 10-6f+ 30 mHz | Laboratory | Active
|
- 100.000 kHz up to 999.999 kHz
| 2.0 X 10-6f+ 91 mHz | Laboratory | Active
|
| 2.0 X 10-6f+ 0.90 Hz | Laboratory | Active
|
- Resistance Calibration of Meters 2-wire
| - CI/01/M1 Based on Euramet cg-15
| | 9.0 mΩ | Laboratory | Active
|
| 13 mΩ | Laboratory | Active
|
| 18 mΩ | Laboratory | Active
|
| 0.12 Ω | Laboratory | Active
|
| 1.2 Ω | Laboratory | Active
|
| 12 Ω | Laboratory | Active
|
| 0.22 kΩ | Laboratory | Active
|
| 7.4 kΩ | Laboratory | Active
|
| 0.13 MΩ | Laboratory | Active
|
- Resistance Calibration of Meters 4-wire
| | 4.5 mΩ | Laboratory | Active
|
| 13 mΩ | Laboratory | Active
|
| 17 mΩ | Laboratory | Active
|
| 0.12 Ω | Laboratory | Active
|
| 1.2 Ω | Laboratory | Active
|
| 12 Ω | Laboratory | Active
|
- Insulation Resistance Calibration of Meters
| - CI/01/I2 Based on SANAS guideline:TG-02-01
| | 80 kΩ | Laboratory | Active
|
| 0.62 MΩ | Laboratory | Active
|
| 60 MΩ | Laboratory | Active
|
| 0.10 GΩ | Laboratory | Active
|
- DC Voltage (Above 1000 V) Calibration of Meters
| - CI/01/I2 Based on SANAS guideline:TG-02-01 U=measured value
| | 33 V | Laboratory | Active
|
| 63 V | Laboratory | Active
|
- AC Voltage @ 60 Hz (Above 1000 V) Calibration of Meters
| - CI/01/I2 Based on SANAS guideline:TG-02-01
| | 0.15 kV | Laboratory | Active
|
| 0.34 kV | Laboratory | Active
|
| Energy Measurement |
| - Direct Method Based on EN50470 E: Measured value
| - (• Voltage: 30 V to 500 V, • Frequency: 40 Hz to 65 Hz, • Power Factor: Cos Φ= 1 to 0.5 (Lead and Lag)) 10 mA to 12 A (Cos Φ = 1)
| 0.07 % E | Laboratory and Customer Premise | Active
|
- (• Voltage: 30 V to 500 V, • Frequency: 40 Hz to 65 Hz, • Power Factor: Cos Φ= 1 to 0.5 (Lead and Lag)) 10 mA to 12 A (Cos Φ = 0.5 to <1)
| 0.08 % E | Laboratory and Customer Premise | Active
|
- (• Voltage: 30 V to 500 V, • Frequency: 40 Hz to 65 Hz, • Power Factor: Cos Φ= 1 to 0.5 (Lead and Lag)) >12 A to 120 A (Cos Φ = 0.5 to 1)
| 0.08 % E | Laboratory and Customer Premise | Active
|
- Reactive Energy (Offline)
| - Direct Method Based on EN50470 Er: Measured value
| - (• Voltage: 30 V to 500 V, • Frequency: 40 Hz to 65 Hz, • Power Factor: Sin Φ= 1 to 0.5 (Lead and Lag)) 10 mA to 12 A (Sin Φ = 1)
| 0.08 % Er | Laboratory and Customer Premise | Active
|
- (• Voltage: 30 V to 500 V, • Frequency: 40 Hz to 65 Hz, • Power Factor: Sin Φ= 1 to 0.5 (Lead and Lag)) 10 mA to 12 A (Sin Φ = 0.5 to <1)
| 0.08 % Er | Laboratory and Customer Premise | Active
|
- (• Voltage: 30 V to 500 V, • Frequency: 40 Hz to 65 Hz, • Power Factor: Sin Φ= 1 to 0.5 (Lead and Lag)) >12 A to 120 A (Sin Φ = 0.5 to 1)
| 0.08 % Er | Laboratory and Customer Premise | Active
|
- Active Energy (Online) Measurement of V, I, PF as set by the Customer
| - Direct Method using Current Clamp Based on EN50470 E: Measured value
| - (• Voltage: 30 V to 500 V, • Frequency: 40 Hz to 65 Hz, • Power Factor: Cos Φ:= 1 to 0.5 (Lead and Lag)) 10 mA to 12 A (Cos Φ = 1)
| 0.24 % E | Customer Premise | Active
|
- (• Voltage: 30 V to 500 V, • Frequency: 40 Hz to 65 Hz, • Power Factor: Cos Φ:= 1 to 0.5 (Lead and Lag)) 10 mA to 12 A (Cos Φ = 0.5 to <1)
| 0.24 % E | Customer Premise | Active
|
- (• Voltage: 30 V to 500 V, • Frequency: 40 Hz to 65 Hz, • Power Factor: Cos Φ:= 1 to 0.5 (Lead and Lag)) >12 A to 120 A (Cos Φ = 0.5 to 1)
| 0.24 % E | Customer Premise | Active
|
- Reactive Energy (Online) Measurement of V, I, PF as set by the Customer
| - Direct Method using Current Clamps Based on EN50470 Er: Measured value
| - (• Voltage: 30 V to 500 V, • Frequency: 40 Hz to 65 Hz, • Power Factor: Sin Φ:= 1 to 0.5 (Lead and Lag)) 10 mA to 12 A (Sin Φ = 1)
| 0.24 % Er | Customer Premise | Active
|
- (• Voltage: 30 V to 500 V, • Frequency: 40 Hz to 65 Hz, • Power Factor: Sin Φ:= 1 to 0.5 (Lead and Lag)) 10 mA to 12 A (Sin Φ = 0.5 to <1)
| 0.24 % Er | Customer Premise | Active
|
- (• Voltage: 30 V to 500 V, • Frequency: 40 Hz to 65 Hz, • Power Factor: Sin Φ:= 1 to 0.5 (Lead and Lag)) >12 A to 120 A (Sin Φ = 0.5 to 1)
| 0.24 % Er | Customer Premise | Active
|
| Dimensions |
- All types of Caliper Digital / Dial / Vernier
| - Procedure (CI/05/V3) based on ISO 13385-1-2019
| | 20 μm | Laboratory | Active
|
- External Micrometer (Digital & Analogue)
| - Procedure (CI/05/O2) based on ISO 3611 :2010
| | 3.0 μm | Laboratory | Active
|
- Dial gauges (Digital & Analogue)
| - Procedure (CI/05/D6) based on ASME B 89-1.10 M-2001 with limited number of calibration points.
| | 4.0 μm | Laboratory | Active
|
| Mass and Balance |
- Non-automatic Weighing Instruments
| - CI/04/W1 in accordance to Euramet cg.18(V 4.0)
| | 2.6 x 10-5 FS of UUC | Customer Premises | Active
|
| 4 x 10-4 FS of UUC | Customer Premises | Active
|
| Temperature & Humidity |
- Direct reading thermometers/Transmitters/Controllers/Temperature chart recorders
| | | 0.3°C | Laboratory/ Customer Premises | Active
|
| 1.3 °C | Laboratory/ Customer Premises | Active
|
- Electrical Calibration of Temperature Indicators Indicators for TC type K
| | | 0.59 °C | Laboratory/ Customer Premises | Active
|
- Electrical Calibration of Temperature Indicators Indicators for TC type J
| | 0.48 °C | Laboratory/ Customer Premises | Active
|
- Electrical Calibration of Temperature Indicators Indicators for TC type T
| | 0.48 °C | Laboratory/ Customer Premises | Active
|
- Electrical Calibration of Temperature Indicators Indicators for TC type R
| | 0.82 °C | Laboratory/ Customer Premises | Active
|
- Electrical Calibration of Temperature Indicators Indicators for TC type S
| 0.93 °C | Laboratory/ Customer Premises | Active
|
- Electrical Calibration of Temperature Indicators Indicators for TC type N
| | 0.59 °C | Laboratory/ Customer Premises | Active
|
- Electrical Calibration of Temperature Indicators Indicators for TC type E
| | 0.49 °C | Laboratory/ Customer Premises | Active
|
- Electrical Calibration of Temperature Indicators Indicators for RTDs
| | 0.39 °C | Laboratory/ Customer Premises | Active
|
| Pressure |
- Pneumatic Pressure (Gauge Pressure)
| - CI-02-PI2, BS 837-1 EURAMET /cg-17/v.04 DKD-R 6-1
| | 0.26 kPa | Laboratory | Active
|
| 3 kPa | Laboratory | Active
|
- Hydraulic Pressure (gauge pressure)
| | 30 kPa | Laboratory/ Customer Premises | Active
|