Directory Details

Procal Measuring & Control Equipment

Initial Accreditation Date:Jan 13 , 2015

Contacts Info

Address

Dubai, United Arab Emirates

PO box / Zip Code

118478

Email

procal@emirates.net.ae

Website

www.procalme.com

Main Field

Calibration Laboratories

Valid Since:Apr 01 , 2024

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Temperature
  • Digital Thermometer (temperture sensor with digital temperature indicator)
  • In-house method No. WI-24
  • ­30 °Ϲ up to 400 °Ϲ
0.3 °ϹLaboratory / Customer PremisesActive
  • > 400 °Ϲ up to 550 °Ϲ
0.7 °ϹLaboratory / Customer PremisesActive
  • > 550 °Ϲ up to 650 °Ϲ
1.0 °ϹLaboratory / Customer PremisesActive
  • Dial Thermometer (temperature gauge)
  • In-house method No. WI-15
  • ­30 °Ϲ up to 140 °Ϲ
0.6 °ϹLaboratory / Customer PremisesActive
  • > 140 °Ϲ up to 300 °Ϲ
0.5 °ϹLaboratory / Customer PremisesActive
  • > 300 °Ϲ up to 650 °Ϲ
1.0 °ϹLaboratory / Customer PremisesActive
  • Temperature Dry Block Calibrator
  • In-house method No.WI-50 based on Euramet cg-13 V 3.0/2015
  • ­30 °Ϲ up to 150 °Ϲ
0.3 °ϹLaboratoryActive
  • > 150 °Ϲ up to 650 °Ϲ
2.0 mK*t/ °ϹLaboratoryActive
  • Chiller , Freezer, Refrigerator and Cooling Cabinate (using single sensor or 9 sensors)
  • In-house method No. WI-62, based on Euramet cg-20, V. 05 (09-2017)
  • -30 to 25 ºC
0.6 ºCCustomer PremisesActive
  • Incubator, Hot Cabinet and Oven (using single sensor or 9 sensors)
  • In-house method No. WI-62, based on Euramet cg-20 V. 05 (09-2017)
  • 25 to 200 ºC
0.8Customer PremisesActive
  • > 200 to 650 ºC
4.9Customer PremisesActive
Pressure
  • Pneumatic Gauge Pressure and Vacuum Sensor, Transmitter
  • Work Instructions WI-17 EURAMET/cg-17/v.04 & DKD-R 6-1 2014sing Digital Pressure Gauge
  • ­95 kPa to 0 kPa
2.8 kPaLaboratory/ Customer PremisesActive
  • 0 up to 4 MPa
3.2 kPaLaboratory/ Customer PremisesActive
  • Hydraulic Gauge Pressure Hydraulic Gauge Pressure, Sensor, Transmitter
  • 0 MPa up to 4 MPa
17 kPaLaboratory/ Customer PremisesActive
  • >4 MPa up to 10 MPa
0.6 MPaLaboratory/ Customer PremisesActive
  • >10 MPa up to 160 MPa
1.1 MPaLaboratory/ Customer PremisesActive
  • >160 MPa up to 250 MPa
1.5 MPaLaboratory/ Customer PremisesActive
Electrical
  • DC Voltage Calibration of meters
  • Direct Method Using Multi-Function Calibrator according to Procal Calibration Procedure WI-19
  • 1 mV to 20 mV
5.6 µVLaboratory/ Customer PremisesActive
  • 21 mV to 200 mV
10 μVLaboratory/ Customer PremisesActive
  • 201 mV to 2 V
0.10 mVLaboratory/ Customer PremisesActive
  • 2.1 V to 20 V
1.0 mVLaboratory/ Customer PremisesActive
  • 20.1 V to 200 V
10 mVLaboratory/ Customer PremisesActive
  • 200.1 V to 1000 V
48 mVLaboratory/ Customer PremisesActive
  • AC Voltage Calibration of meters
  • 1 mV to 200 mV 60 Hz to 1 kHz
0.46 mVLaboratory/ Customer PremisesActive
  • 201 mV to 2 V 60 Hz to 1 kHz
1.2 mVLaboratory/ Customer PremisesActive
  • 2.1 V to 20 V 60 Hz to 1 kHz
13 mVLaboratory/ Customer PremisesActive
  • 20.1 V to 200 V 40 Hz to 1 kHz
0.17 VLaboratory/ Customer PremisesActive
  • 200.1 V to 1000 V 40 Hz to 1 kHz
1.2 VLaboratory/ Customer PremisesActive
  • DC Current Calibration of meters
  • 1 μA to 200 μA
29 nALaboratory/ Customer PremisesActive
  • 201 μA to 2 mA
0.31 μALaboratory/ Customer PremisesActive
  • 2.1 mA to 20 mA
3.1 μALaboratory/ Customer PremisesActive
  • 21 mA to 200 mA
30 μALaboratory/ Customer PremisesActive
  • 0.201 A to 2 A
0.70 mALaboratory/ Customer PremisesActive
  • 2.1 A to 20 A
14 mALaboratory/ Customer PremisesActive
  • Direct Method Using Multi-Function Calibrator according to Procal Calibration Procedure WI-40
  • >20 A to 50 A
0.62 ALaboratory/ Customer PremisesActive
  • 51 A to 100 A
2.2 ALaboratory/ Customer PremisesActive
  • 101 A to 250 A
2.7 ALaboratory/ Customer PremisesActive
  • 251 A to 500 A
5.0 ALaboratory/ Customer PremisesActive
  • 501 A to 750 A
8.7 ALaboratory/ Customer PremisesActive
  • 751 A to 1000 A
9.4 ALaboratory/ Customer PremisesActive
  • AC Current Calibration of meters
  • Direct Method Using Multi-Function Calibrator according to Procal Calibration Procedure WI-19
  • 1 uA to 200 uA 50 Hz to 200 Hz
0.57 µALaboratory/ Customer PremisesActive
  • 201 uA to 2 mA 50 Hz to 200 Hz
2.5 µALaboratory/ Customer PremisesActive
  • 2.1 mA to 20 mA 50 Hz to 200 Hz
22 uALaboratory/ Customer PremisesActive
  • 21 mA to 200 mA 50 Hz to 200 Hz
0.18 mALaboratory/ Customer PremisesActive
  • 0.201 A to 2 A 50 Hz to 200 Hz
3.3.mALaboratory/ Customer PremisesActive
  • 2.1 A to 20 A 50 Hz to 200 Hz
59 mALaboratory/ Customer PremisesActive
  • Direct Method Using Multi-Function Calibrator according to Procal Calibration Procedure WI-40
  • >20 A to 50 A 50 Hz to 200 Hz
0.90 ALaboratory/ Customer PremisesActive
  • 51 A to 100 A 50 Hz to 200 Hz
1.4 ALaboratory/ Customer PremisesActive
  • 101 A to 250 A 50 Hz to 200 Hz
3.7 ALaboratory/ Customer PremisesActive
  • 251 A to 500 A 50 Hz to 200 Hz
6.4 ALaboratory/ Customer PremisesActive
  • 501 A to 750 A 50 Hz to 200 Hz
9.5 ALaboratory/ Customer PremisesActive
  • 751 A to 1000 A 50 Hz to 200 Hz
11 ALaboratory/ Customer PremisesActive
  • Resistance Calibration of meters
  • Direct Method Using Multi-Function Calibrator according to Procal Calibration Procedure WI-19, WI-40 & WI-45
  • 1 Ω
6.8 mΩLaboratory/ Customer PremisesActive
  • 10 Ω
6.0 mΩLaboratory/ Customer PremisesActive
  • 100 Ω
7.1 mΩLaboratory/ Customer PremisesActive
  • 1 kΩ
71 mΩLaboratory/ Customer PremisesActive
  • 10 kΩ
1.2 ΩLaboratory/ Customer PremisesActive
  • 100 kΩ
12 ΩLaboratory/ Customer PremisesActive
  • 1 MΩ
0.21 kΩLaboratory/ Customer PremisesActive
  • 10 MΩ
12 kΩLaboratory/ Customer PremisesActive
  • 100 MΩ
1.2 MΩLaboratory/ Customer PremisesActive
  • 1 GΩ
0.12 GΩLaboratory/ Customer PremisesActive
  • Insulation Resistance Calibration of Meters
  • Direct Method Using Insulation Tester Calibrator according to Procal Calibration Procedure WI-19 & WI-03
  • 100 kΩ
1.2 kΩLaboratory/ Customer PremisesActive
  • 1 MΩ to 10 MΩ
0.13 MΩLaboratory/ Customer PremisesActive
  • 11 MΩ to 100 MΩ
1.2 MΩLaboratory/ Customer PremisesActive
  • 101 MΩ to 500 MΩ
6.6 MΩLaboratory/ Customer PremisesActive
  • 501 MΩ to 990 MΩ
0.12 GΩLaboratory/ Customer PremisesActive
  • 1 GΩ to 10 GΩ
0.14 GΩLaboratory/ Customer PremisesActive
  • 11 GΩ to 100 GΩ
5.8 GΩLaboratory/ Customer PremisesActive
  • Frequency Calibration of Meters
  • Direct Method Using Multi-Function Calibrator according to Procal Calibration Procedure WI-19 & WI-40 f = measured frequency
  • 20 Hz to 50 Hz
0.10 x 10-3 fLaboratory/ Customer PremisesActive
  • 51 Hz to 1000 Hz
0.10 x 10-3 fLaboratory/ Customer PremisesActive
  • >1 kHz to 10 kHz
0.10 x 10-3 fLaboratory/ Customer PremisesActive
  • 11 kHz to 10 MHz
0.58 x 10-3 fLaboratory/ Customer PremisesActive
  • 1.1 MHz to 10 MHz
6 x 10-3 fLaboratory/ Customer PremisesActive
  • Temperature Simulation Calibration of meters
  • Direct Method Using Multi-Function Calibrator according to Procal Calibration Procedure WI-39
  • K-Type -200 °C to 1300 °C
0.48 °CLaboratory/ Customer PremisesActive
  • J-Type -200 °C to 1200 °C
0.48 °CLaboratory/ Customer PremisesActive
  • N-Type -200 °C to 1300 °C
0.98 °CLaboratory/ Customer PremisesActive
  • E-Type -200 °C to 1000 °C
0.67 °CLaboratory/ Customer PremisesActive
  • T-Type -200 °C to 400 °C
0.57 °CLaboratory/ Customer PremisesActive
  • R-Type -50 °C to 1700 °C
2.0 °CLaboratory/ Customer PremisesActive
  • S-Type -50 °C to 1700 °C
1.9 °CLaboratory/ Customer PremisesActive
  • B-Type 300 °C to 1800 °C
2.3 °CLaboratory/ Customer PremisesActive
Mass
  • Mass
  • Standard Weights as per OIML R111-1 by using F1 from 1 mg to 20 kg according to Procal Calibration procedure WI-64
  • 1 mg
0.19 mgLaboratoryActive
  • 2 mg
0.19 mgLaboratoryActive
  • 5 mg
0,19 mgLaboratoryActive
  • 10 mg
0.21 mgLaboratoryActive
  • 20 mg
0.22 mgLaboratoryActive
  • 50 mg
0.22 mgLaboratoryActive
  • 100 mg
0.22 mgLaboratoryActive
  • 200 mg
0.22 mgLaboratoryActive
  • 500 mg
0.22 mgLaboratoryActive
  • 1 g
1 mgLaboratoryActive
  • 2 g
2 mgLaboratoryActive
  • Standard Weights as per OIML R111-1 by using F1 from 1 mg to 20 kg according to Procal Calibration procedure WI-64
  • 5 g
2 mgLaboratoryActive
  • 10 g
2mgLaboratoryActive
  • 20 g
2mgLaboratoryActive
  • 50 g
2mgLaboratoryActive
  • 100 g
2mgLaboratoryActive
  • 200 g
2mgLaboratoryActive
  • 500 g
3mgLaboratoryActive
  • 1000 g
14.0 mgLaboratoryActive
  • 2000 g
14.0 mgLaboratoryActive
  • 5000 g
42.0 mgLaboratoryActive
  • 10000 g
0.40 gLaboratoryActive
  • 20000 g
0.40 gLaboratoryActive
Balance
  • Non Automatic Weighing Scales
  • Using F1 class masses then M1 masses up to 10 ton Procal Calibration Procedure WI-29 & In accordance with Euramet Guide cg 18
  • 1 mg up to 500g
2.2 mgLaboratory/ Customer PremisesActive
  • >500 g up to 5000 g
5.5 mgLaboratory/ Customer PremisesActive
  • >5000 g up to 10000 g
55 mgLaboratory/ Customer PremisesActive
  • >10 kg up to 100 kg
2.2 mgLaboratory/ Customer PremisesActive
  • >100 kg up to 1 ton
2.6 gLaboratory/ Customer PremisesActive
  • >1 ton up to 10 ton
95 gLaboratory/ Customer PremisesActive