Verification Laboratory Verlab Ltd

Initial Accreditation Date:Jul 22 , 2026

Contacts Info

Address

Bosnia and Herzegovina

Email

lejla@verlab.ba

Phone No.

undefined

Mobile No.

undefined

Main Field

Inspection Bodies

Valid Since:Jul 22 , 2026

Inspection Category Inspection field Range of inspection Stage of Inspection Inspection criteria Inspection Activity Type Status
Medical Devices
  • Product
  • Perfusion Pump (Syringe Pump)
  • Flow [0.1 – 100.0] [ml/h]
  • In-service
Rulebook on metrological and technical requirements for infusomats and perfusors (Official Gazette of Bosnia and Herzegovina, number 75/14- Pages 42–44) Badnjević, A., Cifrek, M., Magjarević, R., & Džemić, Z. (Eds.). (2024). Inspection of medical devAActive
  • Infusion Pump
  • Flow [0.1 - 1000.0] [ml/h]
Rulebook on metrological and technical requirements for infusomats and perfusors (Official Gazette of Bosnia and Herzegovina, number 75/14 - Pages 42–44); Badnjević, A., Cifrek, M., Magjarević, R., & Džemić, Z. (Eds.). (2024). Inspection of medical dAActive
  • Pediatric Incubator limited to temperature and Humidity
  • Air temperature [13 – 42] [°C]Skin temperature [13 – 43] [°C]Relative humidity [10 – 99] [%]
Rulebook on metrological and technical requirements for neonatal and pediatric incubators (Official Gazette of Bosnia and Herzegovina, number 75/14 - Pages 46–47); Badnjević, A., Cifrek, M., Magjarević, R., & Džemić, Z. (Eds.). (2024). Inspection ofAActive
  • Blood pressure devices
  • Noninvasive blood pressure [0 – 300] [mmHg]
Rulebook on measuring conditions for manometers for measuring blood pressure -sphygmomanometers (Official Gazette of Bosnia and Herzegovina, no.71/23 - Pages 175–178); Badnjević, A., Cifrek, M., Magjarević, R., & Džemić, Z. (Eds.). (2024). InspectionAActive
  • Dialysis Machine
  • Conductivity [10 – 17] [mS/cm]Temperature [35 – 39] [° C]Pressure [- 300 – 500] [mmHg]
Rulebook on metrological and technical requirements for dialysis devices (Official Gazette of Bosnia and Herzegovina, number 75/14 - – Pages 35–36); Badnjević, A., Cifrek, M., Magjarević, R., & Džemić, Z. (Eds.). (2024). Inspection of medical devicesAActive
  • Mechanical Ventilator
  • Volume [-1.00 - 4.00] [L]Flow [-300 – 200] [L/min]Output pressure [-60 – 140] [cmH2O] (−58,8399 mBar to 137,2931 mBar)
Rulebook on metrological and technical requirements for mechanical ventilators and anesthesia machines (Official Gazette of Bosnia and Herzegovina, number 75/14 – Pages 88–89); Badnjević, A., Cifrek, M., Magjarević, R., & Džemić, Z. (Eds.). (2024). IAActive
  • Anesthesia Machine
  • Volume [-1.00 - 4.00] [l]Flow [-300 – 200] [L/min]Output pressure [-60 – 140] [cmH2O] (−58,8399 mBar to 137,2931 mBar)Concentration of anaesthesia gases {CO2: 0 - 10%, NO2: 0 - 100%, HAL, ISO, ENF: 0 – 8%, SEV: 0 - 10%, DES: 0 - 22%}
Rulebook on metrological and technical requirements for mechanical ventilators and anesthesia machines (Official Gazette of Bosnia and Herzegovina, number 75/14 – Pages 88–89); Badnjević, A., Cifrek, M., Magjarević, R., & Džemić, Z. (Eds.). (2024). IAActive
  • Electrocardiographs
  • Amplitude of voltage signal [0.5 - 2.0] [mV]Heart rate in time interval of 1 min [30 – 300] [BPM]
Rulebook on Metrological and Technical Requirements for Electrocardiographs, Official Gazette of BiH No. 75/14 – Pages 38–40); Badnjević, A., Cifrek, M., Magjarević, R., & Džemić, Z. (Eds.). (2024). Inspection of medical devices: For regulatory purpoAActive
  • Defibrillator
  • Output energy [2-360] [J]
Rulebook on Metrological and Technical Requirements for Defibrillators, Official Gazette of BiH No. 75/14 – page 31); Badnjević, A., Cifrek, M., Magjarević, R., & Džemić, Z. (Eds.). (2024). Inspection of medical devices: For regulatory purposes (2ndAActive
  • Patient Monitors
  • Signal [0.5 - 2.0] [mV]Heart rate in time interval of 1 min [15 – 300] [BPM]Respiration [0 – 150] [brpm]Noninvasive blood pressure [10 – 270] [mmHg]Invasive blood pressure [-50 – 300] [mmHg]Body temperature [30 – 43] [°C]Standard SpO2 [0-100] [%]
Rulebook on Metrological and Technical Requirements for Patient Monitors, Official Gazette of BiH No. 75/14 - Pages 56–60); Badnjević, A., Cifrek, M., Magjarević, R., & Džemić, Z. (Eds.). (2024). Inspection of medical devices: For regulatory purposesAActive
  • Therapeutic Ultrasound Devices
  • Output power [0 – 30] [W]
Rulebook on metrological and technical requirements for therapeutic ultrasound devices (Official Gazette of Bosnia and Herzegovina, number 75/14- pages 93-94); Badnjević, A., Cifrek, M., Magjarević, R., & Džemić, Z. (Eds.). (2024). Inspection of mediAActive