Arabian Calibration Devices Rep Co LLC Sole Proprietorship

Initial Accreditation Date:Oct 07 , 2010

معلومات الاتصال

العنوان

Industrial Area 13, Near Rameez Hypermarket, Sharjah, Sharjah, United Arab Emirates

صندوق بريد / الرمز البريدي

24650

البريد الإلكتروني

qc@arabcal.com eby.joy@arabcal.com

الموقع الإلكتروني

www.arabcal.com

رقم الهاتف

+971 06 534 5747

رقم الجوال

+971 06 534 5727

المجال الرئيسي

Calibration Laboratories

ساري منذ:Nov 11 , 2025

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Temperature
  • Direct reading thermometers/Transmitters/Controllers/Temperature chart recorders
  • Internal Procedure: CI/10/T3 DKD R5-1-2003
  • ­20 °C to 120 °C
0.2 °CLaboratory PremisesActive
  • 120 °Ϲ to 300 °C
0.3 °CLaboratory PremisesActive
  • 300 °Ϲ to 600 °C
0.8 °CLaboratory PremisesActive
  • 600 °Ϲ to 1000 °C
1.5 °CLaboratory PremisesActive
  • ­20 °C to 120 °C
0.27 °CCustomer PremisesActive
  • 120 °Ϲ to 300 °C
0.5 °CCustomer PremisesActive
  • 300 °Ϲ to 600 °C
1.0 °CCustomer PremisesActive
  • 600 °Ϲ to 1000 °C
1.7 °CCustomer PremisesActive
  • Direct reading thermometers
  • ­40 °C to 150 °C
0.07 °CLaboratory PremisesActive
  • Clinical Contact Thermometers
  • Internal Procedure: CI/10/CT1
  • 13 °C to 90 °C
0.2 ° CLaboratory PremisesActive
  • Ovens Refrigerators Freezers Incubators ( 9 points)
  • Internal Procedure: CI/05/CC1 EURAMET- R5-7
  • ­30 °C to 100 °C
0.6 °CLaboratory Premises / Customer PremiseActive
  • 100 °Ϲ to 140 °C
0.8 °CLaboratory Premises / Customer PremiseActive
  • 140 °C to 500 °C
1.5 °CLaboratory Premises / Customer PremiseActive
  • 500 °C – 800°C
2.5 °CLaboratory Premises / Customer PremiseActive
  • 800 °C– 1000 °C
5.0 °CLaboratory Premises / Customer PremiseActive
  • Liquid-in-glass thermometer
  • Internal Procedure: CI/03/G3
  • ­20 °C to 200 °C
0.15 °CLaboratory PremisesActive
  • Autoclave (1 to 5 sensors)
  • Internal Procedure: CI/05/AC3
  • 100 °C to 135 °C
0.8 °CCustomer PremisesActive
  • Dry Block Calibrators
  • Internal Procedure: CI/03/DBC1 Euramet cg-13
  • ­30 °C to 300 °C
0.25 °CLaboratory PremisesActive
  • 300 °C to 500 °C
0.7 °CLaboratory PremisesActive
  • 500 °C to 6500 °C
1.5 °CLaboratory PremisesActive
  • Infrared Thermometer
  • Internal Procedure: CI/03/LTM1 ASTM E2847 - 14
  • ­30 °C to 150 °C
2 °CLaboratory PremisesActive
  • 150 °C to 300 °C
3 °CLaboratory PremisesActive
  • 300 °C to 600 °C
4 °CLaboratory PremisesActive
  • Indicators for TC type K
  • Internal Procedure: CI/03/ECT1 EURAMET cg-11
  • ­270 °C to 1300 °C
0.2 °CLaboratory PremisesActive
  • Indicators for TC type J
  • ­210 °C to 1200 °C
0.2 °CLaboratory PremisesActive
  • Indicators for TC type T
  • ­270 °C to 400 °C
0.2 °CLaboratory PremisesActive
  • Indicators for TC type R
  • ­50 °C to 1340 °C
0.6 °CLaboratory PremisesActive
  • Indicators for TC type S
0.6 °CLaboratory PremisesActive
  • Indicators for TC type N
  • ­270 °C to 1300 °C
0.2 °CLaboratory PremisesActive
  • Indicators for TC type E
  • ­270 °C to 1372 °C
0.2 °CLaboratory PremisesActive
  • Indicators for RTDs
  • ­200 °C to 850 °C
0.07 °CLaboratory PremisesActive
  • Thermo hygrometer
  • Internal Procedure: CI/03/TH1
  • 15 °C to 40 °C
0.25 °CLaboratory PremisesActive
  • RH: 30 % to 70 %
2.5 % RHLaboratory PremisesActive
  • Liquid Bath
  • Internal Procedure: CI/03/B9 Euramet cg-20
  • ­30 °C to 200 °C
0.25 °CLaboratory Premises / Customer PremisesActive
  • Climatic Chamber
  • Internal Procedure: CI/05/CC1 Based on Euramet cg-20
  • 5 °C to 50 °C
1.8 °CLaboratory Premises / Customer PremisesActive
  • 10 % to 90 %RH
2.6 % RHLaboratory Premises / Customer PremisesActive
  • Thermo hygrometer
  • Internal Procedure: CI/03/TH1
  • 5 °C to 50 °C
0.3°CLaboratory PremisesActive
  • 10 % to 50 % rH
1.6% RHLaboratory PremisesActive
  • > 50 % rH to 90 % rH
2.3% RHLaboratory PremisesActive
  • Direct reading thermometers/ Temperature transmitters/ Controllers/ Temperature chart recorders
  • Internal Procedure: CI/10/T3 Based on DKD-R 5-1
  • -80 °C to -40 °C
0.1 °CLaboratory PremisesActive
  • > -40 °C to 150 °C
0.07 °CLaboratory PremisesActive
  • 150 °Ϲ to 300 °C
0.3 °CLaboratory PremisesActive
  • 300 °Ϲ to 600 °C
0.8 °CLaboratory PremisesActive
  • 600 °Ϲ to 1000 °C
1.5 °CLaboratory PremisesActive
  • Liquid Bath
  • Internal Procedure: CI/03/B9 Based on Euramet cg-20
  • -80°C to -20°C
0.25°CLaboratory Premises / Customer PremisesActive
  • -20°C to 200°C
0.15°CLaboratory Premises / Customer PremisesActive
  • Liquid in Glass Thermometer
  • Internal Procedure: CI/03/G3 Based on OIML R-133
  • -80°C to -20°C
0.2°CLaboratory PremisesActive
  • -20°C to 200°C
0.15 °CLaboratory PremisesActive
  • Resistance Thermometer (RTD)
  • Internal Procedure: CI/03/PRT1 Based on DKD-R 5-1
  • -80 °C to 150 °C
0.1 °CLaboratory PremisesActive
  • 150 °Ϲ to 300 °C
0.3 °CLaboratory PremisesActive
  • 300 °Ϲ to 660 °C
0.8 °CLaboratory PremisesActive
  • TC type B
  • Internal Procedure: 'CI/03/ECT1 based on Euramet cg-11
  • 600 °C to 1800 °C
0.4 °CLaboratory PermisesActive
  • TC type C
  • 0 °C to 2316 °C
0.6 °CLaboratory PermisesActive
  • TC type E
  • -250 °C to 1000 °C
0.2 °CLaboratory PermisesActive
  • TC type J
  • -210 °C to 1200 °C
0.3 °CLaboratory PermisesActive
  • TC type K
  • -200 °C to 1372 °C
0.2 °CLaboratory PermisesActive
  • TC type L
  • -200 °C to 900 °C
0.3 °CLaboratory PermisesActive
  • TC type N
  • -200 °C to 1300 °C
0.2 °CLaboratory PermisesActive
  • TC type R
  • 0 °C to 1767 °C
0.5 °CLaboratory PermisesActive
  • TC type S
0.4 °CLaboratory PermisesActive
  • TC type T
  • -250 °C to 400 °C
0.2 °CLaboratory PermisesActive
  • TC type U
  • -200 °C to 600 °C
0.4 °CLaboratory PermisesActive
Electrical
  • DC Voltage Calibration of meters
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15 U=Measured value
  • Up to 0.2 V
35 X10-6U + 4.3 µVLaboratory PremisesActive
  • >0.2 V to 2 V
34X10-6U + 7.7 µVLaboratory PremisesActive
  • >2 V to 20 V
28X10-6U + 67 µVLaboratory PremisesActive
  • >20 V to 200 V
34X10-6U + 0.65 mVLaboratory PremisesActive
  • >200 V to 1025 V
35X10-6U + 5.1 mVLaboratory PremisesActive
  • AC Voltage Calibration of meters
  • (Up to 0.2 V) 10 Hz to 45 Hz
2.3X10-3U + 59 µVLaboratory PremisesActive
  • (Up to 0.2 V) >45 Hz to 1 kHz
0.38X10-3U + 49 µVLaboratory PremisesActive
  • (Up to 0.2 V) >1 kHz to 20 kHz
0.92X10-3U + 75 µVLaboratory PremisesActive
  • (>0.2 V to 2 V) 10 Hz to 45 Hz
2.3X10-3U + 0.44 mVLaboratory PremisesActive
  • (>0.2 V to 2 V) >45 Hz to 999 kHz
0.42X10-3U + 0.23 mVLaboratory PremisesActive
  • (>0.2 V to 2 V) >1 kHz to 20 kHz
0.96X10-3U + 0.35 mVLaboratory PremisesActive
  • (>0.2 V to 2 V) >20 kHz to 50 kHz
2.9X 10-3U + 2.3 mVLaboratory PremisesActive
  • (>2 V to 20 V) 10 Hz to 45 Hz
2.3X 10-3U + 3.9 mVLaboratory PremisesActive
  • (>2 V to 20 V) >45 Hz to 2 kHz
0.36X 10-3U +2.3 mVLaboratory PremisesActive
  • (>2 V to 20 V) >1 kHz to 20 kHz
0.72X10-3U + 3.7 mVLaboratory PremisesActive
  • (>20 V to 200 V) 30 Hz to 45 Hz
0.66X 10-3U + 31 mVLaboratory PremisesActive
  • (>20 V to 200 V) >45 Hz to 1 kHz
0.41X 10-3U + 22 mVLaboratory PremisesActive
  • (>20 V to 200 V) >1 kHz to 19.999 kHz
1.0X 10-3U + 58 mVLaboratory PremisesActive
  • (>200 V to 1025 V) 30 Hz to 45 Hz
0.67X 10-3U + 0.26 VLaboratory PremisesActive
  • (>200 V to 1025 V) >45 Hz to 1 kHz
0.39X 10-3U + 0.20 VLaboratory PremisesActive
  • (>200 V to 1025 V) >1 kHz to 10 kHz
1.7X10-3U + 0.49 VLaboratory PremisesActive
  • DC Current Calibration of meters
  • Up to 200 µA
0.12 X10-3U + 35 nALaboratory PremisesActive
  • >0.2 mA to 2 mA
89X10-6U + 55 nALaboratory PremisesActive
  • >2 mA to 20 mA
53X 10-6U + 0.49 µALaboratory PremisesActive
  • >20 mA to 200 mA
92 X 10-6U + 3.5 µALaboratory PremisesActive
  • >0.2 A to 2 A
0.15X 10-3U + 91 µALaboratory PremisesActive
  • >2 A to 20 A
0.45X 10-3U + 0.50 mALaboratory PremisesActive
  • DC Current, Calibration of Meters (Current coil method 10-turn & 50-turn)
  • Internal Procedure: CI/01/C6 Based on Euramet cg-15 U=Measured value
  • 20 A to 100 A
6.8 X 10-3U + 0.20 ALaboratory PremisesActive
  • >100 A to 500 A
6.8 X 10-3U + 0.70 ALaboratory PremisesActive
  • >500 A to 1000 A
6.8 X 10-3U + 0.70 ALaboratory PremisesActive
  • AC Current Calibration of meters
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15 U=Measured value
  • (Up to 200 µA) 10 Hz to 45 Hz
2.3 X 10-3U + 0.29 µALaboratory PremisesActive
  • (Up to 200 µA) >45 Hz to 1 kHz
0.80 X 10-3U + 0.29 µALaboratory PremisesActive
  • (Up to 200 µA) >1 kHz tp 10 kHz
0.78 X 10-3U + 0.38 µALaboratory PremisesActive
  • (>0.2 mA to 2 mA) 10 Hz to 45 Hz
2.3 X 10-3U + 0.66 µALaboratory PremisesActive
  • (>0.2 mA to 2 mA) >45 Hz to 1 kHz
0.65 X 10-3U + 0.59 µALaboratory PremisesActive
  • (>0.2 mA to 2 mA) >1 kHz tp 10 kHz
0.54 X 10-3U + 2.0 µALaboratory PremisesActive
  • (>2 mA to 20 mA) 10 Hz to 45 Hz
2.3 X 10-3U + 6.6 µALaboratory PremisesActive
  • (>2 mA to 20 mA) >45 Hz to 1 kHz
0.65 X 10-3U + 6.6 µALaboratory PremisesActive
  • (>2 mA to 20 mA) >1 kHz to 10 kHz
5.5 X 10-3U + 5.9 µALaboratory PremisesActive
  • (>20 mA to 200 mA) 10 Hz to 45 Hz
2.3 X 10-3U + 66 µALaboratory PremisesActive
  • (>20 mA to 200 mA) >45 Hz to 1 kHz
0.65 X 10-3U + 59 µALaboratory PremisesActive
  • (>20 mA to 200 mA) >1 kHz to 10 kHz
6.6 X 10-3U + 0.15 mALaboratory PremisesActive
  • (>0.2 A to 2 A) 10 Hz to 45 Hz
2.3 X 10-3U + 0.70 mALaboratory PremisesActive
  • (>0.2 A to 2 A) >45 Hz to 1 kHz
0.98 X 10-3U + 0.64 mALaboratory PremisesActive
  • (>0.2 A to 2 A) >1 kHz to 5 kHz
6.6 X 10-3U + 1.5 mALaboratory PremisesActive
  • (>2 A to 20 A) 30 Hz to 45 Hz
2.3 X 10-3U + 7 mALaboratory PremisesActive
  • (>2 A to 20 A) >45 Hz to 100 Hz
0.88 X 10-3U + 4.9 mALaboratory PremisesActive
  • (>2 A to 20 A) >100 Hz to 1 kHz
3.4 X 10-3U + 5.8 mALaboratory PremisesActive
  • AC Current Calibration of Meters (Current coil method 10-turn & 50-turn)
  • Internal Procedure: CI/01/C6 Based on Euramet cg-15 U=Measured value
  • (>20 A to 100 A) 30 Hz to 60 Hz
6.8 X 10-3U + 0.20 ALaboratory PremisesActive
  • (>100 A to 1000 A) 30 Hz to 60 Hz
6.8 X 10-3U + 0.70 ALaboratory PremisesActive
  • Resistance Calibration of Meters 2-wire
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15
  • 0.2 Ω
5.0 mΩLaboratory PremisesActive
  • 1 Ω
5.0 mΩLaboratory PremisesActive
  • 10 Ω
7.0 mΩLaboratory PremisesActive
  • 100 Ω
11 mΩLaboratory PremisesActive
  • 1 kΩ
0.10 ΩLaboratory PremisesActive
  • 10 kΩ
0.90 ΩLaboratory PremisesActive
  • 100 kΩ
9.0 ΩLaboratory PremisesActive
  • 1 MΩ
0.10 kΩLaboratory PremisesActive
  • 10 MΩ
4.5 kΩLaboratory PremisesActive
  • 100 MΩ
0.60 MΩLaboratory PremisesActive
  • Resistance Calibration of Meters 4-wire
  • 0.1 Ω
0.10 mΩLaboratory PremisesActive
  • 10 Ω
6.0 mΩLaboratory PremisesActive
  • 100 Ω
12 mΩLaboratory PremisesActive
  • 1 kΩ
0.10 ΩLaboratory PremisesActive
  • 10 kΩ
0.90 ΩLaboratory PremisesActive
  • 100 kΩ
9.0 ΩLaboratory PremisesActive
  • Capacitance Calibration of Meters
  • 1 nF
6.0 pFLaboratory PremisesActive
  • 10 nF
40 pFLaboratory PremisesActive
  • 20 nF
0.10 nFLaboratory PremisesActive
  • 50 nF
0.20 nFLaboratory PremisesActive
  • 100 nF
0.30 nFLaboratory PremisesActive
  • 1 uF
10 nFLaboratory PremisesActive
  • 10 uF
0.10 µFLaboratory PremisesActive
  • 100 uF
0.72 µFLaboratory PremisesActive
  • 1 mF
0.01 mFLaboratory PremisesActive
  • Frequency Calibration of Meters
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15 f= Measured Value
  • Up to 100 Hz
2.0 X 10-6f+ 18 µHzLaboratory PremisesActive
  • >100 Hz to 1 k Hz
2.0 X 10-6f+ 90 µHzLaboratory PremisesActive
  • >1 kHz to 10 kHz
2.0 X 10-6f+ 0.90 mHzLaboratory PremisesActive
  • >10 kHz to 20 kHz
2.0 X 10-6f+ 0.60 mHzLaboratory PremisesActive
  • >20 kHz to 50 kHz
2.0 X 10-6f + 1.7 mHzLaboratory PremisesActive
  • >50 kHz to 100 kHz
2.0 X 10-6f+ 30 mHzLaboratory PremisesActive
  • >100 kHz to 1 MHz
2.0 X 10-6f+ 91 mHzLaboratory PremisesActive
  • >1 MHz to 10 MHz
2.0 X 10-6f+ 0.90 HzLaboratory PremisesActive
  • Insulation Resistance Calibration of Meters
  • Internal Procedure: CI/01/I2 Based on SANAS guideline: TG-02-01
  • 10 kΩ
10 ΩLaboratory PremisesActive
  • 100 kΩ
0.10 kΩLaboratory PremisesActive
  • 1 MΩ
1.0 kΩLaboratory PremisesActive
  • 5 MΩ
10 kΩLaboratory PremisesActive
  • 10 MΩ
0.13 MΩLaboratory PremisesActive
  • 50 MΩ
0.59 MΩLaboratory PremisesActive
  • 100 MΩ
1.2 MΩLaboratory PremisesActive
  • 400 MΩ
5.6 MΩLaboratory PremisesActive
  • 1 GΩ
10 MΩLaboratory PremisesActive
  • 2 GΩ
50 MΩLaboratory PremisesActive
  • 4 GΩ
0.24 GΩLaboratory PremisesActive
  • 10 GΩ
0.58 GΩLaboratory PremisesActive
  • 100 GΩ
5.7 GΩLaboratory PremisesActive
  • RCD Current
  • Internal Procedure: CI/01/RCDT1 Direct Method
  • 10 mA @ 50 Hz to 30 mA @ 50 Hz
0.022Laboratory PremisesActive
  • >30 mA @ 50 Hz to 2000 mA @ 50 Hz
0.018Laboratory PremisesActive
  • Residual Current Detector (RCD) Trip Time
  • 20 ms
0.69 msLaboratory PremisesActive
  • 40 ms
0.69 msLaboratory PremisesActive
  • 200 ms
0.69 msLaboratory PremisesActive
  • 390 ms
8.0 msLaboratory PremisesActive
  • 900 ms
8.0 msLaboratory PremisesActive
  • DC Voltage-Measuring Calibration of Calibrators
  • Internal Procedure: CI/01/EC1 Based on Euramet cg-15 U=Measured value
  • Up V to 0.2 V
21 X 10-6U +2.6 µVLaboratory PremisesActive
  • >0.2 V to 2 V
10 X 10-6U + 6.5 µVLaboratory PremisesActive
  • >2 V to 20 V
10 X 10-6U + 38 µVLaboratory PremisesActive
  • >20 V to 200 V
25 X 10-6U + 0.74 µVLaboratory PremisesActive
  • >200 V to 1000 V
25 X 10-6U + 2.6 mVLaboratory PremisesActive
  • Internal Procedure: CI/01/I2 Based on SANAS guideline: TG-02-01 U=measured value
  • >1000 V to 10000 V
0.24 X 10-3ULaboratory PremisesActive
  • AC Voltage-Measuring Calibration of Calibrators
  • Internal Procedure: CI/01/EC1 Based on Euramet cg-15 U=Measured value
  • (Up V to 0.2 V) 20 Hz to 50 Hz
2.9 X 10-3U + 4.8 µVLaboratory PremisesActive
  • (Up V to 0.2 V) >50 Hz to 100 Hz
0.76 X 10-3U + 11 µVLaboratory PremisesActive
  • (Up V to 0.2 V) >100 Hz to 2 kHz
0.16 X 10-3U + 18 µVLaboratory PremisesActive
  • (Up V to 0.2 V) >2 kHz to 10 kHz
0.16 X 10-3U + 18 µVLaboratory PremisesActive
  • (Up V to 0.2 V) >10 kHz to 30 kHz
0.16 X 10-3U + 38 µVLaboratory PremisesActive
  • (Up V to 0.2 V) >30 kHz to 100 kHz
3.2 X 10-3U + 55 µVLaboratory PremisesActive
  • (>0.2 V to 2 V) 20 Hz to 50 Hz
2.9 X 10-3U + 27 µVLaboratory PremisesActive
  • (>0.2 V to 2 V) >50 Hz to 100 Hz
0.79 X 10-3U + 23 µVLaboratory PremisesActive
  • (>0.2 V to 2 V) >100 Hz to 2 kHz
0.21 X 10-3U + 54 µVLaboratory PremisesActive
  • (>0.2 V to 2 V) >2 kHz to 10 kHz
0.21 X 10-3U + 54 µVLaboratory PremisesActive
  • (>0.2 V to 2 V) >10 kHz to 30 kHz
0.25 X 10-3U + 0.10 mVLaboratory PremisesActive
  • (>0.2 V to 2 V) >30 kHz to 100 kHz
3.5 X 10-3U + 41 µVLaboratory PremisesActive
  • (>2 V to 20 V) 20 Hz to 50 Hz
2.9 X 10-3U + 0.27 mVLaboratory PremisesActive
  • (>2 V to 20 V) >50 Hz to 100 Hz
0.80 X 10-3U + 0.21 mVLaboratory PremisesActive
  • (>2 V to 20 V) >100 Hz to 2 kHz
0.33 X 10-3U + 0.41 mVLaboratory PremisesActive
  • (>2 V to 20 V) >2 kHz to 10 kHz
0.45 X 10-3U + 0.33 mVLaboratory PremisesActive
  • (>2 V to 20 V) >10 kHz to 30 kHz
0.55 X 10-3U + 0.72 mVLaboratory PremisesActive
  • (>2 V to 20 V) >30 kHz to 100 kHz
3.5 X 10-3U + 0.31 mVLaboratory PremisesActive
  • (>20 V to 200 V) 20 Hz to 50 Hz
2.9 X 10-3U + 2.7 mVLaboratory PremisesActive
  • (>20 V to 200 V) >50 Hz to 100 Hz
0.80 X 10-3U + 2.9 mVLaboratory PremisesActive
  • (>20 V to 200 V) >100 Hz to 2 kHz
0.56 X 10-3U + 3.7 mVLaboratory PremisesActive
  • (>20 V to 200 V) >2 kHz to 10 kHz
0.45 X 10-3U + 4.4 mVLaboratory PremisesActive
  • (>20 V to 200 V) >10 kHz to 30 kHz
0.55 X 10-3U + 8.1 mVLaboratory PremisesActive
  • (>200 V to 750 V) 20 Hz to 50 Hz
2.9 X 10-3U + 4.5 mVLaboratory PremisesActive
  • (>200 V to 750 V) >50 Hz to 100 Hz
1.1 X 10-3U + 11 mVLaboratory PremisesActive
  • (>200 V to 750 V) >100 Hz to 2 kHz
0.53 X 10-3U + 47 mVLaboratory PremisesActive
  • (>200 V to 750 V) >2 kHz to 10 kHz
0.51 X 10-3U + 0.21 VLaboratory PremisesActive
  • CI/01/I2, Based on SANAS guideline:TG-02-01 U= measured value
  • >1000 V up to 10000 V
0.55 ž 10-3 ULaboratory PremisesActive
  • DC Current-Measuring Calibration of Calibrators
  • Internal Procedure: CI/01/EC1 Based on Euramet cg-15 U=Measured value
  • Up to 200 µA
0.40 X 10-3U + 0.04 µALaboratory PremisesActive
  • >0.2 mA to 2 mA
0.40 X 10-3U + 0.05 µALaboratory PremisesActive
  • >2 mA to 20 mA
0.40 X 10-3U + 0.43 µALaboratory PremisesActive
  • >20 mA to 200 mA
0.43 X 10-3U + 4.4 µALaboratory PremisesActive
  • >0.2 A to 2 A
0.86 X 10-3U + 63 µALaboratory PremisesActive
  • AC Current-Measuring Calibration of Calibrators
  • (Up to 200 µA ) 20 Hz to 50 Hz
4.0 X 10-3U + 4.6 nALaboratory PremisesActive
  • (Up to 200 µA ) >50 Hz to 200 Hz
2.3 X 10-3U + 7.8 nALaboratory PremisesActive
  • (Up to 200 µA ) >200 Hz to 1 kHz
4.6 X 10-3U + 4.1 nALaboratory PremisesActive
  • (Up to 200 µA ) >1 kHz to 10 kHz
5.2 X 10-3U + 0.15 µALaboratory PremisesActive
  • (>0.2 mA to 2 mA ) 20 Hz to 50 Hz
3.5 X 10-3U + 44 nALaboratory PremisesActive
  • (>0.2 mA to 2 mA ) >50 Hz to 200 Hz
1.7 X 10-3U + 82 nALaboratory PremisesActive
  • (>0.2 mA to 2 mA ) >200 Hz to 1 kHz
1.3 X 10-3U + 98 nALaboratory PremisesActive
  • (>0.2 mA to 2 mA ) >1 kHz to 10 kHz
1.4 X 10-3U + 60 µALaboratory PremisesActive
  • (>2 mA to 20 mA) 20 Hz to 50 Hz
3.5 X 10-3U + 0.44 µALaboratory PremisesActive
  • (>2 mA to 20 mA) >50 Hz to 200 Hz
1.7 X 10-3U + 0.82 µALaboratory PremisesActive
  • (>2 mA to 20 mA) >200 Hz to 1 kHz
1.3 X 10-3U + 0.98 µALaboratory PremisesActive
  • (>2 mA to 20 mA) >1 kHz to 10 kHz
1.4 X 10-3U + 0.53 µALaboratory PremisesActive
  • (>20 mA to 200 mA) 20 Hz to 50 Hz
3.5 X 10-3U + 4.1 µALaboratory PremisesActive
  • (>20 mA to 200 mA) >50 Hz to 200 Hz
1.7 X 10-3U + 7.6 µALaboratory PremisesActive
  • (>20 mA to 200 mA) >200 Hz to 1 kHz
1.4 X 10-3U + 9.1 µALaboratory PremisesActive
  • (>20 mA to 200 mA) >1 kHz to 10 kHz
1.4 X 10-3U + 2.6 µALaboratory PremisesActive
  • (>0.2 A to 2 A) 20 Hz to 50 Hz
4.0 X 10-3U + 0.15 mALaboratory PremisesActive
  • (>0.2 A to 2 A) >50 Hz to 200 Hz
2.2 X 10-3U + 0.23 mALaboratory PremisesActive
  • (>0.2 A to 2 A) >200 Hz to 1 kHz
3.4 X 10-3U + 0.17 mALaboratory PremisesActive
  • (>0.2 A to 2 A) >1 kHz to 10 kHz
5.2 X 10-3U + 8.9 µALaboratory PremisesActive
  • Resistance-Measuring Calibration of Calibrators
  • 0.2 Ω to 20 Ω
16 X 10-6U + 0.25 mΩLaboratory PremisesActive
  • >20 Ω to 200 Ω
19 X 10-6U + 1.2 mΩLaboratory PremisesActive
  • >200 Ω to 2 kΩ
8.4 X 10-6U + 6.4 mΩLaboratory PremisesActive
  • >2 kΩ to 20 kΩ
8.4 X 10-6U + 64 mΩLaboratory PremisesActive
  • >20 kΩ to 200 kΩ
39 X 10-6U + 0.60 ΩLaboratory PremisesActive
  • >200 kΩ to 2 MΩ
72 X 10-6U + 30 ΩLaboratory PremisesActive
  • >2 MΩ to 20 MΩ
0.29 X 10-3U + 98 ΩLaboratory PremisesActive
  • >20 MΩ to 200 MΩ
0.63 X 10-3U + 4.0 kΩLaboratory PremisesActive
  • 1 GΩ
2.35 X 10-3U + 41 kΩLaboratory PremisesActive
  • Frequency-Measuring Calibration of Calibrators
  • Internal Procedure: CI/01/EC1 f= Measured Value
  • 1 Hz
0.35 X 10-3 f + 10 nHzLaboratory PremisesActive
  • >1 Hz to 10 Hz
0.35 X 10-3 f + 0.16 µHzLaboratory PremisesActive
  • >10 Hz to 100 Hz
0.35 X 10-3 f + 1.6 µHzLaboratory PremisesActive
  • >100 Hz to 1 kHz
0.35 X 10-3 f + 19 µHzLaboratory PremisesActive
  • >1 kHz to 10 kHz
0.35 X 10-3 f + 0.16 mHzLaboratory PremisesActive
  • >10 kHz to 100 kHz
0.35 X 10-3 f + 1.6 mHzLaboratory PremisesActive
  • >100 kHz to 1 MHz
0.35 X 10-3 f + 20 mHzLaboratory PremisesActive
  • DC Voltage-Sourcing Calibration of Meters
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15 U=Measured value
  • Up to 0.2 V
57 X 10-6U + 5.0 µVCustomer PremisesActive
  • >0.2 V to 2 V
57 X 10-6U + 41 µVCustomer PremisesActive
  • >2 V to 20 V
57 X 10-6U + 0.40 mVCustomer PremisesActive
  • >20 V to 200 V
57 X 10-6U + 3.5 mVCustomer PremisesActive
  • >200 V to 1025 V
57 X 10-6U + 23 mVCustomer PremisesActive
  • AC Voltage-Sourcing Calibration of Meters
  • ( Up to 0.2 V) 10 Hz to 45 Hz
0.80 X 10-3U + 61 µVCustomer PremisesActive
  • ( Up to 0.2 V) >45 Hz to 2 kHz
0.30 X 10-3U + 52 µVCustomer PremisesActive
  • ( Up to 0.2 V) >2 kHz to 19.999 kHz
1.1 X 10-3U + 0.20 mVCustomer PremisesActive
  • (>0.2 V to 2 V) 10 Hz to 45 Hz
0.70 X 10-3U + 0.40 mVCustomer PremisesActive
  • (>0.2 V to 2 V) >45 Hz to 2 kHz
0.40 X 10-3U + 0.30 mVCustomer PremisesActive
  • (>0.2 V to 2 V) >2 kHz to 20 kHz
0.70 X 10-3U + 0.60 mVCustomer PremisesActive
  • (>0.2 V to 2 V) >20 kHz to 50 kHz
2.3 X 10-3U + 3.5 mVCustomer PremisesActive
  • (>2 V to 20 V) 10 Hz to 45 Hz
0.70 X 10-3U + 3.9 mVCustomer PremisesActive
  • (>2 V to 20 V) >45 Hz to 2 kHz
0.40 X 10-3U + 3.4 mVCustomer PremisesActive
  • (>2 V to 20 V) >2 kHz to 20 kHz
0.80 X 10-3U + 5.8 mVCustomer PremisesActive
  • (>2 V to 20 V) >20 kHz to 100 kHz
3.5 X 10-3U + 35 mVCustomer PremisesActive
  • (>20 V to 200 V) 40 Hz to 2 kHz
0.50 X 10-3U + 33 mVCustomer PremisesActive
  • (>20 V to 200 V) >2 kHz to 20 kHz
1.0 X 10-3U + 50 mVCustomer PremisesActive
  • (>200 V to 1025 V) 40 Hz to 2 kHz
0.50 X 10-3U + 0.20 VCustomer PremisesActive
  • (>200 V to 1025 V) >2 kHz to 20 kHz
1.0 X 10-3U + 0.40 VCustomer PremisesActive
  • DC Current-Sourcing Calibration of Meters
  • Up to 200 µA
0.10 Xž 10-6U + 20 nACustomer PremisesActive
  • >0.2 mA to 2 mA
0.10 X 10-6U + 0.10 µACustomer PremisesActive
  • >2 mA to 20 mA
0.10 X 10-6U + 1.0 µACustomer PremisesActive
  • >20 mA to 200 mA
0.10 X 10-3U + 9.0 µACustomer PremisesActive
  • >0.2 A to 2 A
0.60 X 10-3U + 0.10 mACustomer PremisesActive
  • >2 A to 20 A
0.60 X 10-3U + 1.2 mACustomer PremisesActive
  • DC Current-Sourcing Calibration of Meters (Current coil method 10-turn & 50-turn)
  • Internal Procedure: CI/01/C6 Based on Euramet cg-15 U=Measured value
  • 20 A to 100 A
6.8 X 10-3U + 0.20 ACustomer PremisesActive
  • >100 A to 500 A
6.8 X 10-3U + 0.70 ACustomer PremisesActive
  • >500 A to 1000 A
6.8 X 10-3U + 0.70 ACustomer PremisesActive
  • AC Current-Sourcing Calibration of Meters
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15 U=Measured value
  • (Up to 200 µA ) 10 Hz to 45 Hz
1.0 X 10-3U + 0.50 µACustomer PremisesActive
  • (Up to 200 µA ) >45 Hz to 2 kHz
0.80 X 10-3U + 0.40 µACustomer PremisesActive
  • (Up to 200 µA ) >2 kHz tp 10 kHz
0.10 X 10-3U + 0.60 µACustomer PremisesActive
  • (>0.2 mA to 2 mA ) 10 Hz to 45 Hz
1.0 X 10-3U + 1.0 µACustomer PremisesActive
  • (>0.2 mA to 2 mA ) >45 Hz to 2 kHz
0.80 X 10-3U + 1.0 µACustomer PremisesActive
  • (>0.2 mA to 2 mA ) >2 kHz tp 10 kHz
0.90 X 10-3U + 2.0 µACustomer PremisesActive
  • (>2 mA to 20 mA) 10 Hz to 45 Hz
1.0 X 10-3U + 6.0 µACustomer PremisesActive
  • (>2 mA to 20 mA) >45 Hz to 2 kHz
0.80 X 10-3U + 5.0 µACustomer PremisesActive
  • (>2 mA to 20 mA) >2 kHz to 10 kHz
0.90 X 10-3U + 19 µACustomer PremisesActive
  • (>20 mA to 200 mA) 10 Hz to 45 Hz
1.0 X 10-3U + 58 µACustomer PremisesActive
  • (>20 mA to 200 mA) >45 Hz to 2 kHz
0.80 X 10-3U + 50 µACustomer PremisesActive
  • (>20 mA to 200 mA) >2 kHz to 10 kHz
0.90 X 10-3U + 0.20 mACustomer PremisesActive
  • (>0.2 A to 2 A) 10 Hz to 45 Hz
1.1 X 10-3U + 0.70 mACustomer PremisesActive
  • (>0.2 A to 2 A) >45 Hz to 2 kHz
0.70 X 10-3U + 1.8 mACustomer PremisesActive
  • (>2 A to 20 A) 10 Hz to 45 Hz
2.3 X 10-3U + 11 mACustomer PremisesActive
  • (>2 A to 20 A) >45 Hz to 200 Hz
1.7 X 10-3U + 8.2 mACustomer PremisesActive
  • (>2 A to 20 A) >200 Hz to 1 kHz
2.3 X 10-3U + 10 mACustomer PremisesActive
  • AC Current-Sourcing Calibration of Meters (Current coil method 10-turn & 50-turn)
  • Internal Procedure: CI/01/C6 Based on Euramet cg-15 U=Measured value
  • (>20 A to 100 A) 30 Hz to 60 Hz
6.8 X 10-3U + 0.20 ACustomer PremisesActive
  • (>100 A to 1000 A) 30 Hz to 60 Hz
6.8 X 10-3U + 0.70 ACustomer PremisesActive
  • Resistance-Sourcing Calibration of Meters 4-wire
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15
  • 0.2 Ω
5.0 mΩCustomer PremisesActive
  • 1 Ω
5.0 mΩCustomer PremisesActive
  • 10 Ω
7.0 mΩCustomer PremisesActive
  • 100 Ω
11 mΩCustomer PremisesActive
  • 1 kΩ
0.10 ΩCustomer PremisesActive
  • 10 kΩ
0.90 ΩCustomer PremisesActive
  • 100 kΩ
9.0 ΩCustomer PremisesActive
  • 1 MΩ
0.10 kΩCustomer PremisesActive
  • 10 MΩ
4.5 kΩCustomer PremisesActive
  • 100 MΩ
0.60 MΩCustomer PremisesActive
  • 0.1 Ω
0.10 mΩCustomer PremisesActive
  • 10 Ω
6.0 mΩCustomer PremisesActive
  • 100 Ω
12 mΩCustomer PremisesActive
  • 1 kΩ
0.10 ΩCustomer PremisesActive
  • 10 kΩ
0.90 ΩCustomer PremisesActive
  • Capacitance-Sourcing Calibration of Meters
  • 1 nF
6 pFCustomer PremisesActive
  • 10 nF
40 pFCustomer PremisesActive
  • 20 nF
0.10 nFCustomer PremisesActive
  • 50 nF
0.20 nFCustomer PremisesActive
  • 100 nF
0.30 nFCustomer PremisesActive
  • 1 uF
10 nFCustomer PremisesActive
  • Frequency-Sourcing Calibration of Meters
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15 f= Measured Value
  • Up to 100 Hz
2.0 X 10-6f + 18 µHzCustomer PremisesActive
  • >100 Hz to 1 k Hz
2.0 X 10-6f + 90 µHzCustomer PremisesActive
  • >1 kHz to 10 kHz
2.0 X 10-6f + 0.90 mHzCustomer PremisesActive
  • >10 kHz to 20 kHz
2.0 X 10-6f + 0.60 mHzCustomer PremisesActive
  • >20 kHz to 50 kHz
2.0 X 10-6f + 1.7 mHzCustomer PremisesActive
  • >50 kHz to 100 kHz
2.0 X 10-6f + 30 mHzCustomer PremisesActive
  • >100 kHz to 1 MHz
2.0 X 10-6f + 91 mHzCustomer PremisesActive
  • >1 MHz to 10 MHz
2.0 X 10-6f + 0.90 HzCustomer PremisesActive
  • Insulation Resistance-Sourcing Calibration of Meters
  • Internal Procedure: CI/01/I2 Based on SANAS guideline:TG-02-01
  • 10 MΩ
0.10 MΩCustomer PremisesActive
  • 100 MΩ
0.70 MΩCustomer PremisesActive
  • 1 GΩ
0.10 GΩCustomer PremisesActive
  • 10 GΩ
0.10 GΩCustomer PremisesActive
  • DC Voltage-Measuring Calibration of Calibrators
  • Internal Procedure: CI/01/I2 Based on SANAS guideline: TG-02-01 U=measured value
  • Up to 1000 V
0.10 X 10-3U + 24 mVCustomer PremisesActive
  • >1000 V to 5000 V
0.24 X 10-3UCustomer PremisesActive
  • AC Voltage @ 60 Hz-Measurement Calibration of Calibrators
  • 0 to 1000 V
1.0 X 10-3U + 0.20 VCustomer PremisesActive
  • >1000 V to 5000 V
0.55 X 10-3UCustomer PremisesActive
  • DC power Calibration of meters
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15 Using Multiproduct Calibrator
  • (33 mV to 1025 V) 0.33 mA to 2.999 A 10.9 µW to 2.9 kW
0.04%Laboratory Premises/ Customer premisesActive
  • (33 mV to 1025 V) 3 A to 19.999 A 99 mW to 20 kW
0.12%Laboratory Premises/ Customer premisesActive
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15 Using Multiproduct Calibrator + coil
  • 20 A to 59.999 A >20 kW to 60 kW
0.47%Laboratory Premises/ Customer premisesActive
  • 60 A to 1000 A >60 kW to 1000 kW
0.36%Laboratory Premises/ Customer premisesActive
  • AC Power Calibration of Meters @ UPF, 0.5 Lag 0.5 lead 0.8 lead; 45 Hz to 65 Hz
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15
  • (33 mV to 1025 V) 3.3 mA to 8.999 mA 0.11 mW to 2.97 mW
0.20%Laboratory Premises/Customer premisesActive
  • (33 mV to 1025 V) 2.97 mW to 8.99 W
0.23%Laboratory Premises/Customer premisesActive
  • (33 mV to 1025 V) 9 mA to 32.99 mA 0.29 mW to 11 mW
0.17%Laboratory Premises/Customer premisesActive
  • 11 mW to 32.99 W
0.17%Laboratory Premises/ Customer premisesActive
  • 33 mA to 89.99mA 1.1 mW to 29.7 mW
0.20%Laboratory Premises/ Customer premisesActive
  • 29.7 mW to 89.9 W
0.24%Laboratory Premises/ Customer premisesActive
  • 90 mA to 329.99 mA 2.97 mW to 109.9 mW
0.13%Laboratory Premises/ Customer premisesActive
  • 109.9 mW to 329.99 W
0.17%Laboratory Premises/ Customer premisesActive
  • 0.33 A to 0.8999 A 10.9 mW to 297 mW
0.18%Laboratory Premises/ Customer premisesActive
  • 297 mW to 899.9 W
0.21%Laboratory Premises/ Customer premisesActive
  • 0.9 A to 2.1999 A 2.97 mW to 725 mW
0.18%Laboratory Premises/ Customer premisesActive
  • 725 mW to 2.1 kW
0.18%Laboratory Premises/ Customer premisesActive
  • 2.2 A to 4.4999 A 72.6 mW to 1.485 W
0.20%Laboratory Premises/ Customer premisesActive
  • 1.485 W to 4.4 kW
0.22%Laboratory Premises/ Customer premisesActive
  • 4.5 A to 20 A 148.5 mW to 6.6 W
0.17%Laboratory Premises/ Customer premisesActive
  • 6.6 W to 20 kW
0.19%Laboratory Premises/ Customer premisesActive
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15 Using Multiproduct Calibrator + coil
  • 20 A to 59.999 A 20 kW to 60 kW
0.47%Laboratory Premises/ Customer premisesActive
  • 60 A to 1000 A 60 kW to 1000 kW
0.36%Laboratory Premises/ Customer premisesActive
  • Phase angle Calibration of Meters (Voltage to Current)
  • Internal Procedure: CI/01/M1 Based on OEM manual
  • 0°- 90° 10 Hz to 65 Hz
0.2°Laboratory Premises/ Customer premisesActive
  • 0°- 90° 65 Hz to 500 Hz
0.4°Laboratory Premises/ Customer premisesActive
  • 0°- 90° 0.5 kHz to 1kHz
0.8°Laboratory Premises/ Customer premisesActive
  • >90°- 360° 50 Hz
0.4°Laboratory Premises/ Customer premisesActive
  • Active Energy/Reactive Energy Calibration of Meters
  • Internal Procedure: CI/01/EM1 Based on EN 50470-3
  • 40 - 65 Hz 30-500 Volts PF= 1 to 0.5 (Lead and Lag) Current: 1 mA to 12 A
0.07%Laboratory Premises/ Customer premisesActive
  • 40 - 65 Hz 30-500 Volts PF= 1 to 0.5 (Lead and Lag) Current: 10mA to 120 A
0.08%Laboratory Premises/ Customer premisesActive
  • Internal Procedure: CI/01/EM1 Based on EN 50470-3 (Using Current Clamps)
0.24%Laboratory Premises/ Customer premisesActive
  • Resistance Calibration of meters
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15 (Using Standard Calibration Resistors/MPC)
  • 10 µΩ
1.50%Laboratory Premises/ Customer premisesActive
  • 50 μΩ
1.20%Laboratory Premises/ Customer premisesActive
  • 200 μΩ
0.07%Laboratory Premises/ Customer premisesActive
  • 500 μΩ
0.06%Laboratory Premises/ Customer premisesActive
  • 1 mΩ
0.06%Laboratory Premises/ Customer premisesActive
  • 2 mΩ
0.08%Laboratory Premises/ Customer premisesActive
  • 5 mΩ
0.06%Laboratory Premises/ Customer premisesActive
  • 50 mΩ
0.06%Laboratory Premises/ Customer premisesActive
  • 100 mΩ
0.02%Laboratory Premises/ Customer premisesActive
  • 200 mΩ
0.04%Laboratory Premises/ Customer premisesActive
  • 500 mΩ
0.02%Laboratory Premises/ Customer premisesActive
  • 1 Ω
0.02%Laboratory Premises/ Customer premisesActive
  • >1 Ω-10 Ω
0.01%Laboratory Premises/ Customer premisesActive
  • >10 Ω - 100 Ω
0.01%Laboratory Premises/ Customer premisesActive
  • >100 Ω - 1MΩ
0.00%Laboratory Premises/ Customer premisesActive
  • >1 MΩ - 3.3 MΩ
0.007%Laboratory Premises/ Customer premisesActive
  • >3.3 MΩ - 11 MΩ
0.02%Laboratory Premises/ Customer premisesActive
  • >11 MΩ - 33 MΩ
0.03%Laboratory Premises/ Customer premisesActive
  • >33 MΩ - 110 MΩ
0.06%Laboratory Premises/ Customer premisesActive
  • 110 MΩ - 330 MΩ
0.30%Laboratory Premises/ Customer premisesActive
  • >330 MΩ -1 GΩ
1.70%Laboratory Premises/ Customer premisesActive
  • Capacitance Calibration of Meters
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15 (Using Standard Capacitance/MPC)
  • 100 pF - 1nF @ 1 kHz
4.7% to 1.5%Laboratory Premises/ Customer premisesActive
  • 1 nF - 10 µF @ 1 kHz
1.2% to 1.5%Laboratory Premises/ Customer premisesActive
  • 0.19 nF to 0.3999 nF (@10 Hz to 10 kHz)
6.70%Laboratory Premises/ Customer premisesActive
  • 0.4 nF to 1.0999 nF (@10 Hz to 10 kHz)
3.90%Laboratory Premises/ Customer premisesActive
  • 1.1 to 3.2999 nF (@10 Hz to 3 kHz)
2.50%Laboratory Premises/ Customer premisesActive
  • 3.3 to 10.9999 nF (@10 Hz to 1 kHz)
0.50%Laboratory Premises/ Customer premisesActive
  • 11 to 32.9999 nF (@10 Hz to 1 kHz)
0.90%Laboratory Premises/ Customer premisesActive
  • 33 to 109.999 nF (@10 Hz to 1 kHz)
0.50%Laboratory Premises/ Customer premisesActive
  • 110 to 329.999 nF (@10 Hz to 1 kHz)
0.50%Laboratory Premises/ Customer premisesActive
  • 0.33 to 1.09999 µF (@10 Hz to 600 Hz)
0.50%Laboratory Premises/ Customer premisesActive
  • 1.1 to 3.29999 µF (@ 10 Hz to 300 Hz)
0.50%Laboratory Premises/ Customer premisesActive
  • 3.3 to 10.9999 μF (@ 10 Hz to 150 Hz)
0.50%Laboratory Premises/ Customer premisesActive
  • 11 to 32.9999 μF (@ 10 Hz to 120 Hz)
0.70%Laboratory Premises/ Customer premisesActive
  • 33 to 109.999 μF (@ 10 Hz to 80 Hz)
0.70%Laboratory Premises/ Customer premisesActive
  • 110 to 329.999 μF (@ 0 Hz to 50 Hz)
0.70%Laboratory Premises/ Customer premisesActive
  • 0.33 mF to 1.09999 mF (@ 0 Hz to 20 Hz)
0.90%Laboratory Premises/ Customer premisesActive
  • 1.1 to 3.29999 mF (@ 0 Hz to 6 Hz)
0.70%Laboratory Premises/ Customer premisesActive
  • 3.3 to 10.9999 mF (@ 0 Hz to 2 Hz)
0.70%Laboratory Premises/ Customer premisesActive
  • 11 to 32.9999 mF (@ 0 Hz to 0.6 Hz)
1.00%Laboratory Premises/ Customer premisesActive
  • 33 to 110 mF (@ 0 Hz to 0.2 Hz)
1.40%Laboratory Premises/ Customer premisesActive
  • Inductance-Measurement Calibration of Meters
  • Internal Procedure: CI/01/M1 Based on Euramet cg-15 (Using MPC)
  • 1 mH to 10 H (@ 1 kHz)
1.00%Laboratory Premises/ Customer premisesActive
  • 10 µH to 100 µH (10 kHz)
0.19%Laboratory PremisesActive
  • 100 µH to 1 H (100 Hz to 1 kHz)
0.30%Laboratory PremisesActive
  • 100 µH to 1 H (1 kHz to 10 kHz)
0.16%Laboratory PremisesActive
  • 1 H to 10 H (100 Hz to 1 kHz)
0.30%Laboratory PremisesActive
  • DC Current Calibration of Calibrators
  • Internal Procedure: CI/01/EC1 Based on Euramet cg-15 (Using Current shunt and DC Electronic load)
  • >2 A to 30 A
0.15 A to 0.18 ALaboratory PremisesActive
  • >30 A to 240 A
0.6 ALaboratory PremisesActive
  • AC Current Calibration of Calibrators
  • Internal Procedure: CI/01/EC1 Based on Euramet cg-15 (Using Current shunt)
  • > 2A to 30 A (@ 50 Hz to 1 kHz)
0.15 A to 0.18 ALaboratory PremisesActive
  • Resistance Calibration of Calibrators
  • Internal Procedure: CI/01/EC1 Based on Euramet cg-15
  • 1 Ω to 10 MΩ (10 Hz to 1 kHz)
0.15%Laboratory PremisesActive
  • Capacitance Calibration of Calibrators
  • 10 pF to 100 pF (1 kHz to 100 kHz)
0.63% to 0.15%Laboratory PremisesActive
  • 100 pF to 1 nF (100 Hz to 10 kHz)
0.60% to 0.15%Laboratory PremisesActive
  • 1 nF to 10 µF (100 Hz to 10 kHz)
0.15%Laboratory PremisesActive
  • 10 µF to 100 µF (100 Hz to 1 kHz)
0.15%Laboratory PremisesActive
  • Safety Testers/ PAT Testers Earth bond current @ 0.1Ω load
  • Internal Procedure: CI/01/SA1 Based on Transmille 3200 Manual
  • 100 mA
9 mALaboratory PremisesActive
  • 4 A - 25 A
0.5 ALaboratory PremisesActive
  • Safety Testers/ PAT Testers Earth Bond Resistance
  • 36 mΩ to 1 Ω
5 mΩ - 11 mΩLaboratory PremisesActive
  • 5 Ω
29 mΩLaboratory PremisesActive
  • 10 Ω
58 mΩLaboratory PremisesActive
  • 100 Ω
580 mΩLaboratory PremisesActive
  • 1000 Ω
5.8ΩLaboratory PremisesActive
  • Safety Testers/ PAT Testers Leakage current
  • 3.3 µA @ 240 V
0.1 µALaboratory PremisesActive
  • 33 µA @ 240 V
0.4 µALaboratory PremisesActive
  • 115 µA @ 240 V
1.3 µALaboratory PremisesActive
  • 230 µA @ 240 V
2.7 µALaboratory PremisesActive
  • 0.46 mA @ 240 V
5 µALaboratory PremisesActive
  • 2.2 mA @ 240 V
40 μALaboratory PremisesActive
  • 4.7 mA @ 240 V
50 μALaboratory PremisesActive
  • 7.7 mA @ 240 V
50 μALaboratory PremisesActive
  • Safety Testers/ PAT Testers Load for PAT-At nominal UAE mains output
  • 0.13 kVA
0.063Laboratory PremisesActive
  • Continuity Resistance
  • 0.2 Ω - 2 Ω
30 mΩLaboratory PremisesActive
  • 2Ω - 10Ω
40 mΩ - 58 mΩLaboratory PremisesActive
  • 10Ω - 20Ω
58 mΩ - 90 mΩLaboratory PremisesActive
  • 100Ω
0.32 ΩLaboratory PremisesActive
  • 1 kΩ
3 ΩLaboratory PremisesActive
  • 100 mA to 300 mA
3.7 mALaboratory PremisesActive
  • Loop Testers Loop impedence
  • 0.51 Ω
43 mΩLaboratory PremisesActive
  • 1 Ω
44 mΩLaboratory PremisesActive
  • 5 Ω
56 mΩLaboratory PremisesActive
  • 10 Ω
80 mΩLaboratory PremisesActive
  • 99 Ω
0.6 ΩLaboratory PremisesActive
  • 995 Ω
5.8 ΩLaboratory PremisesActive
  • Generation Frequency
  • Internal procedure : CI/14/SPA1 Based on OEM manual; Using Signal Generator, Sweeper
  • 100kHz to 4GHz
0.001 kHz to 1.5 kHzLaboratory PremisesActive
  • >4 GHz to 20 GHz
2.6 kHz to 13.2 kHzLaboratory PremisesActive
  • Generation Power Level (Absolute Power)
  • Internal procedure : CI/14/RFS1 Based on OEM manual; Using Sweeper
  • 50 MHz to 1 GHz -30 dBm to 20 dBm
1.4 dBLaboratory PremisesActive
  • 10 GHz to 18 GHz -30 dBm to 13 dBm
1.6 dBLaboratory PremisesActive
  • Generation Amplitude Modulation (AM) (Mod Rate 1 kHz)
  • Internal procedure : CI/14/SPA1 Based on OEM manual; Using Signal Generator
  • (AM Frequency: 10 MHz to 4 GHz) Depth: 10%
1.70%Laboratory PremisesActive
  • (AM Frequency: 10 MHz to 4 GHz) Depth: 30%
3.00%Laboratory PremisesActive
  • (AM Frequency: 10 MHz to 4 GHz) Depth: 50%
4.30%Laboratory PremisesActive
  • (AM Frequency: 10 MHz to 4 GHz) Depth: 70%
5.60%Laboratory PremisesActive
  • (AM Frequency: 10 MHz to 4 GHz) Depth: 80%
6.30%Laboratory PremisesActive
  • Generation Frequency Modulation (FM) (Mod Rate 1 kHz)
  • Internal procedure : CI/14/SPA1 Based on OEM manual; Using Signal Generator
  • (FM Frequency: 10 MHz to 1 GHz) Deviation: 1 kHz
0.04 kHzLaboratory PremisesActive
  • (FM Frequency: 10 MHz to 1 GHz) Deviation: 50 kHz
2 kHzLaboratory PremisesActive
  • (FM Frequency: 10 MHz to 1 GHz) Deviation: 100 kHz
4 kHzLaboratory PremisesActive
  • Generation Reflection Coefficient (from 0.3 MHz to 10 GHz, N connector)
  • Internal procedure : CI/14/RFS1 Based on OEM manual; Using 6 dB and 20 dB Return Loss standards
  • 0.501 & 0.1 (from 0.3 MHz to 10 GHz)
0.04 & 0.07Laboratory PremisesActive
  • Measure Frequency
  • Internal procedure : CI/14/SG1 Based on OEM manual; Using Counter
  • 10 kHz- 20 GHz
0.0015 kHz to 8.5 kHzLaboratory PremisesActive
  • Measure Power Level (Absolute Power)
  • Internal procedure :CI/14/SG1 Based on OEM manual; Using power meter with sensor and spectrum analyser
  • (10 MHz to 18 GHz) -30 dBm
0.8 dBLaboratory PremisesActive
  • (10 MHz to 18 GHz) -25 dBm to 13 dBm
0.5 dBLaboratory PremisesActive
  • (10 MHz to 6 GHz) -60 dBm to -30 dBm
1.8 dBLaboratory PremisesActive
  • (>6 GHz to 18 GHz) -60 dBm to -30 dBm
2.4 dBLaboratory PremisesActive
  • Measure AM Modulation (Mod Rate- 1kHz)
  • Internal procedure : CI/14/SG1 Based on OEM manual; Using Measuring Receiver
  • (10 MHz to 1GHz) Depth: 1%
0.08%Laboratory PremisesActive
  • (10 MHz to 1GHz) Depth: 10%
0.40%Laboratory PremisesActive
  • (10 MHz to 1GHz) Depth: 50%
1.00%Laboratory PremisesActive
  • (10 MHz to 1GHz) Depth: 90%
1.50%Laboratory PremisesActive
  • (1 MHz to 10 MHz) Deviation:1 kHz to 40 kHz
0.03 kHz to 0.94 kHzLaboratory PremisesActive
  • (1 MHz to 10 MHz) Deviation: >40 kHz to 100 kHz
0.52 kHz to 1.3 kHzLaboratory PremisesActive
  • (>0.01 GHz to 1 GHz) Deviation:1 kHz to 100 kHz
0.15 kHz to 1.4 kHzLaboratory PremisesActive
  • Measure Reflection Coefficient 0.2 to 0.501 (2 MHz to 4 GHz)
  • Internal procedure : CI/14/RFM1 Based on OEM manual; Using RF Analyzer
  • 2 MHz to 3.5 GHz
0.16Laboratory PremisesActive
  • > 3.5 GHz to 4 GHz
0.36Laboratory PremisesActive
  • Insulation Resistance
  • CI/01/I2 Based on SANAS TG-01-02
  • >100 GΩ to 10 TΩ
5.00%Laboratory PremisesActive
  • Frequency (Meassure)
  • CI/01/EC1 Based on EURAMET cg-15
  • > 1 MHz to 15 MHz
0.04%Laboratory PremisesActive
Pressure
  • Gas Pressure (Gauge Pressure)
  • Digital Indicator CI/02/PI2 BS 837-1 EURAMET /cg-17/v.02 DKD-R 6-1
  • ­100 kPa to 100 kPa
30 PaLaboratory PremisesActive
  • 0 MPa up to 2 MPa
0.3 kPaLaboratory PremisesActive
  • > 2 MPa up to 21 MPa
2.5 kPaLaboratory PremisesActive
  • Hydraulic Pressure (gauge pressure)
  • Digital Indicator CI/02/PI2 BS 837-1 EURAMET/ cg-17/v.02 DKD-R 6-1
  • 0 MPa to 70 MPa
30 kPaCustomer premisesActive
  • 0 MPa to 140 MPa
50 kPaLaboratory PremisesActive
  • 140 MPa to 280 MPa
± 0.06 MPaLaboratory PremisesActive
  • Gas Pressure (Gauge Pressure)
  • ­100 kPa to 250 kPa
100 PaCustomer premisesActive
  • >250 kPa to 20 MPa
3 kPaCustomer premisesActive
  • Absolute Pressure
  • CI/02/PI2 Based on DKD R6-1
  • 0 mbar to 3600 mbar abs
0.04%Laboratory PremisesActive
Force
  • UTM, CBD machines
  • ISO 7500-1:2018 - Verification and calibration of ForceMeasuring System, and SOP CI/09/TTM1
  • 0.56 N to 245 kN compression
0.50%Customer PremisesActive
  • 0.56 N to 245 kN tension
0.30%Customer PremisesActive
  • Push pull Gauge
  • VDI/VDE 2624 part 2.1 - Measurement of mechanical quantities - Instructions for calibration of mobile force measurement systems, and SOP CI/04/FG1
  • 0.56 N to 4.5 kN compression
0.20%Laboratory PremisesActive
  • 0.56 N to 4.5 kN Tension
0.20%Laboratory PremisesActive
  • Hand Torque Tools (CW&CCW)
  • ISO 6789-1:2017 - Assembly tools for screws and nuts — Hand torque tools Part 1: Requirements and methods for design conformance testing and quality conformance testing: minimum requirements for declaration of conformance, and ISO 6789-2 Assembly tools fo
  • 0.565 N·m to 2711 N·m
1.00%Laboratory PremisesActive
  • 0.04 N·m to 0.8 N·m
1.00%Laboratory PremisesActive
Mass and Balance
  • Non-automatic Weighing Instruments
  • EURAMET cg 18 full scale of the unit under calibration.
  • 0 to 80 g
4 x 10 -6 FS of UUCCustomer PremisesActive
  • > 80 g to 20 kg
2 x 10·5 FS of UUCCustomer PremisesActive
  • > 20 kg to 100 kg
3.7 x 10-4 FS of UUCCustomer PremisesActive
  • > 100 kg to 1000 kg
4 x 10-4 FS of UUCCustomer PremisesActive
  • Conventional Mass
  • Substitution method with Air Buoyancy Correction
  • 1mg
0.009 mgLaboratory PremisesActive
  • 2 mg
0.012 mgLaboratory PremisesActive
  • 5 mg
0.009 mgLaboratory PremisesActive
  • 10 mg
0.013 mgLaboratory PremisesActive
  • 20 mg
0.009 mgLaboratory PremisesActive
  • 50 mg
0.010 mgLaboratory PremisesActive
  • 100 mg
0.02 mgLaboratory PremisesActive
  • 200 mg
0.011 mgLaboratory PremisesActive
  • 500 mg
0.013 mgLaboratory PremisesActive
  • l g
0.014 mgLaboratory PremisesActive
  • 2 g
0.020 mgLaboratory PremisesActive
  • S g
0.052mgLaboratory PremisesActive
  • 10 g
0.023 mgLaboratory PremisesActive
  • 20 g
0.031 mgLaboratory PremisesActive
  • 50 g
0.044mgLaboratory PremisesActive
  • 100 g
0.13 mgLaboratory PremisesActive
  • 200 g
0.18 mgLaboratory PremisesActive
  • 500 g
1.2 mgLaboratory PremisesActive
  • 1kg
1.1 mgLaboratory PremisesActive
  • 2 kg
2.1 mgLaboratory PremisesActive
  • 5 kg
100 mgLaboratory PremisesActive
  • 10 kg
100 mgLaboratory PremisesActive
  • 20 kg
100 mgLaboratory PremisesActive
  • CI/04/W2 based on OIML R-111 Subsitution Method (OIML Class E2 Class as reference weight)
  • 5 kg
10 mgLaboratory premisesActive
  • 10 kg
16 mgLaboratory premisesActive
  • 20 kg
30 mgLaboratory premisesActive
Volume
  • Micro Pipette
  • Direct (Radiometric Photometry) ISO 8655-7
  • 0.1 μl to 0.5 μl
1.5 %Laboratory PremisesActive
  • > 0.5 μl to 5000 μl
1.2 %Laboratory PremisesActive
  • Piston operated Volumetric Devices (pipettes, dispensers, syringes, burettes)
  • Gravimetric method ISO 8655-6 V=measured valu
  • 10 µl to 500 µI
0.82 µILaboratory PremisesActive
  • > 500 µl to 10000 µI
0.2%*VLaboratory PremisesActive
  • Volumetric Instrument (Glassware) Volumetric Flask
  • Gravimetric method ISO 4787
  • 5 ml
0.04 mlLaboratory PremisesActive
  • 10 ml
0.04 mlLaboratory PremisesActive
  • 20 ml
0.04 mlLaboratory PremisesActive
  • 25 ml
0.04 mlLaboratory PremisesActive
  • 50 ml
0.07 mlLaboratory PremisesActive
  • 100 ml
0.09 mlLaboratory PremisesActive
  • 200 ml
0.1.3 mlLaboratory PremisesActive
  • 250 ml
0.13 mlLaboratory PremisesActive
  • 500 ml
0.20 mlLaboratory PremisesActive
  • 1000 ml
0.29 mlLaboratory PremisesActive
  • 2000 ml
0.41 mlLaboratory PremisesActive
  • Measuring Cylinder
  • 1ml to 5 ml
0.04 mlLaboratory PremisesActive
  • 10 ml
0.08 mlLaboratory PremisesActive
  • 25 ml
0.20 mlLaboratory PremisesActive
  • 50 ml
0.41 mlLaboratory PremisesActive
  • 100 ml
0.41mlLaboratory PremisesActive
  • 250 ml
0.82 mlLaboratory PremisesActive
  • 500 ml
2.0 mlLaboratory PremisesActive
  • 1000 ml
4.1mlLaboratory PremisesActive
  • 2000 ml
8.2 mlLaboratory PremisesActive
  • Measuring Pipette
  • 1 ml
0.004 mlLaboratory PremisesActive
  • 2 ml
0.008 mlLaboratory PremisesActive
  • 5 ml
0.02 mlLaboratory PremisesActive
  • 10 ml
0.04 mlLaboratory PremisesActive
  • 20 ml
0.04 mlLaboratory PremisesActive
  • 25 ml
0.04 mlLaboratory PremisesActive
  • 50 ml
0.04 mlLaboratory PremisesActive
  • 100 ml
0.20 mlLaboratory PremisesActive
  • Burette
  • Gravimetric method ISO 4787
  • 1 ml to 10 ml
0.009 mlLaboratory PremisesActive
  • 25 ml
0.021 mlLaboratory PremisesActive
  • 50 ml
0.033 mlLaboratory PremisesActive
  • 100 ml
0.061 mlLaboratory PremisesActive
  • Bulb pipette
  • 0.5 ml
0.0004 mlLaboratory PremisesActive
  • 1 ml
0.005 mlLaboratory PremisesActive
  • 2 ml
0.005 mlLaboratory PremisesActive
  • 5 ml
0.01 mlLaboratory PremisesActive
  • 10 ml
0.01 mlLaboratory PremisesActive
  • 20 ml
0.01 mlLaboratory PremisesActive
  • 25 ml
0.01mlLaboratory PremisesActive
  • 50 ml
0.01 mlLaboratory PremisesActive
  • 100 ml
0.01 mlLaboratory PremisesActive
  • Beaker
  • Gravimetric method 1S0 4787
  • 1ml to 5 ml
0.02 mlLaboratory PremisesActive
  • 10 ml
0.04 mlLaboratory PremisesActive
  • 25 ml
0.21mlLaboratory PremisesActive
  • 50 ml
0.32 mlLaboratory PremisesActive
  • 100 ml
0.45 mlLaboratory PremisesActive
  • 150 ml
0.65 m lLaboratory PremisesActive
  • 250 ml
0.89 mlLaboratory PremisesActive
  • 400 ml
1.2 mlLaboratory PremisesActive
  • 500 ml
1.5 mlLaboratory PremisesActive
  • 600 ml
1.8 mlLaboratory PremisesActive
  • 800 ml
2.0 mlLaboratory PremisesActive
  • 1000 ml
3.2 mlLaboratory PremisesActive
  • 2000 ml
3.2 mlLaboratory PremisesActive
Dimensions
  • All types of Caliper Digital / Dial / Vernier
  • Procedure (CI/05/V3) based on ISO 13385-1-2019 For determining error of indicated size Comparison with gauge block
  • 0 mm to 300 mm
25 μmLaboratory PremisesActive
  • 300 mm to 600 mm
28 μmLaboratory PremisesActive
  • 600 mm to 1000 mm
35 μmLaboratory PremisesActive
  • External Micrometer
  • Procedure (CI/05/O2) based on ISO 3611 :2010 For determining error of indicated size Comparison with gauge blocks
  • 0 mm to 50 mm
3 μmLaboratory PremisesActive
  • 50 mm to 100 mm
3.5 μmLaboratory PremisesActive
  • Dial gauges (Digital & Analogue)
  • Procedure (CI/05/D6) based on JIS 7503:2017 For determining error of indicated displacement Comparison with dial gauge tester/ gauge blocks
  • 0 mm to 50 mm
3 μmLaboratory PremisesActive
  • Gauge blocks (metric) (Steel, Tungsten Carbide and Ceramic)
  • Procedure-CI/05/GB1 based on ISO 3650 Measurement of central length by mechanical comparison.
  • 0.5 mm to 10 mm
0.12 µmLaboratory PremisesActive
  • >10 mm to 25 mm
0.15 µmLaboratory PremisesActive
  • >25 mm to 50 mm
0.20 µmLaboratory PremisesActive
  • >50 mm to 75 mm
0.25 µmLaboratory PremisesActive
  • >75 mm to 100 mm
0.35 µmLaboratory PremisesActive
Time and Frequency
  • Time Interval
  • Internal Procedure CI/05/SW1 Based on NIST 960-12
  • Up to 86400 s
0.3 sLaboratoryActive
  • Tachometer/ Stroboscope
  • Internal Procedure: CI/01/TCM1 Based on SANAS TR-45-02
  • 30 rpm to 60 rpm
0.00015Laboratory/ Customer PremisesActive
  • >60 rpm to 100 rpm
8.0E-5Laboratory/ Customer PremisesActive
  • >100 rpm to 200,000 rpm
6.0E-5Laboratory/ Customer PremisesActive
  • Rotational Speed Measurement
  • Internal Procedure:CI/05/CFG1 Based on SANAS TR-45-02 (Using Tachometer)
  • 30 rpm to 600 rpm
0.6 rpmLaboratory/ Customer PremisesActive
  • >600 rpm to 6000 rpm
1.8 rpmLaboratory/ Customer PremisesActive
  • >6000 rpm to 30000 rpm
0.0005Laboratory/ Customer PremisesActive
  • Oscilloscope Amplitude-Square wave in to 50 Ω
  • Internal Procedure: CI/01/OSE1 Based on Euramet cg-7
  • 1.8 mVp-p to 200 mVp-p
0.25 % of reading + 0.15 mVLaboratory/ Customer PremisesActive
  • >0.2 Vp-p to 2.2 Vp-p
0.30 % of reading + 0.1 mVLaboratory/ Customer PremisesActive
  • Oscilloscope Amplitude-Square wave) in to 1MΩ
  • 1.8 mVp-p to 105 Vp-p
0.30 % of reading + 0.1 mVLaboratory/ Customer PremisesActive
  • Oscilloscope Time Marker
  • 2 ns
0.03%Laboratory/ Customer PremisesActive
  • 5 ns
0.01%Laboratory/ Customer PremisesActive
  • 10 ns
0.01%Laboratory/ Customer PremisesActive
  • 20 ns
0.00%Laboratory/ Customer PremisesActive
  • 50 ns to 500 ns
0.00%Laboratory/ Customer PremisesActive
  • 1 µs
0.06%Laboratory/ Customer PremisesActive
  • 2 µs
0.03%Laboratory/ Customer PremisesActive
  • 5 µs
0.01%Laboratory/ Customer PremisesActive
  • 10 µs
0.01%Laboratory/ Customer PremisesActive
  • 20 µs
0.00%Laboratory/ Customer PremisesActive
  • 50 µs to 500 µs
0.00%Laboratory/ Customer PremisesActive
  • 1 ms
0.06%Laboratory/ Customer PremisesActive
  • 2 ms
0.03%Laboratory/ Customer PremisesActive
  • 5 ms
0.01%Laboratory/ Customer PremisesActive
  • 10 ms
0.01%Laboratory/ Customer PremisesActive
  • 20 ms
0.00%Laboratory/ Customer PremisesActive
  • 50 ms to 500 ms
0.00%Laboratory/ Customer PremisesActive
  • 1 s
0.06%Laboratory/ Customer PremisesActive
  • 2 s
0.03%Laboratory/ Customer PremisesActive
  • 5 s
0.01%Laboratory/ Customer PremisesActive
  • Oscilloscope Bandwidth
  • 50kHz to 100 MHz
1.70%Laboratory/ Customer PremisesActive
  • >100 MHz to 300 MHz
2.30%Laboratory/ Customer PremisesActive
  • >300 MHz to 4.32GHz
9 kHzLaboratory/ Customer PremisesActive