Aragon Calibrations Laboratory

Initial Accreditation Date:Aug 06 , 2014

معلومات الاتصال

العنوان

G-03, UAE Enterprises Bldg, Airport Road, Dubai, United Arab Emirates, Dubai, United Arab Emirates

صندوق بريد / الرمز البريدي

103912

البريد الإلكتروني

technical@aragoncalibrations.com

الموقع الإلكتروني

www.aragoncalibrations.com

رقم الهاتف

+971 04 885 1098

رقم الجوال

+971 04 885 1097

المجال الرئيسي

Calibration Laboratories

ساري منذ:Aug 23 , 2023

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Balance
  • Weighing Scale
  • EURAMET/cg-18/v.04
  • 1mg to 20g
0.12 mgCustomer PremisesActive
  • > 20 to 200g
0.78 mgCustomer PremisesActive
  • > 200 to 500g
1.8 mgCustomer PremisesActive
  • > 500 to 1000g
2.6 mgCustomer PremisesActive
  • > 1 to 10 kg
15 mgCustomer PremisesActive
  • 0 to 60 kg
8.2 gCustomer PremisesActive
  • > 60 to 120 kg
8.3 gCustomer PremisesActive
  • > 120 to 300 kg
0.70 kgCustomer PremisesActive
  • Mass/ Electronic Balance
  • Comparison method CCM-02/rev.04/01-04-22 in acc. with EURAMET cg 18.v4 (2015)
  • >300 kg to 1000 kg
0.70 kgLaboratory/ Customer PremisesActive
Mass
  • Mass Calibration
  • OIML R111-1
  • 0.001g
0.041 mgLaboratory PremisesActive
  • 0.002g
0.041 mgLaboratory PremisesActive
  • 0.005g
0.041 mgLaboratory PremisesActive
  • 0.01g
0.042 mgLaboratory PremisesActive
  • 0.02g
0.042 mgLaboratory PremisesActive
  • 0.05g
0.043 mgLaboratory PremisesActive
  • 0.1g
0.044 mgLaboratory PremisesActive
  • 0.2g
0.046 mgLaboratory PremisesActive
  • 0.5g
0.048 mgLaboratory PremisesActive
  • 1g
0.050 mgLaboratory PremisesActive
  • 2g
0.061 mgLaboratory PremisesActive
  • 5g
0.065 mgLaboratory PremisesActive
  • 10g
0.085 mgLaboratory PremisesActive
  • 20g
0.09 mgLaboratory PremisesActive
  • 50g
0.12 mgLaboratory PremisesActive
  • 100g
0.23 mgLaboratory PremisesActive
  • 200g
0.64 mgLaboratory PremisesActive
  • 500g
3 mgLaboratory PremisesActive
  • 1kg
7 mgLaboratory PremisesActive
  • Mass/ Weights
  • Comparison method CCM-01/rev.03/01-02-22 in acc. with OIML R 111-1:2004
  • 2 kg
14 mgLaboratory PremisesActive
  • 5 kg
23 mgLaboratory PremisesActive
  • 10 kg
45 mgLaboratory PremisesActive
  • 20 kg
0.11 gLaboratory PremisesActive
Electrical
  • DC Voltage Calibration of meters
  • Direct Method Using Multi-Product Calibrator U = Measured Voltage Value
  • 0.001 mV to 200 mV
31x10-6 U + 4.9 µVLaboratory / Customer PremisesActive
  • >200 mV to 220 mV
31x10-6 U + 5.0 µVLaboratory / Customer PremisesActive
  • >220 mV to 1 V
31x10-6 U + 5.6 µVLaboratory / Customer PremisesActive
  • >1 V to 2 V
31x10-6 U + 10 µVLaboratory / Customer PremisesActive
  • >2 V to 2.2 V
25x10-6 U + 46 µVLaboratory / Customer PremisesActive
  • >2.2 V to 10 V
25x10-6 U + 58 µVLaboratory / Customer PremisesActive
  • >10 V to 20 V
25x10-6 U + 0.10 mVLaboratory / Customer PremisesActive
  • >20 V to 100 V
31x10-6 U + 0.77 mVLaboratory / Customer PremisesActive
  • >100 V to 200 V
31x10-6 U + 1.5 mVLaboratory / Customer PremisesActive
  • >200 V to 220 V
31x10-6 U + 4.8 mVLaboratory / Customer PremisesActive
  • >220 V to 1000 V
31x10-6 U + 8.2 mVLaboratory / Customer PremisesActive
  • AC Voltage Calibration of meters
  • 0.001 mV to 20 mV @40 Hz to 206 Hz
0.35x10-3 U + 25 µVLaboratory / Customer PremisesActive
  • <20 mV to 200 mV @40 Hz to 206 Hz
0.35x10-3 U + 48 µVLaboratory / Customer PremisesActive
  • <200 mV to 2 V @40 Hz to 206 Hz
0.35x10-3 U + 0.41 mVLaboratory / Customer PremisesActive
  • <2 V to 20 V @40 Hz to 206 Hz
0.32x10-3 U + 6.9 mVLaboratory / Customer PremisesActive
  • <20 V to 200 V @40 Hz to 206 Hz
0.35x10-3 U + 0.16 VLaboratory / Customer PremisesActive
  • <200 V to 1000 V @40 Hz to 206 Hz
0.35x10-3 U + 0.50 VLaboratory / Customer PremisesActive
  • 20 mV to 200 mV @1 kHz to 20 kHz
0.81x10-3 U + 69 µVLaboratory / Customer PremisesActive
  • <200 mV to 2 V @1 kHz to 20 kHz
0.69x10-3 U + 0.58 mVLaboratory / Customer PremisesActive
  • <2 V to 20 V@ 1 kHz to 20 kHz
0.58x10-3 U + 7.1 mVLaboratory / Customer PremisesActive
  • <20 V to 200 V @ 1 kHz to 20 kHz
0.81x10-3 U + 0.17 VLaboratory / Customer PremisesActive
  • <200 V to 1000 V @ 1 kHz to 10 kHz
1.4x10-3 U + 0.68 VLaboratory / Customer PremisesActive
  • DC Current Calibration of meters
  • Direct Method Using Multi-Product I : Measured current value
  • Up to 100 µA
92x10-6 I + 0.04 µALaboratory / Customer PremisesActive
  • >100 µA to 200 µA
92x10-6 I + 0.04 µALaboratory / Customer PremisesActive
  • >200 µA to 1 mA
81x10-6 I + 0.05 µALaboratory / Customer PremisesActive
  • >1 mA to 2 mA
81x10-6 I + 0.08 µALaboratory / Customer PremisesActive
  • >2 mA to 10 mA
46x10-6 I + 0.36 µALaboratory / Customer PremisesActive
  • >10 mA to 20 mA
46x10-6 I + 0.7 µALaboratory / Customer PremisesActive
  • >20 mA to 100 mA
58x10-6 I + 3.6 µALaboratory / Customer PremisesActive
  • >100 mA to 200 mA
58x10-6 I + 3.9 µALaboratory / Customer PremisesActive
  • >200 mA to 1 A
92x10-6 I + 41 µALaboratory / Customer PremisesActive
  • >1 A to 2 A
92x10-6 I + 42 µALaboratory / Customer PremisesActive
  • >2 A to 10 A
0.35x10-3 I + 0.58 mALaboratory / Customer PremisesActive
  • >10 A to 20 A
0.35x10-3 I + 0.58 mALaboratory / Customer PremisesActive
  • >20 A to 30 A
0.35x10-3 I + 3.2 mALaboratory / Customer PremisesActive
  • AC Current Calibration of meters
  • Direct Method Using Multi-Product Calibrator I : Measured current value
  • 0 to 25 µA @ 40 Hz to 206 Hz
0.60x10-3 I + 0.29 µALaboratory / Customer PremisesActive
  • <25 µA to 200 µA @ 40 Hz to 1 kHz
0.50x10-3 I + 0.30 µALaboratory / Customer PremisesActive
  • <200 µA to 2 mA @ 40 Hz to 1 kHz
0.50x10-3 I + 0.80 µALaboratory / Customer PremisesActive
  • 2 mA to 20 mA @ 40 Hz to 1 kHz
0.50x10-3 I + 6.5 µALaboratory / Customer PremisesActive
  • <20 mA to 200 mA @ 40 Hz to 1 kHz
0.50x10-3 I + 0.07 mALaboratory / Customer PremisesActive
  • <200 mA to 2 A @ 40 Hz to 1 kHz
0.60x10-3 I + 1.8 mALaboratory / Customer PremisesActive
  • <2 A to 20 A @ 40 Hz to 1 kHz
0.60x10-3 I + 8.7 mALaboratory / Customer PremisesActive
  • <20 A to 30 A @ 40 Hz to 1 kHz
0.60x10-3 I + 12 mALaboratory / Customer PremisesActive
  • Resistance – 2 Wires Calibration of meters
  • Direct Method Using Multi-Product Calibrator R = Measured resistance value
  • 100 mΩ to 1 Ω
0.16x10-3 R + 6.2 mΩLaboratory / Customer PremisesActive
  • <1 Ω to 10 Ω
0.10 x10-3 R + 6.3 mΩLaboratory / Customer PremisesActive
  • <10 Ω to 100 Ω
0.10x10-3 R + 6.0 mΩLaboratory / Customer PremisesActive
  • 100 mΩ
0.16x10-3 R + 5.8 mΩLaboratory / Customer PremisesActive
  • 1 Ω
0.16x10-3 R + 5.8 mΩLaboratory / Customer PremisesActive
  • 10 Ω
0.10 x10-3 R + 5.9 mΩLaboratory / Customer PremisesActive
  • 100 Ω
0.10x10-3 R + 6.2 mΩLaboratory / Customer PremisesActive
  • 1 kΩ
0.05x10-3 R + 7.0 mΩLaboratory / Customer PremisesActive
  • 10 kΩ
0.04x10-3 R + 52 mΩLaboratory / Customer PremisesActive
  • 100 kΩ
0.04x10-3 R + 0.8 ΩLaboratory / Customer PremisesActive
  • 1 MΩ
0.03x10-3 R + 16 ΩLaboratory / Customer PremisesActive
  • 10 MΩ
0.10x10-3 R + 0.58 kΩLaboratory / Customer PremisesActive
  • 100 MΩ
4.6x10-3 R + 12 kΩLaboratory / Customer PremisesActive
  • 1 GΩ
10x10-3 R + 0.50 MΩLaboratory / Customer PremisesActive
  • Resistance – 4 Wires Calibration of meters
  • 400 µΩ
0.12x10-3 R + 0.3 µΩLaboratory / Customer PremisesActive
  • 1 mΩ
0.06 x10-3 R + 63 µΩLaboratory / Customer PremisesActive
  • 100 mΩ
0.16x10-3 R + 6.0 mΩLaboratory / Customer PremisesActive
  • 1 Ω
0.10 x10-3 R + 6.0 mΩLaboratory / Customer PremisesActive
  • 10 Ω
0.10x10-3 R + 6.0 mΩLaboratory / Customer PremisesActive
  • 100 Ω
0.05x10-3 R + 7.0 mΩLaboratory / Customer PremisesActive
  • 1 kΩ
0.04x10-3 R + 50 mΩLaboratory / Customer PremisesActive
  • 10 kΩ
0.04x10-3 R + 0.9 ΩLaboratory / Customer PremisesActive
  • 100 kΩ
0.04x10-3 R + 0.9 ΩLaboratory / Customer PremisesActive
  • Capacitance Calibration of Meters
  • Direct Method Using Multi-Product Calibrator C = Measured capacitance value
  • 1 nF
2.9x10-3 C + 6.1 pFLaboratory / Customer PremisesActive
  • 10 nF
2.9x10-3 C + 0.02 nFLaboratory / Customer PremisesActive
  • 20 nF
2.9x10-3 C + 0.04 nFLaboratory / Customer PremisesActive
  • 50 nF
2.9x10-3 C + 0.10 nFLaboratory / Customer PremisesActive
  • 100 nF
2.9x10-3 C + 0.10 nFLaboratory / Customer PremisesActive
  • 1 µF
4.6x10-3 C + 0.38 nFLaboratory / Customer PremisesActive
  • 10 µF
6.9x10-3 C + 23 nFLaboratory / Customer PremisesActive
  • Frequency Calibration of Meters
  • Direct Method Using Multi-Product Calibrator f = Measured frequency Value
  • 1 Hz to 1 kHz
1.2x10-6 f + 50 µHzLaboratory / Customer PremisesActive
  • >1 kHz to 10 kHz
1.2x10-6 f + 0.4 mHzLaboratory / Customer PremisesActive
  • >10 kHz to 20 kHz
1.2x10-6 f + 4.5 mHzLaboratory / Customer PremisesActive
  • >20 kHz to 50 kHz
1.2x10-6 f + 9.5 mHzLaboratory / Customer PremisesActive
  • >50 kHz to 100 kHz
1.2x10-6 f + 45 mHzLaboratory / Customer PremisesActive
  • >100 kHz to 1 MHz
1.2x10-6 f + 0.45 HzLaboratory / Customer PremisesActive
  • >1 MHz to 10 MHz
1.2x10-6 f + 2.6 HzLaboratory / Customer PremisesActive
  • Continuity Resistance Calibration of meters
  • Direct Method Using Electrical Calibrator
  • 0 to 200 mΩ
30.1 mΩLaboratory / Customer PremisesActive
  • <200 mΩ to 500 mΩ
30.9 mΩLaboratory / Customer PremisesActive
  • 500 mΩ to 800 mΩ
32 mΩLaboratory / Customer PremisesActive
  • <800 mΩ to 1 Ω
32 mΩLaboratory / Customer PremisesActive
  • <1 Ω to 2 Ω
35 mΩLaboratory / Customer PremisesActive
  • <2 Ω to 4 Ω
41 mΩLaboratory / Customer PremisesActive
  • <4 Ω to 6 Ω
47 mΩLaboratory / Customer PremisesActive
  • <6 Ω to 8 Ω
52 mΩLaboratory / Customer PremisesActive
  • <8 Ω to 10 Ω
58 mΩLaboratory / Customer PremisesActive
  • <10 Ω to 20 Ω
87 mΩLaboratory / Customer PremisesActive
  • <20 Ω to 100 Ω
0.3 ΩLaboratory / Customer PremisesActive
  • <100 Ω to 1000 Ω
2.9 ΩLaboratory / Customer PremisesActive
  • Insulation Resistance Calibration of Meters
  • 10 kΩ to 100 kΩ
0.59 kΩLaboratory / Customer PremisesActive
  • <100 kΩ to 200 kΩ
0.62 kΩLaboratory / Customer PremisesActive
  • <200 kΩ to 400 kΩ
5.8 kΩLaboratory / Customer PremisesActive
  • <400 kΩ to 600 kΩ
5.8 kΩLaboratory / Customer PremisesActive
  • <600 kΩ to 1 MΩ
5.9 kΩLaboratory / Customer PremisesActive
  • <1 MΩ to 2 MΩ
6.2 kΩLaboratory / Customer PremisesActive
  • <2 MΩ to 3 MΩ
6.7 kΩLaboratory / Customer PremisesActive
  • <3 MΩ to 5 MΩ
8.2 kΩLaboratory / Customer PremisesActive
  • <5 MΩ to 7 MΩ
81 kΩLaboratory / Customer PremisesActive
  • <7 MΩ to 9 MΩ
0.10 MΩLaboratory / Customer PremisesActive
  • <9 MΩ to 10 MΩ
0.11 MΩLaboratory / Customer PremisesActive
  • <10 MΩ to 20 MΩ
0.23 MΩLaboratory / Customer PremisesActive
  • <20 MΩ to 30 MΩ
0.34 MΩLaboratory / Customer PremisesActive
  • <30 MΩ to 50 MΩ
0.57 MΩLaboratory / Customer PremisesActive
  • <50 MΩ to 70 MΩ
0.80 MΩLaboratory / Customer PremisesActive
  • <70 MΩ to 90 MΩ
1.0 MΩLaboratory / Customer PremisesActive
  • <90 MΩ to 100 MΩ
1.2 MΩLaboratory / Customer PremisesActive
  • <100 MΩ to 200 MΩ
2.8 MΩLaboratory / Customer PremisesActive
  • <200 MΩ to 400 MΩ
5.6 MΩLaboratory / Customer PremisesActive
  • <400 MΩ to 600 MΩ
8.5 MΩLaboratory / Customer PremisesActive
  • <600 MΩ to 800 MΩ
11 MΩLaboratory / Customer PremisesActive
  • <800 MΩ to 1000 MΩ
14 MΩLaboratory / Customer PremisesActive
  • 10GΩ
50MΩLaboratory / Customer PremisesActive
  • DC Voltage Calibration of sources
  • Direct Method Using Precision Multimeter / Comparison Method U = Measured Voltage value
  • 0.001 to 100 mV
6 x10-6 U + 3.8 µVLaboratory / Customer PremisesActive
  • <100 mV to 1 V
5 x10-6 U + 3.9 µVLaboratory / Customer PremisesActive
  • <1 V to 10 V
5 x10-6 U + 38 µVLaboratory / Customer PremisesActive
  • <10 V to 100 V
7 x10-6 U + 0.39 mVLaboratory / Customer PremisesActive
  • <100 V to 1000 V
7 x10-6 U + 5.8 mVLaboratory / Customer PremisesActive
  • AC voltage Calibration of sources
  • 0.001 to 10 mV @ 1 kHz
0.2 x10-3 U + 1.8 µVLaboratory / Customer PremisesActive
  • <10 mV to 100 mV @ 1 kHz
0.1 x10-3 U + 2.8 µVLaboratory / Customer PremisesActive
  • <100 mV to 1 V @ 1 kHz
0.1 x10-3 U +0.23 mVLaboratory / Customer PremisesActive
  • <1 V to 10 V @ 1 kHz
0.1 x10-3 U +0.27 mVLaboratory / Customer PremisesActive
  • <10 V to 100 V @ 1 kHz
0.1 x10-3 U +0.27 mVLaboratory / Customer PremisesActive
  • <100 V to 700 V @ 1 kHz
0.1 x10-3 U +30 mVLaboratory / Customer PremisesActive
  • 0 to 1 V @ 20 Hz
0.1 x10-3 U + 61 µVLaboratory / Customer PremisesActive
  • <1 V to 10 V @ 20 Hz
0.1 x10-3 U + 0.76 mVLaboratory / Customer PremisesActive
  • <10 V to 100 V @ 20 Hz
0.1 x10-3 U + 6.7 mVLaboratory / Customer PremisesActive
  • 0 to 1 V @ 50 Hz
0.1 x10-3 U + 0.38 mVLaboratory / Customer PremisesActive
  • <1 V to 10 V @ 50 Hz
0.1 x10-3 U + 5.03 mVLaboratory / Customer PremisesActive
  • <10 V to 100 V @ 50 Hz
0.2 x10-3 U + 32 mVLaboratory / Customer PremisesActive
  • <100 V to 700 V @ 50 Hz
0.7 x10-3 U + 0.30 VLaboratory / Customer PremisesActive
  • 0 to 1 V @ 20 kHz
0.2 x10-3 U + 0.23 mVLaboratory / Customer PremisesActive
  • <1 V to 10 V @ 20 kHz
0.2 x10-3 U + 0.44 mVLaboratory / Customer PremisesActive
  • <10 V to 100 V @ 20 kHz
0.2 x10-3 U + 7.2 mVLaboratory / Customer PremisesActive
  • <100 V to 700 V @ 20 kHz
0.7 x10-3 U + 55 mVLaboratory / Customer PremisesActive
  • 0 to 1 V @ 100 kHz
0.9 x10-3 U + 0.23 mVLaboratory / Customer PremisesActive
  • <1 V to 10 V @ 100 kHz
0.9 x10-3 U + 0.42 mVLaboratory / Customer PremisesActive
  • <10 V to 100 V @ 100 kHz
1.4 x10-3 U + 11 mVLaboratory / Customer PremisesActive
  • Resistance Calibration of sources
  • Direct Method using Precision Multimeter / Comparison Method R = Reading Resistance
  • (2-Wire) 0.01 Ω to 10 Ω
17 x10-6 R + 0.05 mΩLaboratory / Customer PremisesActive
  • (2-Wire) <10 Ω to 100 Ω
14 x10-6 R + 0.58 mΩLaboratory / Customer PremisesActive
  • (2-Wire) <100 Ω to 1 kΩ
12 x10-6 R + 1.7 mΩLaboratory / Customer PremisesActive
  • (2-Wire) <1 kΩ to 10 kΩ
12 x10-6 R + 8.9 mΩLaboratory / Customer PremisesActive
  • (2-Wire) <10 kΩ to 10 MΩ
58 x10-6 R +0.11 kΩLaboratory / Customer PremisesActive
  • (2-Wire) <10 MΩ to 100 MΩ
0.58 x10-3 R + 1.2 kΩLaboratory / Customer PremisesActive
  • (2-Wire) <100MΩ to 1 GΩ
5.8 x10-3 R + 12 kΩLaboratory / Customer PremisesActive
  • (4-Wire) 0.01 Ω to 10 Ω
17 x10-6 R + 0.05 mΩLaboratory / Customer PremisesActive
  • (4-Wire) <10 Ω to 100 Ω
14 x10-6 R + 0.58 mΩLaboratory / Customer PremisesActive
  • (4-Wire) <100 Ω to 1 kΩ
12 x10-6 R + 1.7 mΩLaboratory / Customer PremisesActive
  • (4-Wire) <1 kΩ to 10 kΩ
12 x10-6 R + 8.9 mΩLaboratory / Customer PremisesActive
  • (4-Wire) <10 kΩ to 100 kΩ
12 x10-6 R + 1.2 ΩLaboratory / Customer PremisesActive
  • DC Current Calibration of sources
  • Direct Method Using Precision Multi-meter / Comparison Method
  • 0 µA to 1 μA
23 x10-6 I + 0.05 nALaboratory/ Customer PremisesActive
  • <1 μA to 10 μA
23 x10-6 I + 0.13 nALaboratory/ Customer PremisesActive
  • <10 μA to 100 μA
23 x10-6 I + 0.41 nALaboratory/ Customer PremisesActive
  • <100 μA to 1 mA
23 x10-6 I + 4.1 nALaboratory/ Customer PremisesActive
  • <1 mA to 10 mA
23 x10-6 I + 0.04 µALaboratory/ Customer PremisesActive
  • <10 mA to 100 mA
40 x10-6 I + 0.5 µALaboratory/ Customer PremisesActive
  • <100 mA to 1 A
0.13 x10-3 I + 22 µALaboratory/ Customer PremisesActive
  • <1 A to 1 0A
17 x10-3 I + 5.1 mALaboratory/ Customer PremisesActive
  • AC Current calibration of sources
  • 0 to 100 μA @ 20 Hz to 1 kHz
0.70 x10-3 I + 0.04 µALaboratory/ Customer PremisesActive
  • <100 μA to 1 mA @ 20 Hz to 1 kHz
0.30 x10-3 I + 0.29 µALaboratory/ Customer PremisesActive
  • < 1 mA to 10 mA @ 20 Hz to 1 kHz
0.30 x10-3 I + 23 µALaboratory/ Customer PremisesActive
  • <10 mA to 100 mA @ 20 Hz to 1 kHz
0.30 x10-3 I + 26 µALaboratory/ Customer PremisesActive
  • <100 mA to 1A @ 20 Hz to 1 kHz
1.2 x10-3 I + 0.33 mALaboratory/ Customer PremisesActive
  • <1 A to 10 A @ 56 kHz
1.7 x10-3 I + 13 mALaboratory/ Customer PremisesActive
  • Insulation Resistance Voltage Measurement
  • Direct Method Using Electrical Calibrator
  • 50 V
1.5 VLaboratory/ Customer PremisesActive
  • 100 V
2.1 VLaboratory/ Customer PremisesActive
  • 250 V
3.8 VLaboratory/ Customer PremisesActive
  • 500 V
6.7 VLaboratory/ Customer PremisesActive
  • 1000 V
13 VLaboratory/ Customer PremisesActive
  • 5000 V
60 VLaboratory/ Customer PremisesActive
  • RCD Current
  • 0 to 10 mA @ 50 Hz
0.20 mALaboratory/ Customer PremisesActive
  • <10 mA to 30 mA @ 50 Hz
0.49 mALaboratory/ Customer PremisesActive
  • <30 mA to 90 mA @ 50 Hz
1.3 mALaboratory/ Customer PremisesActive
  • <90 mA to 100 mA @ 50 Hz
1.5 mALaboratory/ Customer PremisesActive
  • <100 mA to 150 mA @ 50 Hz
2.8 mALaboratory/ Customer PremisesActive
  • <150 mA to 300 mA @ 50 Hz
5.0 mALaboratory/ Customer PremisesActive
  • <300 mA to 1000 mA @ 50 Hz
21 mALaboratory/ Customer PremisesActive
  • <1000 mA to 2000 mA @ 50 Hz
35 mALaboratory/ Customer PremisesActive
  • Portable Appliance Earth Bond Resistance Calibration of meters
  • 30 mΩ
5.0 mΩLaboratory / Customer PremisesActive
  • 100 mΩ
5.3 mΩLaboratory / Customer PremisesActive
  • 200 mΩ
5.6 mΩLaboratory / Customer PremisesActive
  • 300 mΩ
6.3 mΩLaboratory / Customer PremisesActive
  • 400 mΩ
6.9 mΩLaboratory / Customer PremisesActive
  • 500 mΩ
7.9 mΩLaboratory / Customer PremisesActive
  • 1 Ω
11 mΩLaboratory / Customer PremisesActive
  • 5 Ω
34 mΩLaboratory / Customer PremisesActive
  • 10 Ω
63 mΩLaboratory / Customer PremisesActive
  • 100 Ω
0.58 ΩLaboratory / Customer PremisesActive
  • 1000 Ω
5.8 ΩLaboratory / Customer PremisesActive
  • Leakage current
  • 2 mA
38 µALaboratory / Customer PremisesActive
  • 4.7 mA
84 µALaboratory / Customer PremisesActive
  • 7 mA
0.13 mALaboratory / Customer PremisesActive
  • RCD Trip Time
  • 20 ms
0.9 msLaboratory / Customer PremisesActive
  • 40 ms
1.0 msLaboratory / Customer PremisesActive
  • 100 ms
1.0 msLaboratory / Customer PremisesActive
  • 200 ms
1.0 msLaboratory / Customer PremisesActive
  • 390 ms
1.1 msLaboratory / Customer PremisesActive
  • 900 ms
8.2 msLaboratory / Customer PremisesActive
  • Loop Impedance
  • 0.058 Ω
5.1 mΩLaboratory / Customer PremisesActive
  • 0.110 Ω
5.4 mΩLaboratory / Customer PremisesActive
  • 0.227 Ω
6.1 mΩLaboratory / Customer PremisesActive
  • 0.336 Ω
6.7 mΩLaboratory / Customer PremisesActive
  • 0.510 Ω
7.7 mΩLaboratory / Customer PremisesActive
  • 1.005 Ω
11 mΩLaboratory / Customer PremisesActive
  • 5.019 Ω
34 mΩLaboratory / Customer PremisesActive
  • 9.977 Ω
63 mΩLaboratory / Customer PremisesActive
  • 99.231 Ω
0.56 ΩLaboratory / Customer PremisesActive
  • 982.94 Ω
5.7 ΩLaboratory / Customer PremisesActive
  • Time Base
  • Direct Method Using Oscilloscope calibrator / Comparison method using Oscilloscope
  • 5 ns
13 psLaboratory / Customer PremisesActive
  • 5ns to 10 ns
55 psLaboratory / Customer PremisesActive
  • 10 ns to 100ns
86 psLaboratory / Customer PremisesActive
  • 100ns to 1 μs
0.86nsLaboratory / Customer PremisesActive
  • 1 μs to 10 μs
22 nsLaboratory / Customer PremisesActive
  • 10 μs to 100 μs
0.14 μsLaboratory / Customer PremisesActive
  • 100 μs to 1 ms
0.76 μsLaboratory / Customer PremisesActive
  • 1ms to 10 ms
16 μsLaboratory / Customer PremisesActive
  • 10 ms to 100ms
85 μsLaboratory / Customer PremisesActive
  • 100 ms to 1 s
56 msLaboratory / Customer PremisesActive
  • 1s to 5s
0.12 sLaboratory / Customer PremisesActive
  • AC High Current Calibration of meters
  • Direct Method Using Multi-Product Calibrator and Current Coil Adapter
  • <10 A @ 30Hz to 60 Hz
0.42 x10-2 I + 46 mALaboratory / Customer PremisesActive
  • 10 A to 100 A @ 30Hz to 60 Hz
0.42 x10-2 I + 0.12 ALaboratory / Customer PremisesActive
  • 100 A to 300 A @ 30Hz to 60 Hz
0.30 x10-2 I + 1.0 ALaboratory / Customer PremisesActive
  • 300 A to 1000 A @ 30Hz to 60 Hz
0.30 x10-2 I + 1.2 ALaboratory / Customer PremisesActive
  • 1000 A to 1500 A @ 30Hz to 60Hz
0.30 x10-2 I + 1.3 ALaboratory / Customer PremisesActive
  • DC High Current Calibration of mters
  • <10 A
0.42 x10-2 I + 46 mALaboratory / Customer PremisesActive
  • 10 A to 100 A
0.42 x10-2 I + 0.12 ALaboratory / Customer PremisesActive
  • 100 A to 300 A
0.30 x10-2 I + 1.0 ALaboratory / Customer PremisesActive
  • 300 A to 1000 A
0.30 x10-2 I + 1.2 ALaboratory / Customer PremisesActive
  • 1000 A to 1500 A
0.30 x10-2 I + 1.3 ALaboratory / Customer PremisesActive
  • Frequency Measurement
  • Comparison method Using Frequency Counter / Comparison Method F = Measured Frequency
  • < 100 Hz
2.3x10-6 f + 0.12 mHzLaboratory / Customer PremisesActive
  • 100 Hz to 100 MHz
2.3x10-6 f + 58 mHzLaboratory / Customer PremisesActive
  • 100 MHz to 500 MHz
1.2x10-6 f + 0.58 kHzLaboratory / Customer PremisesActive
  • 1 GHz to 4 GHz
1.2x10-6 f + 1.2 kHzLaboratory / Customer PremisesActive
  • 4 GHz to 8 GHz
1.2x10-6 f + 9.2 kHzLaboratory / Customer PremisesActive
  • 8 GHz to 15 GHz
1.2x10-6 f + 17 kHzLaboratory / Customer PremisesActive
  • Comparison method using Frequency Counter
  • 15 GHz to 20 GHz
1.2x10-6 f + 23 kHzLaboratory / Customer PremisesActive
  • Defibrillator Analyzer
  • Comparison method
  • 0 to 20 J
0.23JLaboratory PremisesActive
  • 20 to 50 J
0.58 JLaboratory PremisesActive
  • 50 to 100 J
1.2 JLaboratory PremisesActive
  • 100 to 200 J
2.3 JLaboratory PremisesActive
  • 200 to 300 J
4.2 JLaboratory PremisesActive
  • RF Absolute Power
  • Power Meter with Power sensor
  • 10 MHz -30 dBm to 20 dBm
0.04Laboratory PremisesActive
  • 30 MHz -30 dBm to 20 dBm
0.039Laboratory PremisesActive
  • 100 MHz -30 dBm to 20 dBm
0.038Laboratory PremisesActive
  • 300 MHz -30 dBm to 20 dBm
0.038Laboratory PremisesActive
  • 500 MHz -30dBm to 20 dBm
0.038Laboratory PremisesActive
  • 1.0 GHz -30 dBm to 20 dBm
0.038Laboratory PremisesActive
  • 2.0 GHz -30 dBm to 20 dBm
0.039Laboratory PremisesActive
  • 4.0 GHz -30 dBm to 20 dBm
0.039Laboratory PremisesActive
  • 6.0 GHz -30 dBm to 20 dBm
0.039Laboratory PremisesActive
  • 10.0 GHz -30 dBm to 20 dBm
0.04Laboratory PremisesActive
  • 12.0 GHz -30 dBm to 20 dBm
0.039Laboratory PremisesActive
  • 15.0 GHz -30 dBm to 20 dBm
0.04Laboratory PremisesActive
  • 17.0 GHz -30 dBm to 20 dBm
0.044Laboratory PremisesActive
  • 18.0 GHz -30 dBm to 20 dBm
0.048Laboratory PremisesActive
  • RF Relative Power (2.5MHz to 1.3 GHz)
  • Direct Measurement with Measuring Receiver
  • -10 dB
0.13 dBLaboratory PremisesActive
  • -20 dB
0.13 dBLaboratory PremisesActive
  • -30 dB
0.13 dBLaboratory PremisesActive
  • -40 dB
0.13 dBLaboratory PremisesActive
  • -50 dB
0.15 dBLaboratory PremisesActive
  • -60 dB
0.15 dBLaboratory PremisesActive
  • -70 dB
0.14 dBLaboratory PremisesActive
  • -80 dB
0.15 dBLaboratory PremisesActive
  • -90 dB
0.16 dBLaboratory PremisesActive
  • -100 dB
0.16 dBLaboratory PremisesActive
  • -110 dB
0.16 dBLaboratory PremisesActive
  • -120 dB
0.17 dBLaboratory PremisesActive
  • Amplitude Modulation (5% to 90%)
  • (10 MHz to 1.3 GHz Modulation Rate: 10 kHz) >Depth: 40%
0.0068Laboratory PremisesActive
  • (10 MHz to 1.3 GHz Modulation Rate: 10 kHz) >Depth: 90%
0.012Laboratory PremisesActive
  • Frequency Modulation
  • (250 kHz to 10 MHz (Dev.: 10 kHz peak)) Mod Rate : 10 kHz
0.24 kHz peakLaboratory PremisesActive
  • (250 kHz to 10 MHz (Dev.: 10 kHz peak)) Mod Rate : 20 Hz
0.24 kHz PeakLaboratory PremisesActive
  • (10MHz to 1.3GHz (Dev.: 100kHz peak)) Mod Rate : 100 kHz
1.2 kHz peakLaboratory PremisesActive
  • (10MHz to 1.3GHz (Dev.: 100kHz peak)) Mod Rate : 50 kHz
1.2 kHz peakLaboratory PremisesActive
  • (10MHz to 1.3GHz (Dev.: 100kHz peak)) Mod Rate : 20 Hz
2.5 kHz peakLaboratory PremisesActive
Pressure
  • Gas Pressure (absolute gauge, transmitter, indicators)
  • DKD-R 6-1:2014
  • 75 kPa to 115 kPa
0.029% F.S of UUCLaboratory PremisesActive
  • Gas Pressure (vacuum gauge, transmitter, indicators)
  • -90 kPa to 0
0.053% F.S of UUCLaboratory PremisesActive
  • Gas Pressure (gauge, transmitter, indicators)
  • -300 mmHg to 300 mmHg
0.04 % F.S of UUCLaboratory / Customer PremisesActive
  • 0 to 0.35 MPa
0.015 % F.S of UUCLaboratory / Customer PremisesActive
  • > 0.35 MPa to 2 MPa
0.021 % F.S of UUCLaboratory / Customer PremisesActive
  • Hydraulic Pressure (gauge, transmitter, indicators)
  • 0 bar to 700 bar
0.031% F.S of UUCLaboratory / Customer PremisesActive
  • NIBP Pressure (Mechanical &Automated Sphygmomanometer
  • OIML R16-1(E) 2002 OIML R16-2(E) 2002
  • 0 to 300 mmHg
0.18% F.S of UUCLaboratory / Customer PremisesActive
  • NIBP Pressure (Patient Simulator)
  • Manufacturer procedure
  • 0 to 400 mmHg
0.11 F.S of UUCLaboratory / Customer PremisesActive
  • Pneumatic Pressure (Infusion Device Analyzer)
  • -500 mmHg to 1500 mmHg
0.12 F.S of UUCLaboratory / Customer PremisesActive
Temperature
  • Liquid in Glass Thermometer
  • CMT-01
  • -20 °C to 95 °C
0.3 °CLaboratory PremisesActive
  • Digital Thermometer with sensor, RTD, Thermocouple, (Gauges, Transmitter, switches))
  • CMT-02
  • -39°to 150°C
0.3°CLaboratory PremisesActive
  • 151°C to 400°C
0.3°CLaboratory PremisesActive
  • 400°C to 650°C
0.7°CLaboratory PremisesActive
  • -39°to 150°C
0.5°CCustomer PremisesActive
  • 151°C to 400°C
1.0°CCustomer PremisesActive
  • 400°C to 650°C
2.0°CCustomer PremisesActive
  • Dry Block Calibrators
  • CMT-03
  • -39 to 150°C
1°CLaboratory/ Customer PremisesActive
  • 151 to 400°C
2°CLaboratory/ Customer PremisesActive
  • 400°C to 650°C
5°CLaboratory/ Customer PremisesActive
  • Indicator for TC Type K
  • CMT -04
  • -140°C to 1340°C
0.3 °CLaboratory PremisesActive
  • Indicator for TC Type J
  • 400°C to 750°C
0.3 °CLaboratory PremisesActive
  • Indicator for TC Type T
  • -250°C to 400°C
0.3 °CLaboratory PremisesActive
  • Oven (9 points) Freezer (9 points) Chiller (9 points) Liquid Bath (9 points) Incubator (9 points)
  • CMT -05 (DKD- R.5-7)
  • -39°C to 200°C
0.9°CLaboratory/ Customer PremisesActive
  • 200°C to 400°C
1.5°CLaboratory/ Customer PremisesActive
  • Oven (9 points) Freezer (9 points) Chiller (9 points) Liquid Bath (5 points) Incubator (5-9 points)
  • -39°C to 200°C
0.9°CLaboratory/ Customer PremisesActive
  • 200°C to 400°C
1.5°CLaboratory/ Customer PremisesActive
  • Digital Therometer, Thermo Hygrometer, Data Logger, Humidity Meter, transmitter
  • CMT-06
  • 20 % to 90 % RH
1.1 % RHLaboratory PremisesActive
  • 5°C to 60°C
0.35 °CLaboratory PremisesActive
  • IR Thermometer
  • CMT-07
  • 30 to 100°C
1°CLaboratory PremisesActive
  • 100 to 200°C
2°CLaboratory PremisesActive
  • 200°C to 300°C
3°CLaboratory PremisesActive
Volume
  • Volume Pipette
  • Gravimetric Method for Volume – Calibration of Micropipette no CMV-01 based on ISO 8655-6:2022
  • 10µl to 20µl
0.23 µlLaboratory PremisesActive
  • >20µl to 100µl
0.28 µlLaboratory PremisesActive
  • > 100µl to 500µl
0.43 µlLaboratory PremisesActive
  • > 500µl to <1000 µl
0.85 µlLaboratory PremisesActive
  • 1000 μl to 2500 μl
4.0 μlLaboratory PremisesActive
  • > 2500µl to 5000 µl
13.0 µlLaboratory PremisesActive
Medical Devices
  • NIBP Pressure (Mechanical & Automated Sphygmomanometer)
  • OIML R 16-1 (E) 2002 OIML R 16-2 (E) 2002
  • 0 to 300 mmHg
0.10 % F.S of UUCLaboratory / Customer premisesActive
  • NIBP Pressure (Patient Simulator)
  • Manufacturer procedure
  • 0 to 400 mmHg
0.07 % F.S of UUCLaboratory / Customer premisesActive
  • Pneumatic Pressure (Infusion Device Analyzer)
  • -500 mmHg to 1500 mmHg
0.07 % F.S of UUCLaboratory / Customer premisesActive
  • Defibrillator Analyzer
  • 0 to 20 J
0.24 JLaboratoryActive
  • 20 to 50 J
0.60 JLaboratoryActive
  • 50 to 100 J
1.2 JLaboratoryActive
  • 100 to 200 J
2.3 JLaboratoryActive
  • 200 to 360 J
4.2 JLaboratoryActive
Speed & Time
  • Tachometer Calibration
  • Comparison method using non-contact tachometer and multi-product calibrator Method PM-06
  • 60 up to 240 rpm
0.2 rpmLaboratory PremisesActive
  • 240 up to 6000 rpm
1.9 rpmLaboratory PremisesActive
  • 6000 up to 60000 rpm
1.1 x 10-4 S +0.7 rpmLaboratory PremisesActive
  • Rotation Speed Calibration (Centrifuge, Mixers, Orbital Shaker, Plate Shaker, RPM Apparatus etc..)
  • Comparison method using non-contact tachometer Method PM-11
  • 60 up to 800 rpm
1.0 rpmLaboratory/ Customer PremisesActive
  • 800 up to 13000 rpm
5.5 x 10-4 S +1.7 rpmLaboratory/ Customer PremisesActive
  • 13000 up to 30000 rpm
5.9 x 10-4 S + 2.1 rpmLaboratory/ Customer PremisesActive
Torque
  • Hand torque Tools
  • Based on ISO 6789-1:2017 and ISO 6789-2:2017
  • 0.1 N·m to 1 N·m
1.5 %Laboratory PremisesActive
  • Based on ISO 6789-1:2017 and ISO 6789-2:20187
  • > 1 N·m to 2000 N·m
1.0 %Laboratory PremisesActive