Total Weighing And Measuring LLC

Initial Accreditation Date:Jan 09 , 2019

معلومات الاتصال

العنوان

Office 308, Dubai Financial Support Fund - Al Quoz Third., Dubai, United Arab Emirates

صندوق بريد / الرمز البريدي

10239

البريد الإلكتروني

info@tcal.ae qaqc@tcal.ae

الموقع الإلكتروني

www.tacl.ae

رقم الهاتف

+971 054 426 6403

المجال الرئيسي

Calibration Laboratories

ساري منذ:Jan 09 , 2025

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Balance
  • Weighing Balance
  • TCAL-WI-11 using Standard Weights based on Euramet cg 18
  • 0 up to 500 g
0.86 mgCustomers PremisesActive
  • >500g up to 5000g
0.39 gCustomers PremisesActive
  • >5000g up to 10000g
2.0 gCustomers PremisesActive
  • >10 kg up to 100 kg
20 gCustomers PremisesActive
  • >100 kg up to 500 kg
98 gCustomers PremisesActive
  • TCAL-WI-11 Using Standard Weights based on Euramet cg-18 v 4.0
  • >500kg to 1000kg
0.19 kgCustomer PremisesActive
Pressure
  • Gauge Pressure: Medium: Pneumatic
  • TCAL-WI-01 by Comparison method using Digital Pressure Gauge as per DKD-R 6-1
  • -90 kPa to 0
5.8 kPaLaboratory and Customer PremisesActive
  • 0 to 700 kPa
0.50 kPaLaboratory and Customer PremisesActive
  • Gauge Pressure: Medium: Hydraulic
0.50 kPaLaboratory and Customer PremisesActive
  • >0.7 MPa to 7 MPa
2.7 kPaLaboratory and Customer PremisesActive
  • >7 MPa up to 100 MPa
29 kPaLaboratory and Customer PremisesActive
  • Gauge Pressure Pneumatic Analog and Digital Pressure Gauges
  • TCAL-WI-01 by Comparison method based on requirement of DKD-R 6-1 (03/2014)
  • 0.7 MPa to 6 MPa
2.9 kPaLaboratory and Customer PremisesActive
  • Gauge Pressure Hydraulic Pressure Transmitters
  • TCAL-WI-02 by Comparison method based on requirement of DKD-R 6-1 (03/2014)
  • 0 to 10 MPa
42 kPaLaboratory and Customer PremisesActive
Electrical
  • DC Voltage
  • Direct Method Using Multi-product Calibrator according to TCAL Calibration Procedures TCAL-WI-28 & 30
  • Up to 200 mV
0.02 mVLaboratoryActive
  • < 200 mV to 2 V
0.17 mVLaboratoryActive
  • < 2 V to 20 V
1.6 mVLaboratoryActive
  • < 20 V to 200 V
16 mVLaboratoryActive
  • < 200 V to 1000 V
81 mVLaboratoryActive
  • DC Current
  • Up to 200 µA
0.05 µALaboratoryActive
  • < 200 µA to 2 mA
0.33 µALaboratoryActive
  • < 2 mA to 20 mA
3.3 µALaboratoryActive
  • < 20 mA to 200 mA
37 µALaboratoryActive
  • < 200 mA to 2 A
1.3 mALaboratoryActive
  • < 2 A to 20 A
13 mALaboratoryActive
  • Direct Method Using Multi product Calibrator and Current Coil Station
  • < 20 A to 50 A
0.41 ALaboratoryActive
  • < 50 A to 100 A
0.87 ALaboratoryActive
  • < 100 A to 250 A
2.0 ALaboratoryActive
  • < 250 A to 500 A
3.4 ALaboratoryActive
  • < 500 A to 750 A
4.8 ALaboratoryActive
  • < 750 A to 1000 A
6.2 ALaboratoryActive
  • AC Voltage
  • Direct Method Using Multi-product Calibrator according to TCAL Calibration Procedures TCAL-WI-28 & 30
  • Up to 200 mV @ 40 Hz to 10 kHz
0.12 mVLaboratoryActive
  • (< 200 mV to 2 V) @ 40 Hz to 10 kHz
1.1 mVLaboratoryActive
  • (< 2 V to 20 V) @ 40 Hz to 10 kHz
11 mVLaboratoryActive
  • <20V to 200 V @ 40 Hz to 10 kHz
0.13 VLaboratoryActive
  • (< 200 V to 700 V) @ 40 Hz to 10 kHz
0.51 VLaboratoryActive
  • (< 200 V to 1000 V ] @ 40 Hz to 10 kHz
0.68 VLaboratoryActive
  • AC Current
  • Up to 200 µA @ 50 Hz
0.51 µALaboratoryActive
  • (< 200 µA to 2 mA) @ 50 Hz
2.2 µALaboratoryActive
  • (< 2mA to 20 mA) @ 50 Hz
20 µALaboratoryActive
  • (< 20mA to 200 mA) @ 50 Hz
0.2 mALaboratoryActive
  • (< 200 mA to 2 A) @ 50 Hz
2.6 mALaboratoryActive
  • (< 2 A to 20 A) @ 50 Hz
42 mALaboratoryActive
  • Direct Method Using Multi product Calibrator and Current Coil Station
  • (< 20 A to 50 A) @ 50 Hz
0.42 ALaboratoryActive
  • (< 50 A to 100 A) @ 50 Hz
0.89 ALaboratoryActive
  • (< 100 A to 250 A) @ 50 Hz
2.0 ALaboratoryActive
  • (< 250 A to 500 A) @ 50 Hz
3.5 ALaboratoryActive
  • (< 500 A to 750 A) @ 50 Hz
5.0 ALaboratoryActive
  • (< 500 A to 1000 A) @ 50 Hz
6.4 ALaboratoryActive
  • Resistance 2 Wire
  • Direct Method Using Multi-product Calibrator according to TCAL Calibration Procedures TCAL-WI-28 & 30
  • < 0.1 Ω to 1 Ω
7 mΩLaboratoryActive
  • < 1 Ω to 10 Ω
13 mΩLaboratoryActive
  • < 10 Ω to 100 Ω
18 mΩLaboratoryActive
  • < 100 Ω to 1 kΩ
0.14 ΩLaboratoryActive
  • < 1 kΩ to 10 kΩ
1.4 ΩLaboratoryActive
  • < 10 kΩ to 100 kΩ
14 ΩLaboratoryActive
  • < 100 kΩ to 1 MΩ
0.2 kΩLaboratoryActive
  • < 1 MΩ to 10 MΩ
11 kΩLaboratoryActive
  • < 10 MΩ to 100 MΩ
0.96 MΩLaboratoryActive
  • Direct Method Using Multi product Calibrator
  • < 1 Ω to 10 Ω
13 mΩLaboratoryActive
  • < 10 Ω to 100 Ω
18 mΩLaboratoryActive
  • < 100 Ω to 1 kΩ
0.14 ΩLaboratoryActive
  • < 1 kΩ to 10 kΩ
1.4 ΩLaboratoryActive
  • < 10 kΩ to 100 kΩ
14 ΩLaboratoryActive
  • Insulation Tester Calibration Box @ 5000 V
  • 10 MΩ
0.12 MΩLaboratoryActive
  • 100 MΩ
1.3 MΩLaboratoryActive
  • 1 GΩ
22 MΩLaboratoryActive
  • Capacitance
  • Direct Method Using Multi-product Calibrator according to TCAL Calibration Procedures TCAL-WI-28 & 30
  • < 1 nF to 10 nF
0.05 nFLaboratoryActive
  • < 10 nF to 20 nF
0.14 nFLaboratoryActive
  • < 20 nF to 50 nF
0.33 nFLaboratoryActive
  • < 50 nF to 100 nF
0.65 nFLaboratoryActive
  • < 100 nF to 1 µF
9.7 nFLaboratoryActive
  • 10 Hz to 100 Hz
3.3 mHzLaboratoryActive
  • < 100 Hz to 1 kHz
33 mHzLaboratoryActive
  • < 1 kHz to 10 kHz
0.33 HzLaboratoryActive
  • < 10 kHz to 50 kHz
1.7 HzLaboratoryActive
  • < 50 kHz to 100 kHz
3.3 HzLaboratoryActive
  • < 100 kHz to 1 MHz
30 HzLaboratoryActive
Temperature
  • Direct Reading Thermometers a) Digital Thermometer with PRT Sensor b) Temperature Transmitter with PRT Sensor
  • In-house methods TCAL-WI-05 & TCAL-WI-06 based on DKD-R-5-1
  • -30℃ to 200℃
0.27℃Laboratory / Customer PremisesActive
  • > 200℃ to 400℃
0.29℃Laboratory / Customer PremisesActive
  • > 400℃ to 600℃
0.41℃Laboratory / Customer PremisesActive
  • Direct Reading Thermometers a) Digital Thermometer with thermocouple Sensor, Type K, J, N b) Transmitter with Thermocouple Sensor, type K, J, N c) Bi-Metal & Dial Thermometer
  • In-house methods TCAL-WI-05 & TCAL-WI-06 based on EURAMET CG No.8
  • -30℃ to 200℃
0.3℃Laboratory / Customer PremisesActive
  • > 200℃ to 400℃
0.4℃Laboratory / Customer PremisesActive
  • > 400℃ to 600℃
0.5℃Laboratory / Customer PremisesActive
  • Temperature Indicators & Controllers for PRTs
  • Electrical simulation method TCAL-WI-07 based on EURAMET cg-11
  • -30℃ to 200℃
0.22℃Laboratory / Customer PremisesActive
  • 200℃ to 400℃
0.25℃Laboratory / Customer PremisesActive
  • 400℃ to 600℃
0.28℃Laboratory / Customer PremisesActive
  • Temperature Indicators & Controllers for Thermocouples Types J, K and N
  • -30℃ to 200℃
0.3℃Laboratory / Customer PremisesActive
  • > 200℃ to 600℃
0.4℃Laboratory / Customer PremisesActive
  • Temperature Chart Recorders Connected to Thermocouples
  • In-house methods TCAL-WI-05 based on DKD-R-5-1
  • -30℃ to 600℃
1.2℃Laboratory / Customer PremisesActive
  • Chiller / Freezer / Refrigerator
  • In-house methods TCAL-WI-10 & TCAL-WI-26 based on DKD-R-5-1
  • -25℃ to 50℃
0.9℃Customer PremisesActive
  • Drying Oven / Hot Cabinet
  • > 50℃ to 300℃
1.5℃Customer PremisesActive
  • Chiller / Freezer / Refrigerator / Drying Oven / Hot Cabinet / Incubator
  • -25℃ to 50℃
0.6℃Customer PremisesActive
  • > 50 ℃ to 400℃
1.3℃Customer PremisesActive
  • Muffle Furnace
  • > 400℃ to 900℃
3.5℃Customer PremisesActive