| Measured Quantity / Calibration Instrument |
Calibration Method |
Range and Specification |
Expanded Measurement |
Location |
Status |
| Balance |
| - TCAL-WI-11 using Standard Weights based on Euramet cg 18
| | 0.86 mg | Customers Premises | Active
|
| Pressure |
- Gauge Pressure: Medium: Pneumatic
| - TCAL-WI-01 by Comparison method using Digital Pressure Gauge as per DKD-R 6-1
| | 5.8 kPa | Laboratory and Customer Premises | Active
|
- Gauge Pressure: Medium: Hydraulic
| | 0.50 kPa | Laboratory and Customer Premises | Active
|
- Gauge Pressure Pneumatic Analog and Digital Pressure Gauges
| - TCAL-WI-01 by Comparison method based on requirement of DKD-R 6-1 (03/2014)
| | 2.9 kPa | Laboratory and Customer Premises | Active
|
- Gauge Pressure Hydraulic Pressure Transmitters
| - TCAL-WI-02 by Comparison method based on requirement of DKD-R 6-1 (03/2014)
| | 42 kPa | Laboratory and Customer Premises | Active
|
| Electrical |
| - Direct Method Using Multi-product Calibrator according to TCAL Calibration Procedures TCAL-WI-28 & 30
| | 0.02 mV | Laboratory | Active
|
| | 0.05 µA | Laboratory | Active
|
| - Up to 200 mV @ 40 Hz to 10 kHz
| 0.12 mV | Laboratory | Active
|
| | 0.51 µA | Laboratory | Active
|
| | 7 mΩ | Laboratory | Active
|
| | 0.05 nF | Laboratory | Active
|
| Temperature |
- Direct Reading Thermometers a) Digital Thermometer with PRT Sensor b) Temperature Transmitter with PRT Sensor
| - In-house methods TCALWI-05 & TCAL-WI-06 based on DKD-R-5-1
| | 0.27℃ | Laboratory / Customer Premises | Active
|
- Direct Reading Thermometers a) Digital Thermometer with thermocouple Sensor, Type K, J, N b) Transmitter with Thermocouple Sensor, type K, J, N c) Bi-Metal & Dial Thermometer
| - In-house methods TCALWI-05 & TCAL-WI-06 based on EURAMET CG No.8
| 0.3℃ | Laboratory / Customer Premises | Active
|
- Temperature Indicators & Controllers for PRTs
| - Electrical simulation method TCAL-WI-07 based on EURAMET cg11
| 0.22℃ | Laboratory / Customer Premises | Active
|
- Temperature Indicators & Controllers for Thermocouples Types J, K and N
| 0.3℃ | Laboratory / Customer Premises | Active
|
- Temperature Chart Recorders Connected to Thermocouples
| - In-house method TCAL-WI-05 based on DKD-R5-1
| | 1.2℃ | Laboratory / Customer Premises | Active
|
- Climatic Chambers (9 Point Method) for Chiller / Freezer / Refrigerator / Incubator
| - In-house methods TCALWI-10 & TCAL-WI-26 based on DKD-R 5-7
| | 0.9 ℃ | Customer Premises | Active
|
- Climatic Chambers (9 Point Method) for Drying Oven / Hot Cabinet
| | | | | Active
|
- Climatic Chambers (Single Point Method) for Chiller / Freezer / Refrigerator / Incubator
| | | Active
|
- Climatic Chambers (Single Point Method) for Drying Oven / Hot Cabinet
| - In-house methods TCALWI-10 & TCAL-WI-26 based on DKD-R 5-7
| | 1.5 ℃ | Customer Premises | Active
|
- Muffle Furnace (Single Point Method)
| | 3.5 ℃ | Customer Premises | Active
|
- Wideband Infrared Thermometers
| - In-house method TCAL-WI-36 based on ASTM E2847-21
| | 1.6 oC | Laboratory | Active
|