Almoayara Althakyah laboratory (SmartCal)

Initial Accreditation Date:Nov 27 , 2014

معلومات الاتصال

العنوان

Sharjah, United Arab Emirates

صندوق بريد / الرمز البريدي

82356

البريد الإلكتروني

info@smartcaluae.com

الموقع الإلكتروني

www.smartcaluae.com

المجال الرئيسي

Calibration Laboratories

ساري منذ:Feb 06 , 2024

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Temperature
  • Resistance thermometers with display unit
  • SCP-05 Comparison with platinum resistance thermometer in bath
  • -30 to 200°C
0.04 °CLaboratoryActive
  • >200 °C to 250 °C
0.09 °CLaboratoryActive
  • SCP-05 Comparison with platinum resistance thermometer in a dry block calibrator
  • -90 °C to -45 °C
0.15 °CLaboratory/ Customer PremisesActive
  • -45 °C to 100 °C
0.15 °CLaboratory/ Customer PremisesActive
  • >100 °C to 150 °C
0.20 °CLaboratory/ Customer PremisesActive
  • Thermocouple thermometers with display unit
  • SCP-05 Comparison with platinum resistance thermometer in bath
  • -30 °C to 100 °C
0.30 °CLaboratoryActive
  • >100 °C to 200 °C
0.50 °CLaboratoryActive
  • >200 °C to 250 °C
0.55 °CLaboratoryActive
  • SCP-05 Comparison with platinum resistance thermometer in a dry block calibrator
  • -90 °C to -45 °C
0.30 °CLaboratory/ Customer PremisesActive
  • -45 °C to 100 °C
0.40 °CLaboratory/ Customer PremisesActive
  • >100 °C to 150 °C
0.60 °CLaboratory/ Customer PremisesActive
  • Liquid in glass thermomer / Glass thermometer
  • SCP-28 Comparison with platinum resistance thermometer in bath
  • -30 °C to 100 °C
0.20 °CLaboratoryActive
  • 100 °C to 200 °C
0.30 °CLaboratoryActive
  • Thermo-hygrometer
  • SCP-23 Comparison with Relative Humidity Sensor in Climatic Chamber
  • 30% to 95%
1.5% rHLaboratoryActive
  • 10 °C to 45 °C
±0.23 °CLaboratoryActive
  • Temperature Chambers (Chillers, Refrigerators, Climatic Chambers, Ovens) includes chambers above 2 m³ of volume
  • SCP- 04 DKD R-5-7 Comparison with platinum resistance thermometers /Data loggers. Temperature profile measurement 9 points.
  • -80 °C to -40 °C
1.0 °CLaboratory/ Customer PremisesActive
  • -40 °C to 200 °C
0.60 °CLaboratory/ Customer PremisesActive
  • Incubators
  • SCP- 04 DKD R-5-7 Comparison with platinum resistance thermometers / Data loggers. 5 to 9 points
  • -40 °C to 150 °C
0.25 °CLaboratory/ Customer PremisesActive
  • Liquid baths
  • SCP- 04 Comparison with platinum resistance thermometers /Data loggers in bath Temperature profile measurement at 5 to 9 points
  • 5 °C to 90 °C
0.40 °CLaboratory/ Customer PremisesActive
  • Autoclaves
  • SCP-20 Comparison with platinum resistance thermometers /Data loggers inside autoclaveTemperature profile measurement at 1 to 5 points
  • 110 °C to 140 °C
0.60 °CLaboratory/ Customer PremisesActive
  • Non contact thermometers (IR thermometers / Ear thermometers / Pyrometer / thermal imager etc.)
  • SCP-32 based on E287-14
  • -30 °C to -15 °C
1.0 °CLaboratory/ Customer PremisesActive
  • > -15 °C to 150 °C
0.80 °CLaboratory/ Customer PremisesActive
  • Temperature Controlled surface plates / stations: (Thermal cycler/ Hot plate/ Slide warmer/ slide cooler/ PCR system / Reaction station/ Paraffin embedding Station/ Tissue processor/ Hybridization oven/ Block Heater/ immunology analyzer/ Coagulation analyzer/ Thermobrite etc. )
  • SCP- 33 Performance study using Data loggers
  • -30 °C to 200 °C
0.45 °CLaboratory/ Customer PremisesActive
  • High temperature chambers / oven / muffle funaces
  • SCP-21
  • >200 °C to 1000 °C
3.0 °CLaboratory/ Customer PremisesActive
  • Temperature Data Logger
  • SCP-05 Comparison with platinum resistance thermometer in Climatic Chamber
  • "-45 °C to -20 °C
±0.7LaboratoryActive
  • ">-20 °C to 0 °C
±0.5LaboratoryActive
  • >0°C to 45 °C
±0.2LaboratoryActive
  • >45 °C to 100 °C
±0.3LaboratoryActive
  • >100 °C to 150 °C
±0.5LaboratoryActive
  • >150 °C to 180 °C
±0.7LaboratoryActive
Mass and Balance
  • Mass/ Electronic Balances
  • Use of Mass Standards SCP-11 according to EURAMET cg18-v4, using by E2, F1, M1 masses
  • 1 mg to 100 mg
2.2 E-4Customer PrimasesActive
  • >100 mg to 1 g
4.0 E-5Customer PrimasesActive
  • >1 g to 5 g
1.3 E-5Customer PrimasesActive
  • >5 g to 60 g
2.6 E-6Customer PrimasesActive
  • >60 g to 600 g
2.0 E-6Customer PrimasesActive
  • >600 g to 6 kg
6.7 E-6Customer PrimasesActive
  • >6 kg to 320 kg
8.7 E-5Customer PrimasesActive
Volume
  • Liquid Volume/ Piston operated volumetric apparatus (Pipettes/ Dispensers /Repeaters)
  • Gravimetric method SCP-06 according to ISO 8655-6:2022
  • 0.5 µL to 20 µL
0.05 µlLaboratoryActive
  • >20 µL to 100 µL
0.2 µlLaboratoryActive
  • >100 µL to 500 µL
0.5 µlLaboratoryActive
  • >500 µL to 1000 µL
1 µlLaboratoryActive
  • >1 mL to 2 mL
2 µlLaboratoryActive
  • >2 mL to 5ml
4 µlLaboratoryActive
  • >5 ml to 10 mL
8 µlLaboratoryActive
  • Liquid Volume/ Piston operated volumetric apparatus (Pipettes)
  • Photometric method SCP-08 according to ISO 8655-7:2022
  • 0.1 μL to 5 µl
1.5% of readingLaboratoryActive
  • >5 μL to 5000 µl
1.2% of readingLaboratoryActive
  • Liquid Volume/ Laboratory Glassware (Bulb pipettes, Graduated pipettes)
  • Gravimetric method SCP-29 according to ISO 4787:2021
  • 0.1 ml to 5 ml
0.004 mlLaboratoryActive
  • >5 ml to 10 ml
0.008 mlLaboratoryActive
  • >10 ml to 50 ml
0.012 mlLaboratoryActive
  • >50 ml to 100 ml
0.02 mlLaboratoryActive
  • Liquid Volume/ Laboratory Glassware (Burrettes)
  • 1 ml to 10 ml
0.005 mlLaboratoryActive
  • >10 ml to 100 ml
0.03 mlLaboratoryActive
  • Liquid Volume/ Laboratory Glassware (Flasks)
  • 1 ml to 25 ml
0.02 mlLaboratoryActive
  • >25 ml to 100 ml
0.05 mlLaboratoryActive
  • >100 ml to 500 ml
0.18 mlLaboratoryActive
  • >500 ml to 2000 ml
0.26 mlLaboratoryActive
  • >2000 ml to 5000 ml
0.35 mlLaboratoryActive
  • Liquid Volume/ Laboratory Glassware (Graduated cylinders)
  • Gravimetric method SCP-29 according to ISO 4787:2021
  • 1 ml to 5 ml
0.04 mlLaboratoryActive
  • >5 ml to 20 ml
0.08 mlLaboratoryActive
  • >20 ml to 100 ml
0.12 mlLaboratoryActive
  • >100 ml to 250 ml
0.35 mlLaboratoryActive
  • >250 ml to 500 ml
0.8 mlLaboratoryActive
  • >500 ml to 1000 ml
1.2 mlLaboratoryActive
  • >1000 ml to 2000 ml
2.6 mlLaboratoryActive
  • Liquid Volume/ Laboratory Glassware (Beakers)
  • 1 ml to 5 ml
0.3 mlLaboratoryActive
  • >5 ml to 10 ml
0.5 mlLaboratoryActive
  • >10 ml to 25 ml
0.7 mlLaboratoryActive
  • >25 ml to 50 ml
1 mlLaboratoryActive
  • >50 ml to 100 ml
1.5 mlLaboratoryActive
  • Gravimetric method SCP-29 according to ISO 4787:2021
  • >100 ml to 250 ml
3 mlLaboratoryActive
  • >250 ml to 500 ml
5 mlLaboratoryActive
  • >500 ml to 1000 ml
7 mlLaboratoryActive
  • >1000 ml to 2000 ml
10 mlLaboratoryActive
  • >2000 ml to 5000 ml
20 mlLaboratoryActive
  • Liquid volume / Piston operated volumetric apparatus (Burettes, Diluters)
  • Gravimetric method SCP-06 according to ISO 8655-6:2022
  • 10 ml to 100ml
0.03 % rdg.LaboratoryActive
  • Liquid volume / Piston operated volumetric apparatus (Single stroke dispensers)
0.3 % rdg.LaboratoryActive
  • >100 ml to 200ml
0.1 % rdg.LaboratoryActive
Time & Speed
  • Rotational Speed Tachometer
  • Comparison with Reference tachometer SCP-26
  • Up to 50 rpm
0.1 rpmLaboratoryActive
  • 50 < range ≤ 200 rpm
0.13 rpmLaboratoryActive
  • 200 < range ≤ 1000 rpm
1.6 rpmLaboratoryActive
  • 1000 < range≤ 10000rpm
1.7 rpmLaboratoryActive
  • 10000 < range≤ 90000rpm
5.4 rpmLaboratoryActive
  • Rotational Speed eg. Centrifuge , mixers , orbital shaker . Plate shaker etc
  • Comparison with calibrated tachometer SCP-25
  • Up to 50 rpm
0.1 rpmLaboratory/ Customer PremisesActive
  • 50 < range ≤ 200 rpm
0.13 rpmLaboratory/ Customer PremisesActive
  • 200 < range ≤ 1000 rpm
1.6 rpmLaboratory/ Customer PremisesActive
  • 1000 < range≤ 10000rpm
1.7 rpmLaboratory/ Customer PremisesActive
  • 10000 < range≤ 90000rpm
5.4 rpmLaboratory/ Customer PremisesActive
  • Time Calibration - (Stop watches/ Timers ; and Time indication in instruments like centrifuges/ speed machines /Thermoclock / Autoclaves/ PCR system or other such machines .etc)
  • Comparison with calibrated stopwatch SCP- 34
  • 0 to 30 min
0.35 SecLaboratory/ Customer PremisesActive
  • Comparison with calibrated stopwatch SCP- 34
  • Above 30 min to 2 hr
1.6 SecLaboratory/ Customer PremisesActive