Abhath Weights and Measuring Laboratory

Initial Accreditation Date:Sep 22 , 2013

معلومات الاتصال

العنوان

Dubai, United Arab Emirates

صندوق بريد / الرمز البريدي

48402

البريد الإلكتروني

nf@abhathuae.com

الموقع الإلكتروني

www.abhathuae.com

المجال الرئيسي

Calibration Laboratories

ساري منذ:Oct 06 , 2025

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Temperature
  • Liquid in Glass Thermometer
  • Comparison Method by using Digital Temperature Indicator with Standard RTD Procedure WI/ABHATH-T04
  • ­10 °C to 100 °C
0.15 °CLaboratoryActive
  • Direct reading thermometers, temperature gauges, dial thermometers, with a resistance or a thermocouple sensor
  • Comparison Method by using Standard RTD with Digital Temperature Indicator Procedure WI/ABHATH-T04
  • ­10 °C to 420 °C
0.35 °CPermanent Laboratory & Customer PremisesActive
  • Comparison Method by using Standard Thermocouple with Digital Temperature Indicator Procedure WI/ABHATH-T11
  • 420 °C to 600 °C
4.9 °CPermanent Laboratory & Customer PremisesActive
  • Dry block temperature Calibrator
  • Comparison Method by using Standard RTD or Standard Thermocouple with Digital Temperature Indicator Procedure WI/ABHATH-T06
  • ­10 °C to 420 °C
0.26 °CPermanent Laboratory & Customer PremisesActive
  • 420 °C to 600 °C
4.7 °CPermanent Laboratory & Customer PremisesActive
  • Temperature Gauge and dial thermometers
  • Comparison Method by using Standard RTD with Digital Temperature Indicator Procedure WI/ABHATH-T04
  • ­10 °C to 420 °C
0.15 °CPermanent Laboratory & Customer PremisesActive
  • Comparison Method by using Standard Thermocouple with Digital Temperature Indicator Procedure WI/ABHATH-T11
  • 420 °C to 600 °C
4.9 °CPermanent Laboratory & Customer PremisesActive
  • Temperature liquid bath
  • Multipoint comparison Method by using Standard RTD/Standard Thermocouple with Digital Temperature Indicator Procedure WI/ABHATH-T02
  • ­10 °C to 200 °C
0.88 °CPermanent Laboratory & Customer PremisesActive
  • Chiller / Freezer / Refrigerator (9 point calibration)
  • Comparison Method by using Digital Temperature Indicator with Standard RTD Procedure WI/ABHATH-T03
  • ­25 °C to 15 °C
0.5 °CCustomer PremisesActive
  • Oven / Furnaces (9 point calibration)
  • Comparison Method by using Standard RTD or Standard Thermocouple with Digital Temperature Indicator Procedure WI/ABHATH-T03
  • 20 °C to 200 °C
0.5 °CCustomer PremisesActive
  • Incubator (9 point calibration)
  • Comparison Method by using Digital Temperature Indicator with Standard RTD Procedure WI/ABHATH-T03
  • 10 °C to 100 °C
0.5 °CCustomer PremisesActive
  • Infrared Thermometers and Thermal Imagers
  • Comparison Method by using infrared calibrator Procedure WI/ABHATH-T08
  • 50 °C to 150 °C
2.9 °CLaboratoryActive
  • 150 °C to 500 °C
4.5 °CLaboratoryActive
  • Thermo-Hygrometers for relative humidity with Indicator or Transmitter; Relative humidity Data Loggers
  • Comparison Method with standard hygrometers using a humidity generator Air temperature: 20 °C to 30 °C Procedure WI/ABHATH-T09
  • 20 % rH to 90 % rH
3.1 % rHLaboratoryActive
  • Comparison Method with standard hygrometers using a humidity generator Air temperature: 30 °C to 50 °C Procedure WI/ABHATH-T09
3.5 % rHLaboratoryActive
  • Thermo-Hygrometers for air temperature with Indicator or Transmitter; Air temperature Data Loggers
  • Comparison Method with standard thermometers in a humidity generator or climatic chamber Procedure WI/ABHATH-T09
  • 20 °C to 50 °C
1.2 °CLaboratoryActive
  • measuring point in Chiller / Freezer / Refrigerator (single point calibration)
  • Comparison Method by using Digital Temperature Indicator with Standard RTD Procedure WI/ABHATH-T05
  • -25 °C to 15 °C
0.5 °CCustomer PremisesActive
  • measuring point in Oven / Furnaces (single point calibration)
  • Comparison Method by using Standard RTD or Standard Thermocouple with Digital Temperature Indicator Procedure WI/ABHATH-T05
  • 20 °C to 200 °C
0.65 °CCustomer PremisesActive
  • measuring point in Incubator (single point calibration)
  • Comparison Method by using Digital Temperature Indicator with Standard RTD Procedure WI/ABHATH-T05
  • 10 °C to 100 °C
0.65 °CCustomer PremisesActive
  • Dish washers,Glass Washer
  • Comparison Method by using standard dish washer thermometer Procedure WI/ABHATH-T12
  • 20 °C to 100 °C
0.7 °CCustomer PremisesActive
Balance
  • Mass/ Electronic Balance
  • Using F1 Class Weight WI/ABHTH-M01 in accordance to: EURAMET cg-18/4.0
  • Up to 35 kg
0.18 g but not less than 1 d of the balanceCustomers PremisesActive
  • Using M1 Class Weight WI/ABHTH-M01 in accordance to: EURAMET cg-18/4.0
  • > 35 kg to 1000 kg
0.19 kgCustomer PremisesActive
Electrical-Source
  • DC Voltage
  • Direct Method Using 8846 A according to Calibration Procedure WI/ABHATH-E03
  • 100 mV to 10 V
0.14%LaboratoryActive
  • > 10 V to 100 V
0.01%LaboratoryActive
  • > 100 V to 1000 V
0.01%LaboratoryActive
  • DC High Voltage
  • In-house Method Using 8846 A according to Calibration Procedure WI/ABHATH-E03
  • 1 kV to 40 kV
5.80%LaboratoryActive
  • AC Voltage (50 Hz to 1 kHz)
  • Direct Method Using 8846 A according to Calibration Procedure WI/ABHATH-E03
  • 100 mV to 10 V
0.75%LaboratoryActive
  • > 10 V to 1000 V
1.20%LaboratoryActive
  • AC Voltage(@50 Hz)
  • In-house Method Using 8846 A according to Calibration Procedure WI/ABHATH-E03
  • 1 kV to 20 kV
6.90%LaboratoryActive
  • DC Current
  • Direct Method Using 8846 A according to Calibration Procedure WI/ABHATH-E03
  • 100 µA to 100 mA
0.59%LaboratoryActive
  • > 0.1 A to 1 A
0.49%LaboratoryActive
  • > 1 A to 10 A
0.20%LaboratoryActive
  • AC Current (50 Hz to 1kHz)
  • Direct Method Using 8846 A according to Calibration Procedure WI/ABHATH-E03
  • 100 µA to 100 mA
2.60%LaboratoryActive
  • > 0.1 A to 1 A
0.82%LaboratoryActive
  • > 1 A to 10 A
0.82%LaboratoryActive
  • Resistance(2-Wire)
  • Direct Method Using 8846 A according to Calibration Procedure WI/ABHATH-E02
  • 10Ω to 100 Ω
0.36%LaboratoryActive
  • > 100 Ω to 100 KΩ
0.15%LaboratoryActive
  • > 100 kΩ to 100 MΩ
0.94%LaboratoryActive
  • > 100 MΩ to 1 GΩ
2.48%LaboratoryActive
  • Resistance(4-Wire)
  • 10Ω to 100 Ω
0.36%LaboratoryActive
  • > 100 Ω to 100 KΩ
0.02%LaboratoryActive
  • > 100 kΩ to 10 MΩ
0.10%LaboratoryActive
  • Capacitance
  • 1nF to 100 nF
7.70%LaboratoryActive
  • > 100 nF to 100 µF
1.90%LaboratoryActive
  • > 100 μF to 100 mF
5.00%LaboratoryActive
  • Temperature RTD
  • -200 ºC to 600 ºC
0.26%LaboratoryActive
  • Frequency
  • > 5 Hz to 1 MHz
0.11%LaboratoryActive
Electrical-Measuring
  • DC Voltage
  • Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E01
  • 1mV to 320 mV
0.67%LaboratoryActive
  • > 0.32 V to 32 V
0.02%LaboratoryActive
  • > 32 V to 320 V
0.03%LaboratoryActive
  • > 320 V to 1000 V
0.03%LaboratoryActive
  • AC Voltage (10 Hz to 3 kHz)
  • 10 mV to 320 mV
1.16%LaboratoryActive
  • > 0.32 V to 30 V
1.20%LaboratoryActive
  • AC Voltage (40 Hz to 3 kHz)
  • > 30 V to 300 V
1.10%LaboratoryActive
  • > 300 V to 1000 V
1.20%LaboratoryActive
  • DC Current
  • 10 µA to 300 µA
1.9%LaboratoryActive
  • > 0.3 mA to 300 mA
0.15%LaboratoryActive
  • > 0.3 A to 3 A
0.15%LaboratoryActive
  • > 3 A to 10 A
0.15%LaboratoryActive
  • Using 10 turns & 50 turns Coil (WI/ABHATH-E05)
  • > 10 A to 1000 A
1.50%LaboratoryActive
  • AC Current (10 Hz to 3 kHz)
  • Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E01
  • 10 µA to 300 µA
3.6%LaboratoryActive
  • > 0.3 mA to 300 mA
0.59%LaboratoryActive
  • > 0.3 A to 3 A
0.59%LaboratoryActive
  • > 3 A to 10 A
0.41%LaboratoryActive
  • AC Current (@ 50 Hz)
  • Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E05
  • > 10 A to 1000 A
0,76% 1.9%LaboratoryActive
  • Resistance (2 Wire)
  • Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E01
  • 1 Ω to 40 Ω
1.2%LaboratoryActive
  • > 40 Ω to 400 Ω
0.08%LaboratoryActive
  • > 0.4 kΩ to 400 kΩ
0.05%LaboratoryActive
  • > 0.4 MΩ to 40 MΩ
0.08%LaboratoryActive
  • > 40 MΩ to 400 MΩ
1.00%LaboratoryActive
  • Resistance (4 Wire)
  • 1Ω to 40 Ω
1.2%LaboratoryActive
  • > 40 Ω to 400 Ω
0.08%LaboratoryActive
  • > 0.4 kΩ to 400 kΩ
0.05%LaboratoryActive
  • > 0.4 MΩ to 30 MΩ
0.60%LaboratoryActive
  • Capacitance
  • 1nF to 4nF
6.10%LaboratoryActive
  • > 4 nF to 400 nF
2.00%LaboratoryActive
  • > 0.4 μF to 40 μF
1.90%LaboratoryActive
  • > 40 μF to 400 μF
1.70%LaboratoryActive
  • > 0.4 mF to 40 mF
4.6%LaboratoryActive
  • Insulation Resistance (50 V to 5 kV)
  • Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E04 Fixed point
  • 10 MΩ
1.30%LaboratoryActive
  • 100 MΩ
1.30%LaboratoryActive
  • 1 GΩ
1.2%LaboratoryActive
  • 10 GΩ
1.2%LaboratoryActive
  • AC Power (@ 50 Hz)
  • Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E07 (Using 10 X and 50 X Current Coil)
  • 10 W to to 5 kW up to 500 V up to 10 A
0.30%LaboratoryActive
  • > 5 kW to 350 Kw > 500 V to 700 V > 10 A to 500 A
0.60%LaboratoryActive
  • Temperature Simulation RTD
  • Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E06
  • -200 ºC to 800 ºC
0.44 ºCLaboratoryActive
  • Temperature Simulation J-type
  • -200 ºC to 1200 ºC
0.50 ºCLaboratoryActive
  • Temperature Simulation K-type
  • -200 ºC to 1300 ºC
0.51 ºCLaboratoryActive
  • Frequency
  • Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E01
  • > 10 Hz to 1 MHz
0.01%LaboratoryActive
Pressure
  • Gas Pressure Gauge, Pressure Indicating and electrical output devices
  • Comparison Method by using Digital Pressure Gauge with Pneumatic comparatorWI/ABHATH-P01 in accordance to: DKD-R 6-1:2013
  • (- 90 kPa to 200 kPa)
1.2 kPaLaboratory/ Customer PremisesActive
  • >200 k Pa to 3.5 MPa
10 kPaLaboratory/ Customer PremisesActive
  • Liquid Pressure Gauge, Pressure Indicating and electrical output devices
  • WI/ABHATH-P02 Based on DKD-R-6-1: 2013
  • Up to 6 Mpa
19 kPaLaboratory/ Customer PremisesActive
  • >6 MPa to 70 MPa
34 kPaLaboratory/ Customer PremisesActive
  • >70 MPa to 100 MPa
40 kPaLaboratory/ Customer PremisesActive
  • >100 MPa to 120 MPa
64 kPaLaboratory/ Customer PremisesActive
  • 2013WI/ABHATH-P02 Based on DKD-R-6-1: 2013
  • Up to 70 Mpa
34 kPaCustomer PremisesActive
  • >70 MPa to 100 MPa
40 kPaCustomer PremisesActive