| Measured Quantity / Calibration Instrument |
Calibration Method |
Range and Specification |
Expanded Measurement |
Location |
Status |
| Balance & Scale |
- Top loading weighing scale
| | | | | Active
|
| Temperature |
- Liquid in Glass Thermometer
| - Comparison Method by using Digital Temperature Indicator with Standard RTD Procedure WI/ABHATH-T04
| | 0.15 °C | Laboratory | Active
|
- Direct reading thermometers, temperature gauges, dial thermometers, with a resistance or a thermocouple sensor
| - Comparison Method by using Standard RTD with Digital Temperature Indicator Procedure WI/ABHATH-T04
| | 0.35 °C | Permanent Laboratory & Customer Premises | Active
|
- Comparison Method by using Standard Thermocouple with Digital Temperature Indicator Procedure WI/ABHATH-T11
| | 4.9 °C | Permanent Laboratory & Customer Premises | Active
|
- Dry block temperature Calibrator
| - Comparison Method by using Standard RTD or Standard Thermocouple with Digital Temperature Indicator Procedure WI/ABHATH-T06
| | 0.26 °C | Permanent Laboratory & Customer Premises | Active
|
| 4.7 °C | Permanent Laboratory & Customer Premises | Active
|
- Temperature Gauge and dial thermometers
| - Comparison Method by using Standard RTD with Digital Temperature Indicator Procedure WI/ABHATH-T04
| | 0.15 °C | Permanent Laboratory & Customer Premises | Active
|
- Comparison Method by using Standard Thermocouple with Digital Temperature Indicator Procedure WI/ABHATH-T11
| | 4.9 °C | Permanent Laboratory & Customer Premises | Active
|
| - Multipoint comparison Method by using Standard RTD/Standard Thermocouple with Digital Temperature Indicator Procedure WI/ABHATH-T02
| | 0.88 °C | Permanent Laboratory & Customer Premises | Active
|
- Chiller / Freezer / Refrigerator (9 point calibration)
| - Comparison Method by using Digital Temperature Indicator with Standard RTD Procedure WI/ABHATH-T03
| | 0.5 °C | Customer Premises | Active
|
- Oven / Furnaces (9 point calibration)
| - Comparison Method by using Standard RTD or Standard Thermocouple with Digital Temperature Indicator Procedure WI/ABHATH-T03
| | 0.5 °C | Customer Premises | Active
|
- Incubator (9 point calibration)
| - Comparison Method by using Digital Temperature Indicator with Standard RTD Procedure WI/ABHATH-T03
| | 0.5 °C | Customer Premises | Active
|
- Infrared Thermometers and Thermal Imagers
| - Comparison Method by using infrared calibrator Procedure WI/ABHATH-T08
| | 2.9 °C | Laboratory | Active
|
| 4.5 °C | Laboratory | Active
|
- Thermo-Hygrometers for relative humidity with Indicator or Transmitter; Relative humidity Data Loggers
| - Comparison Method with standard hygrometers using a humidity generator Air temperature: 20 °C to 30 °C Procedure WI/ABHATH-T09
| | 3.1 % rH | Laboratory | Active
|
- Comparison Method with standard hygrometers using a humidity generator Air temperature: 30 °C to 50 °C Procedure WI/ABHATH-T09
| 3.5 % rH | Laboratory | Active
|
- Thermo-Hygrometers for air temperature with Indicator or Transmitter; Air temperature Data Loggers
| - Comparison Method with standard thermometers in a humidity generator or climatic chamber Procedure WI/ABHATH-T09
| | 1.2 °C | Laboratory | Active
|
- measuring point in Chiller / Freezer / Refrigerator (single point calibration)
| - Comparison Method by using Digital Temperature Indicator with Standard RTD Procedure WI/ABHATH-T05
| | 0.5 °C | Customer Premises | Active
|
- measuring point in Oven / Furnaces (single point calibration)
| - Comparison Method by using Standard RTD or Standard Thermocouple with Digital Temperature Indicator Procedure WI/ABHATH-T05
| | 0.65 °C | Customer Premises | Active
|
- measuring point in Incubator (single point calibration)
| - Comparison Method by using Digital Temperature Indicator with Standard RTD Procedure WI/ABHATH-T05
| | 0.65 °C | Customer Premises | Active
|
- Dish washers,Glass Washer
| - Comparison Method by using standard dish washer thermometer Procedure WI/ABHATH-T12
| | 0.7 °C | Customer Premises | Active
|
| Balance |
| - Using F1 Class Weight WI/ABHTH-M01 in accordance to: EURAMET cg-18/4.0
| | 0.18 g but not less than 1 d of the balance | Customers Premises | Active
|
- Using M1 Class Weight WI/ABHTH-M01 in accordance to: EURAMET cg-18/4.0
| | 0.19 kg | Customer Premises | Active
|
| Electrical-Source |
| - Direct Method Using 8846 A according to Calibration Procedure WI/ABHATH-E03
| | 0.14% | Laboratory | Active
|
| 0.01% | Laboratory | Active
|
| 0.01% | Laboratory | Active
|
| - In-house Method Using 8846 A according to Calibration Procedure WI/ABHATH-E03
| | 5.80% | Laboratory | Active
|
- AC Voltage (50 Hz to 1 kHz)
| - Direct Method Using 8846 A according to Calibration Procedure WI/ABHATH-E03
| | 0.75% | Laboratory | Active
|
| 1.20% | Laboratory | Active
|
| - In-house Method Using 8846 A according to Calibration Procedure WI/ABHATH-E03
| | 6.90% | Laboratory | Active
|
| - Direct Method Using 8846 A according to Calibration Procedure WI/ABHATH-E03
| | 0.59% | Laboratory | Active
|
| 0.49% | Laboratory | Active
|
| 0.20% | Laboratory | Active
|
- AC Current (50 Hz to 1kHz)
| - Direct Method Using 8846 A according to Calibration Procedure WI/ABHATH-E03
| | 2.60% | Laboratory | Active
|
| 0.82% | Laboratory | Active
|
| 0.82% | Laboratory | Active
|
| - Direct Method Using 8846 A according to Calibration Procedure WI/ABHATH-E02
| | 0.36% | Laboratory | Active
|
| 0.15% | Laboratory | Active
|
| 0.94% | Laboratory | Active
|
| 2.48% | Laboratory | Active
|
| | 0.36% | Laboratory | Active
|
| 0.02% | Laboratory | Active
|
| 0.10% | Laboratory | Active
|
| | 7.70% | Laboratory | Active
|
| 1.90% | Laboratory | Active
|
| 5.00% | Laboratory | Active
|
| | 0.26% | Laboratory | Active
|
| | 0.11% | Laboratory | Active
|
| Electrical-Measuring |
| - Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E01
| | 0.67% | Laboratory | Active
|
| 0.02% | Laboratory | Active
|
| 0.03% | Laboratory | Active
|
| 0.03% | Laboratory | Active
|
- AC Voltage (10 Hz to 3 kHz)
| | 1.16% | Laboratory | Active
|
| 1.20% | Laboratory | Active
|
- AC Voltage (40 Hz to 3 kHz)
| | 1.10% | Laboratory | Active
|
| 1.20% | Laboratory | Active
|
| | 1.9% | Laboratory | Active
|
| 0.15% | Laboratory | Active
|
| 0.15% | Laboratory | Active
|
| 0.15% | Laboratory | Active
|
- Using 10 turns & 50 turns Coil (WI/ABHATH-E05)
| | 1.50% | Laboratory | Active
|
- AC Current (10 Hz to 3 kHz)
| - Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E01
| | 3.6% | Laboratory | Active
|
| 0.59% | Laboratory | Active
|
| 0.59% | Laboratory | Active
|
| 0.41% | Laboratory | Active
|
| - Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E05
| | 0,76% 1.9% | Laboratory | Active
|
| - Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E01
| | 1.2% | Laboratory | Active
|
| 0.08% | Laboratory | Active
|
| 0.05% | Laboratory | Active
|
| 0.08% | Laboratory | Active
|
| 1.00% | Laboratory | Active
|
| | 1.2% | Laboratory | Active
|
| 0.08% | Laboratory | Active
|
| 0.05% | Laboratory | Active
|
| 0.60% | Laboratory | Active
|
| | 6.10% | Laboratory | Active
|
| 2.00% | Laboratory | Active
|
| 1.90% | Laboratory | Active
|
| 1.70% | Laboratory | Active
|
| 4.6% | Laboratory | Active
|
- Insulation Resistance (50 V to 5 kV)
| - Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E04 Fixed point
| | 1.30% | Laboratory | Active
|
| 1.30% | Laboratory | Active
|
| 1.2% | Laboratory | Active
|
| 1.2% | Laboratory | Active
|
| - Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E07 (Using 10 X and 50 X Current Coil)
| - 10 W to to 5 kW up to 500 V up to 10 A
| 0.30% | Laboratory | Active
|
- > 5 kW to 350 Kw > 500 V to 700 V > 10 A to 500 A
| 0.60% | Laboratory | Active
|
- Temperature Simulation RTD
| - Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E06
| | 0.44 ºC | Laboratory | Active
|
- Temperature Simulation J-type
| | 0.50 ºC | Laboratory | Active
|
- Temperature Simulation K-type
| | 0.51 ºC | Laboratory | Active
|
| - Direct Method Using 9100/3041A according to Calibration Procedure WI/ABHATH-E01
| | 0.01% | Laboratory | Active
|
| Pressure |
- Gas Pressure Gauge, Pressure Indicating and electrical output devices
| - Comparison Method by using Digital Pressure Gauge with Pneumatic comparatorWI/ABHATH-P01 in accordance to: DKD-R 6-1:2013
| | 1.2 kPa | Laboratory/ Customer Premises | Active
|
| 10 kPa | Laboratory/ Customer Premises | Active
|
- Liquid Pressure Gauge, Pressure Indicating and electrical output devices
| - WI/ABHATH-P02 Based on DKD-R-6-1: 2013
| | 19 kPa | Laboratory/ Customer Premises | Active
|
| 34 kPa | Laboratory/ Customer Premises | Active
|
| 40 kPa | Laboratory/ Customer Premises | Active
|
| 64 kPa | Laboratory/ Customer Premises | Active
|
- 2013WI/ABHATH-P02 Based on DKD-R-6-1: 2013
| | 34 kPa | Customer Premises | Active
|
| 40 kPa | Customer Premises | Active
|