Omega Measuring & Calibration Laboratory

Initial Accreditation Date:Jul 01 , 2014

معلومات الاتصال

العنوان

Dubai, United Arab Emirates

صندوق بريد / الرمز البريدي

P.O Box No : 125349

البريد الإلكتروني

sbem@emirates.net.ae

الموقع الإلكتروني

NA

المجال الرئيسي

Calibration Laboratories

ساري منذ:Aug 22 , 2023

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Temperature
  • Resistance Thermometers with Display Unit / Temperature Transmitters with Resistance Sensor / Resistance Thermometers
  • Comparison with platinum resistance thermometer in a bath. DKD R5-1
  • ­ 30 °C up to 150 °C
0.1 °CLaboratory/ Customer PremisesActive
  • 150 °C up to 250 °C
0.2 °CLaboratory/ Customer PremisesActive
  • Comparison with platinum resistance thermometer in a block calibrator DKD R5-1
  • ­30 °C up to 150 °C
0.2 °CLaboratory/ Customer PremisesActive
  • 150 °C up to 400 °C
0.4 °CLaboratory/ Customer PremisesActive
  • 400 °C up to 600 °C
0.6 oCLaboratory/ Customer PremisesActive
  • Thermocouples with Display Unit / Temperature Transmitters with Thermocouple Sensor / Thermocouples
  • Comparison with a thermocouple thermometer in a block calibrator
  • ­ 30 °C up to 150 °C
0.6 °CLaboratory/ Customer PremisesActive
  • 150 °C up to 400 °C
2.0 °CLaboratory/ Customer PremisesActive
  • 400 °C up to 600 °C
2.5 oCLaboratory/ Customer PremisesActive
  • 600 °C up to 1200 °C
3.0 oCLaboratory/ Customer PremisesActive
  • 600 °C to 900 °C
8.0 °CLaboratory/ Customer PremisesActive
  • 900 °C to 1200 °C
12.0 °CLaboratory/ Customer PremisesActive
  • Liquid in Glass Thermometers
  • Comparison with platinum resistance thermometer in bath
  • ­ 30 °C up to 150 °C
0.4 °CLaboratoyActive
  • 150 °C up to 250 °C
0.6 °CLaboratoyActive
  • -30 °C to 150 °C
0.4 °CLaboratoyActive
  • 150 °C to 250 °C
0.6 °CLaboratoyActive
  • Oil Bath, Dry Block Furnace
  • Euramet cg-13
  • 30°C to 250°C
0.2°CLaboratoyActive
  • 250°C to 400°C
0.4°CLaboratoyActive
  • 250°C to 600°C
0.6°CLaboratoyActive
  • 600°C to 900°C
8,0°CLaboratoyActive
  • 900°C to 1200°C
12,0°CLaboratoyActive
  • Measuring location in Oven (Single Point)
  • DKD-R 5-7 method C
  • 0°C to 250°C
0,2°CLaboratory/ Customer PremisesActive
  • Oven (Multi Points)
  • DKD-R 5-7 method A or B
1,9°CLaboratory/ Customer PremisesActive
  • Measuring location in Furnace (Single Point)
  • DKD-R 5-7 method C
  • 250°C to 600°C
0,7°CLaboratory/ Customer PremisesActive
  • 600°C to 900°C
8,0°CLaboratory/ Customer PremisesActive
  • 900°C to 1200°C
12,0°CLaboratory/ Customer PremisesActive
  • Measuring location in Freezer, Chiller, Incubator, Cold Room, Environment Chamber ( Single Point )
  • -30°C to 80°C
0,2°CLaboratory/ Customer PremisesActive
  • Freezer, Chiller, Incubator, Cold Room, Environment Chamber ( Multi Points )
  • DKD-R 5-7 method A or B
  • -30°C to 30°C
0,5°CLaboratory/ Customer PremisesActive
  • 30°C to 80°C
1,4°CLaboratory/ Customer PremisesActive
  • Measuring location in Humidity Chamber, Environment Chamber ( Single Point )
  • DKD-R 5-7 method C At air temperature 5°C to 85°C
  • 20%RH to 95%RH
1,9%RHLaboratory/ Customer PremisesActive
  • Humidity Chamber, Environment Chamber ( Multi Points)
  • DKD-R 5-7 method A or B At air temperature: 5°C to 85°C
3,2%RHLaboratory/ Customer PremisesActive
  • Temp Data logger (Inbuild Sensor)
  • BS:1041-2.2, BS:1041-4
  • -30°C to 80°C
0,2°CLaboratoryActive
  • Humidity Data logger, Hygrometer (Inbuild Sensor)
  • BS:1041-2.2, BS:1041-4 At air temperature: 5°C to 50°C
  • 20%RH to 95%RH
3,0%RHLaboratoryActive
Pressure
  • Gas Pressure (negative gauge)/ Digital and analogue indicating devices
  • WI 10 acc.to DKD R 6-1 (03-2014)
  • -0.9 bar to 0 bar
20 mbarLaboratory/ Customer PremisesActive
  • Gas Pressure (gauge)/ Digital and analogue indicating devices
  • > 0 bar to 20 bar
0.20 barLaboratory/ Customer PremisesActive
  • Liquid Pressure (gauge)/ Digital and analogue indicating devices
  • 0 bar to 350 bar
0.20 barLaboratory/ Customer PremisesActive
  • > 350 bar to 700 bar
0.25 barLaboratory/ Customer PremisesActive
  • > 700 bar to 2500 bar
3.0 barLaboratory/ Customer PremisesActive
Dimensions
  • All types of Caliper Digital / Dial / Vernier
  • In-House method WI-22 based on BS EN ISO 13385 For determining error of indicated size Comparison with reference gauge blocks
  • 0 mm up to 150 mm
15 µmLaboratoryActive
  • External Micrometer
  • In-House method WI-07 based on BS EN ISO 3611 For determining error of indicated size Comparison with reference gauge blocks
  • 0 mm up to 25 mm
3.3 µmLaboratoryActive
  • Dial/Digital Indicators
  • In-House method WI-03 Based on BS EN ISO 463 For determining error of indicated displacement, Comparison with dial gauge calibrator
6.7 µmLaboratoryActive
  • Height Gauge
  • In-House method Based on BS EN ISO 31225 For determining error of indicated vertical size Comparison with reference gauge blocks or calliper checker
  • 0 mm up to 300 mm
18 µmLaboratoryActive
Electrical-Calibrators
  • DC Voltage
  • EURAMET- cg-15 Version 3.0
  • ≥ 1 mV to 100 mV
10 µVLaboratory/ Customer PremisesActive
  • > 100 mV to 1 V
0.14 mVLaboratory/ Customer PremisesActive
  • > 1 V to 10 V
0.71 mVLaboratory/ Customer PremisesActive
  • > 10 V to 100 V
6.0 mVLaboratory/ Customer PremisesActive
  • > 100 mV to 1000 V
0.38 VLaboratory/ Customer PremisesActive
  • AC Voltage
  • EURAMET- cg-15 Version 3.0 ¦ = 50 Hz
  • ≥ 1 mV to 100 mV
0.12 mVLaboratory/ Customer PremisesActive
  • > 100 mV to 1 V
1.0 mVLaboratory/ Customer PremisesActive
  • > 1 V to 10 V
10 mVLaboratory/ Customer PremisesActive
  • > 10 V to 100 V
0.10 VLaboratory/ Customer PremisesActive
  • > 100 V to 1000 V
1.2 VLaboratory/ Customer PremisesActive
  • Resistance
  • EURAMET- cg-15 Version 3.0
  • ≥ 1 Ω to 100 Ω
35 mΩLaboratory/ Customer PremisesActive
  • > 100 Ω to 1 kΩ
0.18 ΩLaboratory/ Customer PremisesActive
  • > 1 kΩ to 10 kΩ
1.4 ΩLaboratory/ Customer PremisesActive
  • > 10 kΩ to 100 kΩ
12 ΩLaboratory/ Customer PremisesActive
  • > 100 kΩ to 1 MΩ
0.19 kΩLaboratory/ Customer PremisesActive
  • > 1MΩ to 10 MΩ
4.8 kΩLaboratory/ Customer PremisesActive
  • > 10 M Ω to 100 MΩ
96 kΩLaboratory/ Customer PremisesActive
  • DC Current
  • ≥ 1 mA to 10 mA
30 µALaboratory/ Customer PremisesActive
  • > 10 mA to 100 mA
70 µALaboratory/ Customer PremisesActive
  • > 100 mA to 1 A
1.3 mALaboratory/ Customer PremisesActive
  • > 1 A to 3 A
5.0 mALaboratory/ Customer PremisesActive
  • AC Current
  • EURAMET- cg-15 Version 3.0 ¦ = 50 Hz
  • ≥ 100 mA to 1 A
3.8 mALaboratory/ Customer PremisesActive
  • > 1 A to 3 A
7.3 mALaboratory/ Customer PremisesActive
Electrical-Instruments
  • DC Voltage
  • EURAMET- cg-15 Version 3.0
  • ≥ 1 mV to 200 mV
14 µVlaboratory/ Customer PremisesActive
  • > 200 mV to 2 V
72 µVlaboratory/ Customer PremisesActive
  • > 2 V to 20 V
0.60 mVlaboratory/ Customer PremisesActive
  • > 20 V to 200 V
14 mVlaboratory/ Customer PremisesActive
  • > 200 V to 1000 V
92 mVlaboratory/ Customer PremisesActive
  • > 20 µA to 2 mA
0.29 µAlaboratory/ Customer PremisesActive
  • > 2 mA to 20 mA
2.2 µAlaboratory/ Customer PremisesActive
  • > 20 mA to 200 mA
30 µAlaboratory/ Customer PremisesActive
  • > 200 mA to 2 A
0.64 mAlaboratory/ Customer PremisesActive
  • > 2 A to 20 A
16 mAlaboratory/ Customer PremisesActive
  • > 20 A to 100 A
0.58 Alaboratory/ Customer PremisesActive
  • > 100 A to 1000 A
5.8 Alaboratory/ Customer PremisesActive
  • Resistance
  • Up to 1 mΩ
0.23 mΩlaboratory/ Customer PremisesActive
  • > 1 mΩ to 10 mΩ
0.23 mΩlaboratory/ Customer PremisesActive
  • > 10 mΩ to 100 mΩ
0.31 mΩlaboratory/ Customer PremisesActive
  • > 100 mΩ to 1 Ω
0.35 mΩlaboratory/ Customer PremisesActive
  • > 1 Ω to 10 Ω
1.4 mΩlaboratory/ Customer PremisesActive
  • > 10 Ω to 100 Ω
59 mΩlaboratory/ Customer PremisesActive
  • > 100 Ω to 1 kΩ
0.13 Ωlaboratory/ Customer PremisesActive
  • > 1 kΩ to 10 kΩ
0.31 Ωlaboratory/ Customer PremisesActive
  • > 10 kΩ to 100 kΩ
4.0 Ωlaboratory/ Customer PremisesActive
  • > 100 kΩ to 1 MΩ
0.17 kΩlaboratory/ Customer PremisesActive
  • > 1 MΩ to 10 MΩ
12 kΩlaboratory/ Customer PremisesActive
  • > 10 MΩ to 100 MΩ
1.2 MΩlaboratory/ Customer PremisesActive
  • > 100 MΩ to 1 GΩ
0.12 GΩlaboratory/ Customer PremisesActive
  • AC Voltage
  • EURAMET- cg-15 Version 3.0 ¦=60 Hz
  • ≥ 2 mV to 200 mV
0.21 mVlaboratory/ Customer PremisesActive
  • > 200 mV to 2 V
0.90 mVlaboratory/ Customer PremisesActive
  • > 2 V to 20 V
9.5 mVlaboratory/ Customer PremisesActive
  • > 20 V to 200 V
0.16 Vlaboratory/ Customer PremisesActive
  • > 200 V to 1000 V
1.0 Vlaboratory/ Customer PremisesActive
  • AC Current
  • ≥ 20 µA to 2 mA
1.6 µAlaboratory/ Customer PremisesActive
  • > 2 mA to 20 mA
20 µAlaboratory/ Customer PremisesActive
  • > 20 mA to 200 mA
0.15 mAlaboratory/ Customer PremisesActive
  • > 200 mA to 2 A
3.0 mALaboratory/ Customer PremisesActive
  • > 2 A to 20 A
52 mALaboratory/ Customer PremisesActive
  • > 20 A to 100 A
0.58 ALaboratory/ Customer PremisesActive
  • > 100 A to 1000 A
5.8 ALaboratory/ Customer PremisesActive
  • Temperature Simulation Method, RTD (Pt-100)
  • ITS-90 EURAMET- cg-11
  • ­ 200 °Ϲ to 800 °Ϲ
0.6 °ϹLaboratory/ Customer PremisesActive
Balance
  • Mass/ Electronic Balances
  • WI-23 based on EURAMET cg18:2015
  • Up to 210 g
0.0005 gCustomer PremisesActive
  • Up to 1 kg
0.011 gCustomer PremisesActive
  • Up to 30 kg
0.6 gCustomer PremisesActive
  • Up to 60 kg
1.5 gCustomer PremisesActive
  • Up to 100 kg
6 gCustomer PremisesActive
  • Up to 300 kg
0.1 kgCustomer PremisesActive
  • Up to 1000 kg
0.18 kgCustomer PremisesActive