Al Taif Technical Services Company L.L.C

Initial Accreditation Date:Nov 01 , 2015

معلومات الاتصال

العنوان

Abu Dhabi, United Arab Emirates

صندوق بريد / الرمز البريدي

43337

البريد الإلكتروني

Elias.Cruz@altaif.ae

الموقع الإلكتروني

www.altaif.ae

المجال الرئيسي

Calibration Laboratories

ساري منذ:Sep 23 , 2024

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Temperature
  • Wideband Infrared Thermometers
  • Infrared Thermometer Calibration Work Instruction RDE – WI 12
  • 100 °C to 200 °C
1.5 °CLaboratoryActive
  • >200°C to 350 °C
2.5 °CLaboratoryActive
  • >350 °C to 500 °C
3.5 °CLaboratoryActive
Electrical
  • DC Voltage
  • Direct Comparison with Multifunction Calibrator Fluke 5520A method METCAL, ATI - CAL04 U = measured voltage value
  • 0 V to 330 mV
21x10-6 U +1.4 μVLaboratoryActive
  • >330 mV to 3.3 V
11x10-6 U + 2.0 μVLaboratoryActive
  • >3.3 V to 33 V
12x10-6 U + 16 μVLaboratoryActive
  • >33 V to 330 V
16x10-6 U + 0.13 mVLaboratoryActive
  • >330 V to 1000 V
16x10-6 U + 1.3 mVLaboratoryActive
  • AC Voltage
  • (1.0 mV to 33 mV) 50 Hz to 10 kHz
0.41x10-3 U + 4.6 μVLaboratoryActive
  • (1.0 mV to 33 mV) >10 kHz to 50 kHz
0.42x10-3 U + 4.6 μVLaboratoryActive
  • (>33 mV to 330 mV) 50 Hz to 10 kHz
0.14x10-3 U + 6.1 μVLaboratoryActive
  • (>33 mV to 330 mV) >10 kHz to 50 kHz
0.15x10-3 U + 6.1 μVLaboratoryActive
  • (>330 mV to 3.3 V) 50 Hz to 10 kHz
0.13 x 10¯³ U + 46 µVLaboratoryActive
  • (>330 mV to 3.3 V) >10 kHz to 50 kHz
0.16 x 10¯³ U + 46 µVLaboratoryActive
  • (>3.3 V to 33 V) 50 Hz to 10 kHz
0.13 x 10¯³ U + 0.46 mVLaboratoryActive
  • (>3.3 V to 33 V) >10 kHz to 50 kHz
0.20 x 10¯³ U + 0.46 mVLaboratoryActive
  • (>33 V to 330 V) 50 Hz to 10 kHz
0.16 x 10¯³ U + 1.5 mVLaboratoryActive
  • (>33 V to 330 V) >10 kHz to 50 kHz
0.20 x 10¯³ U + 4.6 mVLaboratoryActive
  • (>330 V to 1000 V) 50 Hz to 10 kHz
0.24 x 10¯³ U + 7.7 mVLaboratoryActive
  • DC Current
  • Direct Comparison with Multifunction Calibrator Fluke 5520A method METCAL, ATI - CAL04 I = measured current value
  • 0 µA to 330 µA
0.13 x 10¯³ I + 0.02 µALaboratoryActive
  • >330 µA to 3.3 mA
80 x 10¯⁶ I + 0.04 µALaboratoryActive
  • >3.3 mA to 33 mA
80 81 x 10¯⁶ I + 0.19 µALaboratoryActive
  • >33 mA to 330 mA
81 80 x 10¯⁶ I + 1.9 µALaboratoryActive
  • >330 mA to 3 A
0.16 x 10¯³ I + 31 µALaboratoryActive
  • >3 A to 11 A
0.39 x 10¯³ I + 0.39 mALaboratoryActive
  • >11 A to 20 A
0.39 x 10¯³ I + 0.38 mALaboratoryActive
  • AC Current
  • Direct Comparison with Multifunction Calibrator Fluke 5520A method METCAL, ATI - CAL04 I = measured current value
  • 0 29 µA to 330 µA
1.1 x 10¯³ I + 0.08 µALaboratoryActive
  • 50 Hz to 10 kHz
1.1 x 10¯³ I + 0.08 µALaboratoryActive
  • >330 µA to 3.3 mA
0.86 x 10¯³ I + 0.12 µALaboratoryActive
  • 50 Hz to 10 kHz
0.86 x 10¯³ I + 0.12 µALaboratoryActive
  • >3.3 mA to 33 mA
0.36 x 10¯³ I + 1.5 µALaboratoryActive
  • 50 Hz to 10 kHz
0.36 x 10¯³ I + 1.5 µALaboratoryActive
  • >33 mA to 330 mA
0.42 x 10¯³ I + 15 µALaboratoryActive
  • 50 Hz to 10 kHz
0.42 x 10¯³ I + 15 µALaboratoryActive
  • >330 mA to 3 A
0.48 x 10¯³ I + 77 µALaboratoryActive
  • 50 Hz to 1 kHz
0.48 x 10¯³ I + 77 µALaboratoryActive
  • >3 A to 11 A
1.6 x 10¯³ I + 1.5 mALaboratoryActive
  • 50 Hz to 1 kHz
1.6 x 10¯³ I + 1.5 mALaboratoryActive
  • >11 A to 20 A
0.87 x 10¯³ I + 1.5 mALaboratoryActive
  • 50 Hz to 1 kHz
0.87 x 10¯³ I + 1.5 mALaboratoryActive
  • Resistance
  • Direct Comparison with Multifunction Calibrator Fluke 5520A method METCAL, ATI - CAL04 R = measured resistance value
  • 0 Ω to 11 Ω
66 x 10¯⁶ R + 6.9 mΩLaboratoryActive
  • >11 Ω to 33 Ω
67 x 10¯⁶ R + 6.9 mΩLaboratoryActive
  • >33 Ω to 110 Ω
26 x 10¯⁶ R + 7.4 mΩLaboratoryActive
  • >110 Ω to 330 Ω
26 x 10¯⁶ R + 7.4 mΩLaboratoryActive
  • >330 Ω to 1.1 kΩ
23 x 10¯⁶ R + 8.1 mΩLaboratoryActive
  • >1.1 kΩ to 3.3 kΩ
23 x 10¯⁶ R + 0.08 ΩLaboratoryActive
  • >3.3 kΩ to 11 kΩ
23 x 10¯⁶ R + 0.08 ΩLaboratoryActive
  • >11 kΩ to 33 kΩ
23 x 10¯⁶ R + 0.81 ΩLaboratoryActive
  • >33 kΩ to 110 kΩ
23 x 10¯⁶ R + 0.81 ΩLaboratoryActive
  • >110 kΩ to 330 kΩ
23 x 10¯⁶ R + 0.81 ΩLaboratoryActive
  • >330 kΩ to 1.1 MΩ
29 x 10¯⁶ R + 8.1 ΩLaboratoryActive
  • >1.1 MΩ to 3.3 MΩ
29 x 10¯⁶ R + 8.1 ΩLaboratoryActive
  • >3.3 MΩ to 11 MΩ
0.10 x 10¯³ R + 0.08 kΩLaboratoryActive
  • >11 MΩ to 33 MΩ
0.10 x 10¯³ R + 0.08 kΩLaboratoryActive
  • >33 MΩ to 110 MΩ
0.42 x 10¯³ R + 4.3 kΩLaboratoryActive
  • >110 MΩ to 330 MΩ
0.41 x 10¯³ R + 4.3 kΩLaboratoryActive
Mass and Balance
  • Mass/ Electronic Balances
  • RD-WI-08 Based on the EURAMET cg 18/v.04:2015
  • 0 to 100 g
1.3 mgClient PremisesActive
  • < 100 to 1000 g
24 mgClient PremisesActive
Pressure
  • Pressure/ Gauge, hydraulic
  • RD-WI-113 Based on the EURAMET cg 17/v.03:2017 and DKD-R 6-1:2014
  • 0 & 2 - 100 MPa
8 X 0-5 * pe + 1.2 kPaLaboratoryActive
Torque
  • Torque/ Hand torque tools
  • RD-WI-09 Based on ISO 6789:2017 pt1 and pt2
  • 1 N.m up to 10 N.m
0.0061LaboratoryActive
  • 10 N.m up to 100 N.m
0.0071LaboratoryActive
  • 110 N.m up to 1100 N.m
0.0067LaboratoryActive
  • 300 N.m up to 3000 N.m
0.0092LaboratoryActive
Dimensions
  • All types of Caliper Digital / Dial / Vernier
  • ATI - CAL 02 Based on JIS B 7507
  • 0-200 mm
15 µmLaboratoryActive
  • >200 mm up to 600 mm
20 µmLaboratoryActive
  • Calibration of External Micrometers (Digital-Analogue
  • ATI - CAL 05 Based on JIS B 7502
  • 0-25 mm
2.4 µmLaboratoryActive
  • Calibration of Dial Gauge (Digital/Analogue)
  • ATI - ECAL 03 Based on JIS B 7503
4.7 µmLaboratoryActive