Zenner Calibration Laboratory

Initial Accreditation Date:Feb 02 , 2020

معلومات الاتصال

العنوان

Dubai, United Arab Emirates

صندوق بريد / الرمز البريدي

62880

البريد الإلكتروني

quality@zennerservices.com

الموقع الإلكتروني

www.zennerservices.com

المجال الرئيسي

Calibration Laboratories

ساري منذ:Sep 01 , 2023

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Electrical
  • DC Current
  • Based on the calibration procedure (ZCL-CP-01) Using 5025 C- MultiFunction Calibrator
  • Up to 200 µA
18 nALaboratoryActive
  • > 200 µA to 2 mA
1.6 µALaboratoryActive
  • > 2 mA to 20 mA
15 µALaboratoryActive
  • > 20 mA to 200 mA
0.18 mALaboratoryActive
  • > 200 mA to 2 A
0.19 mALaboratoryActive
  • > 2 A to 20 A
2.3 mALaboratoryActive
  • AC Current @ 50 Hz
  • Up to 200 µA
0.33 µALaboratoryActive
  • > 200 µA to 2 mA
0.95 µALaboratoryActive
  • > 2 mA to 20 mA
25 µALaboratoryActive
  • > 20 mA to 200 mA
0.11 mALaboratoryActive
  • > 200 mA to 2 A
81 µALaboratoryActive
  • > 2 A to 20 A
36 mALaboratoryActive
  • AC Current @ 1 kHz
  • Up to 200 µA
0.33 µALaboratoryActive
  • > 200 µA to 2 mA
0.95 µALaboratoryActive
  • > 2 mA to 20 mA
25 µALaboratoryActive
  • > 20 mA to 200 mA
0.11 mALaboratoryActive
  • > 200 mA to 2 A
81 µALaboratoryActive
  • > 2 A to 20 A
36 mALaboratoryActive
  • DC Voltage
  • Up to 20 mV
5.2 µVLaboratoryActive
  • > 20 mV to 200 mV
16 µVLaboratoryActive
  • > 200 mV to 2 V
1.8 mVLaboratoryActive
  • > 2 V to 20 V
2.4 mVLaboratoryActive
  • > 20 V to 200 V
11 mVLaboratoryActive
  • > 200 V to 1000 V
27 mVLaboratoryActive
  • AC Voltage @ 50 Hz
  • Based on the calibration procedure (ZCL-CP-01) Using 5025 C- MultiFuncation Calibrator
  • Up to 20 mV
12 µvLaboratoryActive
  • > 20 mV to 200 mV
24 µVLaboratoryActive
  • > 200 mV to 2 V
0.1 mVLaboratoryActive
  • > 2 V to 20 V
7.6 mVLaboratoryActive
  • > 20 V to 200 V
78 mVLaboratoryActive
  • > 200 V to 1000 V
0.88 VLaboratoryActive
  • AC Voltage @1 kHz
  • Up to 20 mV
19 µVLaboratoryActive
  • > 20 mV to 200 mV
84 µVLaboratoryActive
  • > 200 mV to 2 V
0.4 mVLaboratoryActive
  • > 2 V to 20 V
2.8 mVLaboratoryActive
  • > 20 V to 200 V
28 mVLaboratoryActive
  • > 200 V to 1000 V
0.19 VLaboratoryActive
  • Fixed Resistance 4-W
  • 10 Ω
2.5 mΩLaboratoryActive
  • 100 Ω
5.0 mΩLaboratoryActive
  • 1 KΩ
15 mΩLaboratoryActive
  • 10 KΩ
0.2 ΩLaboratoryActive
  • 100 KΩ
3.0 ΩLaboratoryActive
  • Fixed Resistance 2-W
  • 1 MΩ
0.1 KΩLaboratoryActive
  • 10 MΩ
0.01 MΩLaboratoryActive
  • 100 MΩ
0.88 MΩLaboratoryActive
  • 1 GΩ
0.03 GΩLaboratoryActive
  • Thermocouple "K" Type
  • -200 °C
0.6 °CLaboratoryActive
  • 0 °C
0.6 °CLaboratoryActive
  • 500 °C
0.6 °CLaboratoryActive
  • 1190 °C
0.6 °CLaboratoryActive
  • Thermocouple "J" Type
  • -200 °C
0.6 °CLaboratoryActive
  • 500 °C
0.6 °CLaboratoryActive
  • 1190 °C
0.6 °CLaboratoryActive
  • Thermocouple "T" Type
  • -200 °C
0.6 °CLaboratoryActive
  • 0 °C
0.6 °CLaboratoryActive
  • 390 °C
0.7 °CLaboratoryActive
  • Thermocouple "R" Type
  • -50 °C
0.9 °CLaboratoryActive
  • 100 °C
0.9 °CLaboratoryActive
  • 1750 °C
0.9 °CLaboratoryActive
  • Thermocouple "S" Type
  • -50 °C
0.9 °CLaboratoryActive
  • 100 °C
0.9 °CLaboratoryActive
  • 1750 °C
0.9 °CLaboratoryActive
  • Thermocouple "B" Type
  • 300 °C
0.6 °CLaboratoryActive
  • 1000 °C
0.6 °CLaboratoryActive
  • 1800 °C
0.7 °CLaboratoryActive
  • Thermocouple "E" Type
  • -200 °C
0.6 °CLaboratoryActive
  • 0 °C
0.6 °CLaboratoryActive
  • 1000 °C
0.6 °CLaboratoryActive
  • Thermocouple "N" Type
  • -200 °C
0.6 °CLaboratoryActive
  • 100 °C
0.6 °CLaboratoryActive
  • 1295 °C
0.6 °CLaboratoryActive
  • Capacitance @ 1 kHz
  • 1 nF
1.4 pFLaboratoryActive
  • 10 nF
14 pFLaboratoryActive
  • 20 nF
74 pFLaboratoryActive
  • 50 nF
0.09 nFLaboratoryActive
  • 100 nF
0.17 nFLaboratoryActive
  • 200 nF
0.76 nFLaboratoryActive
  • 500 nF
1.1 nFLaboratoryActive
  • 1 µF
1.2 nFLaboratoryActive
  • Frequency
  • >20 Hz to 200 Hz
2.4 mHzLaboratoryActive
  • > 1 KHz to 100 KHz
2.3 HzLaboratoryActive
  • > 500 KHz to 1000 KHz
2.5 HzLaboratoryActive
  • AC Current @ 50 Hz
  • Based on the calibration procedure (ZCL-CP-02) Using 5025 C- MultiFuncation Calibrator & Current coil
  • > 10 A to 20 A
0.3 ALaboratoryActive
  • > 20 A to 100 A
0.7 ALaboratoryActive
  • > 100 A to 250 A
1.8 ALaboratoryActive
  • > 250 A to 500 A
3.9 ALaboratoryActive
  • > 500 A to 750 A
5.7 ALaboratoryActive
  • > 750 A to 1000 A
7.6 ALaboratoryActive
  • DC Current
  • > 10 A to 20 A
0.3 ALaboratoryActive
  • > 20 A to 100 A
0.6 ALaboratoryActive
  • > 100 A to 250 A
1.7 ALaboratoryActive
  • > 250 A to 500 A
3.3 ALaboratoryActive
  • > 500 A to 750 A
4.8 ALaboratoryActive
  • > 750 A to 1000 A
6.4 ALaboratoryActive
  • Based on the calibration procedure (ZCL-CP-02) Using 8846A, 6.5" Digital Multimeter
  • Up to 100 µA
0.1 µALaboratoryActive
  • > 100 µA to 1 mA
0.7 µALaboratoryActive
  • > 1 mA to 10 mA
8.9 µALaboratoryActive
  • > 10 mA to 100 mA
69 µALaboratoryActive
  • > 100 mA to 400 mA
0.3 mALaboratoryActive
  • > 400 mA to 1 A
0.8 mALaboratoryActive
  • > 1 A to 3 A
4.3 mALaboratoryActive
  • > 3 A to 10 A
19 mALaboratoryActive
  • AC Current @ 50 Hz
  • Up to 100 µA
0.5 µALaboratoryActive
  • > 100 µA to 1 mA
3.1 µALaboratoryActive
  • > 1 mA to 10 mA
27 µALaboratoryActive
  • > 10 mA to 100 mA
0.2 mALaboratoryActive
  • > 100 mA to 400 mA
0.9 mALaboratoryActive
  • > 400 mA to 1 A
1.7 mALaboratoryActive
  • > 1 A to 3 A
7.4 mALaboratoryActive
  • > 3 A to 10 A
25 mALaboratoryActive
  • DC Voltage
  • Up to 100 mV
13 µVLaboratoryActive
  • > 100 mV to 1 V
1.9 mVLaboratoryActive
  • > 1 V to 10 V
1.9 mVLaboratoryActive
  • > 10 V to 100 V
7.0 mVLaboratoryActive
  • > 100 V to 1000 V
61 mVLaboratoryActive
  • AC Voltage @ 50 Hz
  • Up to 100 mV
0.1 mVLaboratoryActive
  • > 100 mV to 1 V
1.1 mVLaboratoryActive
  • > 1 V to 10 V
11 mVLaboratoryActive
  • > 10 V to 100 V
0.12 VLaboratoryActive
  • > 100 V to 1000 V
1.0 VLaboratoryActive
  • Resistance 4-W
  • 10 Ω
5.0 mΩLaboratoryActive
  • 100 Ω
16 mΩLaboratoryActive
  • 1 KΩ
0.1 ΩLaboratoryActive
  • 10 KΩ
1.3 ΩLaboratoryActive
  • 100 KΩ
13 ΩLaboratoryActive
  • Resistance 2-W
  • 1 MΩ
0.2 kΩLaboratoryActive
  • 10 MΩ
5.0 kΩLaboratoryActive
  • 100 MΩ
0.9 MΩLaboratoryActive
  • 1000 MΩ
26 MΩLaboratoryActive
  • Capacitance
  • 1 nF
0.07 nFLaboratoryActive
  • 10 nF
0.19 nFLaboratoryActive
  • 100 nF
1.9 nFLaboratoryActive
  • 1 µF
0.02 µFLaboratoryActive
  • 10 µF
0.04 µFLaboratoryActive
  • 100 µF
0.6 µFLaboratoryActive
  • 1 mF
0.02 mFLaboratoryActive
  • 10 mF
0.04 mFLaboratoryActive
  • 100 mF
0.9 mFLaboratoryActive
  • Frequency
  • 1 MHz
9.0 HzLaboratoryActive
Temperature
  • Direct Reading Thermometers, Digital Thermometers, Pen Type Thermometer, Temperature Indicator / Datalogger with Sensor, Transmitters, Transducers, Controllers with Sensors, Chart Recorders.
  • ZCL-CP-05
  • 0 °C up to  200 °C
0.65 °CLaboratory & Customer PremisesActive
  • 300 °C up to 500° C
1.32 °CLaboratory & Customer PremisesActive
  • Liquid baths (5 sensors)
  • ZCL-CP-06
  • -30 up to 300 °C
0.5 °CLaboratory & Customer PremisesActive
  • Dry Block Calibrators
  • 40 °C up to 500° C
0.75 °CLaboratory & Customer PremisesActive
  • PRT / RTD/ Thermocouples (B,J,K,R,S,T,N)without display unit
  • ZCL-CP-07
  • 0 °C up to 150° C
0.41 °CLaboratory & Customer PremisesActive
  • >150 °C to 500 °C
1.1 °CLaboratory & Customer PremisesActive
  • Environmental Chamber / Climatic Chamber / Temperature & Humidity Chambers ( Single Point/ Three Point / Nine Point)
  • ZCL-CP-08
  • 20°C to 60°C & 10%rH to 90% rH
0.7 °C & 4 %rHCustomer PremisesActive
  • Freezer, Chiller, Refrigerator, Hot Oven, Hot Cabinet, Hot Plates / Boxes, Warmers, Auto Clave, Dishwasher, Glass Washer. ( Single Point to Nine Point)
  • -30°C to 500°C
0.75 °CCustomer PremisesActive
  • Furnace
  • 550 °C to 1100 °C
7 °CCustomer PremisesActive
  • Temperature & Humidity Meter
  • ZCL-CP-09
  • 10°C to 60°C & 20%rH to 90% rH
0.7 °C , 3.5 %rHLaboratory PremisesActive
  • Infrared / Laser Thermometer, Ear Gun Thermometer, Thermal Imager
  • ZCL-CP-10
  • 50 °C up to 500 °C
3.5 °CLaboratory PremisesActive
Pressure
  • Vacuum Pressure Gauge, Chart Recorders, Monitors, Indicating, Transmitters and Transducers.
  • DKD-R6-1-2014
  • -1 bar to 0 bar
0.04 barLaboratory & Customer PremisesActive
  • Pneumatic Pressure Gauge, Chart Recorders, Monitors, Indicating, Transmitters and Transducers.
  • 0 bar to 40 bar
0.024 barLaboratory & Customer PremisesActive
  • Hydraulic Pressure Gauge, Chart Recorders, Monitors, Indicating, Transmitters and Transducers.
  • 0 bar to 700 bar
0.26 barLaboratory & Customer PremisesActive
  • > 700 bar to 2000 bar
1.5 barLaboratory & Customer PremisesActive