Metrology Laboratories Section- Dubai Central Laboratory Department (Dubai Municipality)

Initial Accreditation Date:May 30 , 2007

معلومات الاتصال

العنوان

Dubai, United Arab Emirates

صندوق بريد / الرمز البريدي

67

البريد الإلكتروني

AAABDULSALAM@dm.gov.ae

الموقع الإلكتروني

www.dm.gov.ae

المجال الرئيسي

Calibration Laboratories

ساري منذ:Oct 31 , 2025

Measured Quantity / Calibration Instrument Calibration Method Range and Specification Expanded Measurement Location Status
Balance & Scale
  • Top loading weighing scale
Active
Temp, Humidity and Autoclave
  • Resistance thermometers and direct reading thermometers with RTD
  • DMS 2003 (comparison method using calibrated PRTs, Thermocouples, Data Loggers , liquid fluidized baths, reference ovens and fixed ice point
  • -80 °C up to -40 °C
0,020 °CDCL laboratory/ Customers PremisesActive
  • DMS 2003 (comparison method using calibrated PRTs, Thermocouples, Data Loggers , liquid fluidized baths, reference ovens and fixed ice point
  • -40 °C up to 250 °C
0,030 °CDCL laboratory/ Customers PremisesActive
  • 250 °C up to 400 °C
0,070 °CDCL laboratory/ Customers PremisesActive
  • 400 °C up to 550 °C
0,10 °CDCL laboratory/ Customers PremisesActive
  • 550 °C up to 650 °C
0,15 °CDCL laboratory/ Customers PremisesActive
  • liquid in glass thermometers
  • DMS 2004 (comparison method) Using calibrated PRTs, Thermocouples, Data loggers, liquid fluidized baths, and Fixed Ice point.
  • -40 °C up to 200 °C
0,10 °CDCL laboratoryActive
  • 200 °C up to 300 °C
0,30 °CDCL laboratoryActive
  • Nobel metal thermocouples and direct reading thermometers with nobel metal Thermocouple sensors
  • DMS 2007 (comparison method using calibrated PRTs, Thermocouples, Data Loggers , liquid fluidized bathes, reference ovens and fixed ice point)
  • 0 °C up to 600 °C
0,50 °CDCL laboratoryActive
  • 600 °C up to 900 °C
0,70 °CDCL laboratoryActive
  • 900 °C up to 1100 °C
1,5 °CDCL laboratoryActive
  • Enviromental Chambers and Ovens
  • DMS 2008 (comparison method using single or multiple calibrated probes)
  • -40 °C up to 200 °C
0,30 °CCustomers PremisesActive
  • 50 °C up to 250 °C
0,30 °CCustomers PremisesActive
  • Base metal thermocouples and direct reading thermometers with TC sensors
  • DMS 2007(comparison method using calibrated PRTs, Thermocouples, Data Loggers , liquid fluidized baths, reference ovens and fixed ice point)
  • -40 °C up to 250 °C
0,20 °CDCL laboratory and Customers Premises (direct reading thermometers only)Active
  • 250 °C up to 400 °C
0,50 °CDCL laboratory and Customers Premises (direct reading thermometers only)Active
  • 400 °C up to 550 °C
0,90 °CDCL laboratory and Customers Premises (direct reading thermometers only)Active
  • 550 °C up to 650 °C
1,5 °CDCL laboratory and Customers Premises (direct reading thermometers only)Active
  • 650 °C up to 900 °C
2,5 °CDCL laboratory and Customers Premises (direct reading thermometers only)Active
  • 900 °C up to 1100 °C
4,0 °CDCL laboratory and Customers Premises (direct reading thermometers only)Active
  • Infrared thermometers
  • DM-DCLD-SOP-MEL-IM T07 (Comparison method using Blackbody Simulator)
  • -20 °C up to -15 °C
0,80 °CDCL laboratoryActive
  • -15 °C up to 40 °C
0,30 °CDCL laboratoryActive
  • >40 °C up to 80 °C
0,80 °CDCL laboratoryActive
  • >80 °C up to 120 °C
1,0 °CDCL laboratoryActive
  • Refregerators , Freezers
  • DMS 2009 (comparison method using single and multiple calibrated probes)
  • -40 °C up to 20 °C
0,20 °CCustomers PremisesActive
  • Incubators
  • 20 °C up to 70 °C
0,20 °CCustomers PremisesActive
  • Liquid bathes
  • 5 °C up to 90 °C
0,40 °CCustomers PremisesActive
  • 90 °C up to 250 °C
0,60 °CCustomers PremisesActive
  • Furnaces
  • DMS 2010 (comparison method using single and multiple calibrated probes)
  • 300 °C up to 450 °C
0,80 °CCustomers PremisesActive
  • 450 °C up to 800 °C
2,0 °CCustomers PremisesActive
  • 800 °C up to 1100 °C
5,0 °CCustomers PremisesActive
  • Climatic chambers
  • DM-DCLD-SOP-MEL-IM T09 (comparison method using wireless temperture & humidity loggers)
  • 10 °C up to 70 °C (for temperature) 10% up to 95% (for relative Humidity)
0,20 °C 1,6% rHCustomers PremisesActive
  • Autoclaves
  • DM-DCLD-SOP-MEL-IM T10 (comparison method using wireless temperture & pressure loggers with timer)
  • 100 °C up to 135 °C (1 - 30 Mins) (0 - 150 kPa)
0,20 °C (35,4 ms) (1,8 kPa)Customers PremisesActive
  • Hygrometers
  • DM-DCLD-SOP-MEL-IM T08 (comparison using humidity generator)
  • 0 °C up to 70 °C (for temperature) 10% up to 96% (for relative Humidity)
0,10 °C 0,80% RHDCL laboratoryActive
  • Dry Block Calibrator / Dry Well Bath
  • DM-DCLD-SOP-MEL-IM T11 Comparison method using calibrated PRTs, SPRTs and Thermocouples
  • -45 °C up to 150 °C
0,05 °CDCL laboratoryActive
  • 150 °C up to < 650 °C
0,1 °CDCL laboratoryActive
  • 650 °C up to 1000 °C
0,5 °CDCL laboratoryActive
  • >1000 °C up to 1200 °C
1,0 °CDCL laboratoryActive
Pressure
  • Negative Gauge Pressure Gas Analogue and digital pressure gauges
  • Comparison method using standard transducer modules BS EN 837-1, EURAMET /cg-17 and DKD-R 6-1
  • -95 kPa ≤ p ≤ 0 kPa
0.050 kPaDCL LaboratoryActive
  • Gauge Pressure Gas Analogue and digital pressure gauges
  • 0 kPa ≤ p ≤ 7000 kPa
1.0 kPaDCL LaboratoryActive
  • Gauge Pressure Hydraulic Analogue and digital pressure gauges
  • Comparison method using dead weight tester, standard transducer BS EN 837-1 EURAMET /cg-17 and DKD-R 6-1
  • 100 kPa ≤ p ≤ 600 kPa
0.50 kPaDCL LaboratoryActive
  • 600 kPa < p ≤ 6 Mpa
1.2 kPaDCL LaboratoryActive
  • 6 MPa < p ≤ 70 Mpa
5 kPaDCL LaboratoryActive
  • Comparison method using Digital indicator / Calibrator BS EN 837-1, EURAMET /cg-17 and DKD-R 6-1
  • 0 MPa ≤ p ≤ 70 Mpa
1.2 kPa +1.83·10-3·pDCL Laboratory/ Customer PremisesActive
  • Negative Gauge Pressure Gas Analogue and digital pressure gauges
  • -100 kPa ≤ p ≤ 0 kPa
1000 PaCustomers PremisesActive
  • Barometer
  • DM-DCLD-SOP-MEL-IM P06 Comparison method using Automated Pressure Calibrator
  • 90 kPa to 110 kpa
0.025 kPaDCL LaboratoryActive
Electrical
  • DC Voltage Calibration of meters
  • DM-DCLD-SOP-MEL-IM E01 Comparison method using multi-product calibrator (Fluke 5520A MPC) U = measured voltage value
  • 0 mV to 330 mV
11•10-6•U + 2,9 µVDCL LaboratoryActive
  • >330 mV to 3,3 V
6,5•10-6 •U + 9,6 µVDCL LaboratoryActive
  • >3,3 V to 33 V
7,7•10-6 •U + 85 µVDCL LaboratoryActive
  • >33 V to 330 V
12•10-6•U + 0,74 mVDCL LaboratoryActive
  • >330 V to 1020 V
13•10-6•U + 2,1 mVDCL LaboratoryActive
  • DC Current Calibration of meters
  • DM-DCLD-SOP-MEL-IM E01 Comparison method using multi-product calibrator (Fluke 5520A MPC) I = measured current value
  • 0 to 329,999 μA
0,18•10-3 •I + 16 nADCL LaboratoryActive
  • >330 μA to 3,29999 mA
76•10-6 •I + 46 nADCL LaboratoryActive
  • >3,30 mA to 32,9999 mA
75•10-6 •I + 0,30 µADCL LaboratoryActive
  • >33 mA to 329,999 mA
68•10-6 •I + 5,7 µADCL LaboratoryActive
  • >330 mA to 1,09999 A
0,15•10-3 •I + 35 µADCL LaboratoryActive
  • >1.1 A to 2,99999 A
0,28•10-3 •I + 95 µADCL LaboratoryActive
  • >3 A to 10,9999 A
0,38•10-3 •I + 0,55 mADCL LaboratoryActive
  • >11 A to 20,5 A
0,77•10-3 •I + 0,88 mADCL LaboratoryActive
  • Resistance Calibration of meters
  • DM-DCLD-SOP-MEL-IM E01 Comparison method using multi-product calibrator (Fluke 5520A MPC) R = measured resistance value
  • 0 Ω to 10,9999 Ω
31•10-6 •R + 0,78 mΩDCL LaboratoryActive
  • >11 Ω to 32,999 Ω
23•10-6 •R + 1,2 mΩDCL LaboratoryActive
  • >33 Ω to 109,9999 Ω
21•10-6 •R + 1,2 mΩDCL LaboratoryActive
  • >110 Ω to 329,9999 Ω
20•10-6 •R + 2,6 mΩDCL LaboratoryActive
  • >330 Ω to 1,099999 kΩ
21•10-6 •R + 2,3 mΩDCL LaboratoryActive
  • >1,1 kΩ to 3,299999 kΩ
20•10-6 •R + 27 mΩDCL LaboratoryActive
  • >3,3 kΩ to 10,99999 kΩ
21•10-6 •R + 22 mΩDCL LaboratoryActive
  • >11 kΩ to 32,99999 kΩ
19•10-6 •R + 0,28 ΩDCL LaboratoryActive
  • >33 kΩ to 109,9999 kΩ
21•10-6 •R + 0,19 ΩDCL LaboratoryActive
  • >110 kΩ to 329,9999 kΩ
23•10-6 •R + 2,6 ΩDCL LaboratoryActive
  • >330 kΩ to 1,099999 MΩ
24•10-6 •R + 2,9 ΩDCL LaboratoryActive
  • >1,1 MΩ to 3,299999 MΩ
42•10-6 •R + 47 ΩDCL LaboratoryActive
  • >3,3 MΩ to 10,99999 MΩ
90•10-6 •R + 0,21 kΩDCL LaboratoryActive
  • >11 MΩ to 32,99999 MΩ
0,19 •10-3 •R + 2,3 kΩDCL LaboratoryActive
  • >33 MΩ to 109,9999 MΩ
0,38•10-3 •R + 3,9 kΩDCL LaboratoryActive
  • >110 MΩ to 329,9999 MΩ
2,3•10-3 •R + 0,082 MΩDCL LaboratoryActive
  • >330 MΩ to 1100 MΩ
11 •10-3•R + 0,60 MΩDCL LaboratoryActive
  • 1,0 mV to 32,999 mV 10 Hz to 45 Hz
0,62•10-3 •U + 4,9 µVDCL LaboratoryActive
  • 1,0 mV to 32,999 mV >45 Hz to 10 kHz
0,11•10-3 •U + 5,5 µVDCL LaboratoryActive
  • 1,0 mV to 32,999 mV >10 kHz to 20 kHz
0,14•10-3 •U + 5,5 µVDCL LaboratoryActive
  • 1,0 mV to 32,999 mV >20 kHz to 50 kHz
0,71 •10-3 •U + 7,5 µVDCL LaboratoryActive
  • 1,0 mV to 32,999 mV >50 kHz to 100 kHz
2,7 •10-3 •U + 12 µVDCL LaboratoryActive
  • 1,0 mV to 32,999 mV >100 kHz to 500 kHz
6,1 •10-3 •U + 44 µVDCL LaboratoryActive
  • >33 mV to 329,999 mV 10 Hz to 45 Hz
0,23•10-3 •U + 9,2 µVDCL LaboratoryActive
  • >33 mV to 329,999 mV >45 Hz to 10 kHz
0,10 •10-3 •U + 10 µVDCL LaboratoryActive
  • >33 mV to 329,999 mV >10 kHz to 20 kHz
0,12•10-3 •U + 10 µVDCL LaboratoryActive
  • AC Voltage Calibration of meters
  • DM-DCLD-SOP-MEL-IM E01 Comparison method using multi-product calibrator (Fluke 5520A MPC) U = measured voltage value
  • >33 mV to 329,999 mV >20 kHz to 50 kHz
0,25•10-3 •U + 13 µVDCL LaboratoryActive
  • >33 mV to 329,999 mV >50 kHz to 100 kHz
0,60•10-3 •U + 32 µVDCL LaboratoryActive
  • >33 mV to 329,999 mV >100 kHz to 500 kHz
1,5 •10-3 •U + 81 µVDCL LaboratoryActive
  • >0,33 V to 3,29999 V 10 Hz to 45 Hz
0,22•10-3•U + 76 µVDCL LaboratoryActive
  • >0,33 V to 3,29999 V >45 Hz to 10 kHz
0,11•10-3•U + 90 µVDCL LaboratoryActive
  • >0,33 V to 3,29999 V >10 kHz to 20 kHz
0,14•10-3•U + 86 µVDCL LaboratoryActive
  • >0,33 V to 3,29999 V >20 kHz to 50 kHz
0,22•10-3•U + 87 µVDCL LaboratoryActive
  • >0,33 V to 3,29999 V >50 kHz to 100 kHz
0,53•10-3•U + 0,16 mVDCL LaboratoryActive
  • >0,33 V to 3,29999 V >100 kHz to 500 kHz
1,8•10-3•U + 0,84 mVDCL LaboratoryActive
  • >3,3 V to 32,9999 V 10 Hz to 45 Hz
0,22•10-3•U + 0,83 mVDCL LaboratoryActive
  • >3,3 V to 32,9999 V >45 Hz to 10 kHz
0,11•10-3•U + 0,89 mVDCL LaboratoryActive
  • >3,3 V to 32,9999 V >10 kHz to 20 kHz
0,18•10-3•U + 0,83 mVDCL LaboratoryActive
  • >3,3 V to 32,9999 V >20 kHz to 50 kHz
0,26•10-3•U + 1,0 mVDCL LaboratoryActive
  • >3,3 V to 32,9999 V >50 kHz to 100 kHz
0,68•10-3•U + 1,9 mVDCL LaboratoryActive
  • >33 V to 329,999 V 10 Hz to 45 Hz
0,13•10-3•U + 9,4 mVDCL LaboratoryActive
  • >33 V to 329,999 V >45 Hz to 10 kHz
0,14•10-3•U + 11 mVDCL LaboratoryActive
  • >33 V to 329,999 V >10 kHz to 20 kHz
0,18•10-3•U + 10 mVDCL LaboratoryActive
  • >33 V to 329,999 V >20 kHz to 50 kHz
0,18•10-3•U + 26 mVDCL LaboratoryActive
  • >33 V to 329,999 V >50 kHz to 100 kHz
1,5•10-3•U + 44 mVDCL LaboratoryActive
  • >330 V to 1020 V 45 Hz to 1 kHz
0,23•10-3 •U + 15 mVDCL LaboratoryActive
  • >330 V to 1020 V >1 kHz to 5 kHz
0,19•10-3 •U + 16 mVDCL LaboratoryActive
  • >330 V to 1020 V >5 kHz to 10 kHz
0,23 •10-3 •U + 15 mVDCL LaboratoryActive
  • AC Current Calibration of meters
  • DM-DCLD-SOP-MEL-IM E01 Comparison method using multi-product calibrator (Fluke 5520A MPC) I = measured current value
  • 29 µA to 329,99 µA 10 Hz to 20 Hz
1,5•10-3 •I + 0,11 µADCL LaboratoryActive
  • 29 µA to 329,99 µA >20 Hz to 45 Hz
1,1•10-3 •I + 0,10 µADCL LaboratoryActive
  • 29 µA to 329,99 µA >45 Hz to 1 kHz
0,95•10-3 •I + 0,10 µADCL LaboratoryActive
  • 29 µA to 329,99 µA >1 kHz to 5 kHz
2,3•10-3 •I + 0,14 µADCL LaboratoryActive
  • 29 µA to 329,99 µA >5 kHz to 10 kHz
6,1•10-3 •I + 0,19 µADCL LaboratoryActive
  • 29 µA to 329,99 µA >10 kHz to 30 kHz
12•10-3 •I + 0,39 µADCL LaboratoryActive
  • >0,33 mA to 3,2999 mA 10 Hz to 20 Hz
1,4•10-3 •I + 0,59 µADCL LaboratoryActive
  • >0,33 mA to 3,2999 mA >20 Hz to 45 Hz
0,90•10-3 •I + 0,39 µADCL LaboratoryActive
  • >0,33 mA to 3,2999 mA >45 Hz to 1 kHz
0,73•10-3 •I + 0,30 µADCL LaboratoryActive
  • >0,33 mA to 3,2999 mA >1 kHz to 5 kHz
1,5•10-3 •I + 0,33 µADCL LaboratoryActive
  • >0,33 mA to 3,2999 mA >5 kHz to 10 kHz
3,4•10-3 •I + 2,3 µADCL LaboratoryActive
  • >0,33 mA to 3,2999 mA >10 kHz to 30 kHz
7,4•10-3 •I + 2,0 µADCL LaboratoryActive
  • >3,3 mA to 32,999 mA 10 Hz to 20 Hz
1,4•10-3 •I + 3,2 µADCL LaboratoryActive
  • >3,3 mA to 32,999 mA 20 Hz to 45 Hz
0,65•10-3 •I + 3,6 µADCL LaboratoryActive
  • >3,3 mA to 32,999 mA >45 Hz to 1 kHz
0,25•10-3 •I + 4,4 µADCL LaboratoryActive
  • >3,3 mA to 32,999 mA >1 kHz to 5 kHz
0,56•10-3 •I + 4,1 µADCL LaboratoryActive
  • >3,3 mA to 32,999 mA >5 kHz to 10 kHz
1,4•10-3 •I + 6,7 µADCL LaboratoryActive
  • >3,3 mA to 32,999 mA >10 kHz to 30 kHz
2,6•10-3 •I + 20 µADCL LaboratoryActive
  • >33 mA to 329,99 mA 10 Hz to 20 Hz
1,4•10-3 •I + 31 µADCL LaboratoryActive
  • >33 mA to 329,99 mA >20 Hz to 45 Hz
0,65•10-3 •I + 36 µADCL LaboratoryActive
  • >33 mA to 329,99 mA >45 Hz to 1 kHz
0,26•10-3 •I + 38 µADCL LaboratoryActive
  • >33 mA to 329,99 mA >1 kHz to 5 kHz
0,74 •10-3 •I + 56 µADCL LaboratoryActive
  • >33 mA to 329,99 mA >5 kHz to 10 kHz
1,5•10-3 •I + 0,11 mADCL LaboratoryActive
  • >33 mA to 329,99 mA >10 kHz to 30 kHz
2,8•10-3 •I + 0,29 mADCL LaboratoryActive
  • >0,33 A to 1,09999 A 10 Hz to 45 Hz
1,4•10-3 •I + 0,10 mADCL LaboratoryActive
  • >0,33 A to 1,09999 A >45 Hz to 1 kHz
0,36•10-3 •I + 0,12 mADCL LaboratoryActive
  • >0,33 A to 1,09999 A >1 kHz to 5 kHz
4,7•10-3 •I + 0,78 mADCL LaboratoryActive
  • >0,33 A to 1,09999 A >5 kHz to 10 kHz
19•10-3 •I + 3,9 mADCL LaboratoryActive
  • >1,1 A to 2,99999 A 10 Hz to 45 Hz
1,3•10-3 •I + 0,38 mADCL LaboratoryActive
  • >1,1 A to 2,99999 A >45 Hz to 1 kHz
0,35•10-3 •I + 0,65 mADCL LaboratoryActive
  • >1,1 A to 2,99999 A >1 kHz to 5 kHz
4,6•10-3 •I + 0,91 mADCL LaboratoryActive
  • >1,1 A to 2,99999 A >5 kHz to 10 kHz
19•10-3 •I + 4,0 mADCL LaboratoryActive
  • >3 A to 10,9999 A 45 Hz to 100 Hz
0,33•10-3 •I + 2,8 mADCL LaboratoryActive
  • >3 A to 10,9999 A >100 Hz to 1 kHz
0,56•10-3 •I + 2,6 mADCL LaboratoryActive
  • >3 A to 10,9999 A >1 kHz to 5 kHz
19•10-3 •I + 1,7 mADCL LaboratoryActive
  • >11 A to 20,5 A 45 Hz to 100 Hz
0,90•10-3 •I + 4,9 mADCL LaboratoryActive
  • >11 A to 20,5 A 100 Hz to 1 kHz
1,1•10-3 •I + 4,8 mADCL LaboratoryActive
  • >11 A to 20,5 A 1 kHz to 5 kHz
23•10-3 •I + 4,0 mADCL LaboratoryActive
  • Capacitance
  • DM-DCLD-SOP-MEL-IM E01 Comparison method using multi-product calibrator (Fluke 5520A MPC) C = measured capacitance value
  • 0,19 nF to 0,3999 nF
2,3•10-3 •C + 14 pFDCL LaboratoryActive
  • >0,4 nF to 1.0999 nF
2,6•10-3 •C + 14 pFDCL LaboratoryActive
  • >1,1 nF to 3,2999 nF
3,1•10-3 •C + 14 pFDCL LaboratoryActive
  • >3,3 nF to 10,9999 nF
1,5•10-3 •C + 17 pFDCL LaboratoryActive
  • >11 nF to 32,9999 nF
1,4•10-3 •C + 0,14 nFDCL LaboratoryActive
  • >33 nF to 109,999 nF
1,5•10-3 •C + 0,17 nFDCL LaboratoryActive
  • >110 nF to 329,999 nF
1,6•10-3 •C + 0,47 nFDCL LaboratoryActive
  • >0,33 µF to 1,09999 µF
1,5•10-3 •C + 1,7 nFDCL LaboratoryActive
  • >1,1 µF to 3,29999 µF
1,6•10-3 •C + 4,7 nFDCL LaboratoryActive
  • <3,3 µF to 10,9999 µF
1,5•10-3 •C + 17 nFDCL LaboratoryActive
  • >11 µF to 32,9999 µF
2,7•10-3 •C + 44 nFDCL LaboratoryActive
  • >33 µF to 109,999 µF
3,1•10-3 •C + 0,15 µFDCL LaboratoryActive
  • >110 µF to 329,999 µF
3,1•10-3 •C + 0,43 µFDCL LaboratoryActive
  • >0,33 µF to 1,09999 mF
3,2•10-3 •C + 1,6 µFDCL LaboratoryActive
  • >1,1 mF to 3,2999 mF
3,2•10-3 •C + 3,6 µFDCL LaboratoryActive
  • >3,3 mF to 10,9999 mF
3,2•10-3 •C + 12 µFDCL LaboratoryActive
  • >11 mF to 32,9999 mF
5,6•10-3 •C + 33 µFDCL LaboratoryActive
  • >33 mF to 110 mF
8,4•10-3 •C + 0,10 mFDCL LaboratoryActive
  • DC Clamp Meter
  • DM-DCLD-SOP-MEL-IM E01 Comparison method using multi-product calibrator Fluke 5520A with 50 – turns current coil I = measured current value
  • 20 A to 1000 A
4,3•10-3 •I + 0,065 mADCL LaboratoryActive
  • AC Clamp Meter
  • 20 A to 1000 A 45 Hz to 65 Hz
4,1•10-3 •I + 1,6 mADCL LaboratoryActive
  • 20 A to 1000 A >65 Hz to 200 Hz
9,5•10-3 •I + 0,8 mADCL LaboratoryActive
  • 20 A to 400 A >65 to 400 Hz
9,5•10-3 •I + 0,077 mADCL LaboratoryActive
  • DC power Calibration of meters
  • DM-DCLD-SOP-MEL-IM E03 Comparison method using multi-product calibrator (Fluke 5520A MPC) P = measured power value
  • 0 W to 10 kW
0,50•10-3 • PDCL LaboratoryActive
  • >10 kW to 20 kW
0,80•10-3 • PDCL LaboratoryActive
  • 0 W to 10 kW 45 Hz to 1 kHz
0,7•10-3 • PDCL LaboratoryActive
  • DM-DCLD-SOP-MEL-IM E03 Comparison method using multi-product calibrator (Fluke 5520A MPC) P = measured power value
  • >10 kW to 20 kW 45 Hz to 1 kHz
1,3•10-3 • PDCL LaboratoryActive
  • DC Voltage Calibration of sources
  • DM-DCLD-SOP-MEL-IM E04 Fluke 8508A Reference Multimeter U = measured voltage value
  • 0,0 mV to 200 mV
1,6 •10-6 •U+ 2,6 µVDCL LaboratoryActive
  • >200 mV to 2 V
2,9•10-6 •U+ 2,3 µVDCL LaboratoryActive
  • DM-DCLD-SOP-MEL-IM E04 4Fluke 8508A Reference Multimeter U = measured voltage value
  • >2 V to 20 V
3,5•10-6 •U + 4,2 µVDCL LaboratoryActive
  • DM-DCLD-SOP-MEL-IM E04 Fluke 8508A Reference Multimeter U = measured voltage value
  • >20 V to 200 V
5,5•10-6 •U + 38 µVDCL LaboratoryActive
  • >200V to 1000 V
5,5•10-6 •U + 0,46 mVDCL LaboratoryActive
  • DC Current Calibration of sources
  • DM-DCLD-SOP-MEL-IM E04 Fluke 8508A Reference Multimeter I = measured current value
  • 0,0 µA to 200 µA
13•10-6 •I + 0,30 nADCL LaboratoryActive
  • >200 µA to 2 mA
13•10-6 •I + 3,0 nADCL LaboratoryActive
  • >2 mA to 20 mA
14•10-6 •I + 31 nADCL LaboratoryActive
  • >20 mA to 200 mA
48•10-6 •I + 0,61 µADCL LaboratoryActive
  • >200 mA to 2 A
0,18•10-3 •I + 12 µADCL LaboratoryActive
  • >2 A to 20 A
0,40•10-3 •I + 0,30 mADCL LaboratoryActive
  • Resistance Calibration of sources
  • DM-DCLD-SOP-MEL-IM E04 Fluke 8508A Reference Multimeter R = measured resistance value
  • 0,0 Ω to 2 Ω
19•10-6 •R + 4,0 µΩDCL LaboratoryActive
  • >2 Ω to 20 Ω
10•10-6 •R + 14 µΩDCL LaboratoryActive
  • >20 Ω to 200 Ω
8,1•10-6 •R + 45 µΩDCL LaboratoryActive
  • >200 Ω to 2 kΩ
7,9•10-6 •R + 0,46 mΩDCL LaboratoryActive
  • >2 kΩ to 20 kΩ
7,9•10-6 •R + 4,6 mΩDCL LaboratoryActive
  • >20 kΩ to 200 kΩ
8,1•10-6 •R + 45 mΩDCL LaboratoryActive
  • >200 kΩ to 2 MΩ
10•10-6 •R + 0,87 ΩDCL LaboratoryActive
  • >2 MΩ to 20 MΩ
22•10-6 •R + 89 ΩDCL LaboratoryActive
  • >20 MΩ to 200 MΩ
0,12•10-3 •R + 9,2 kΩDCL LaboratoryActive
  • >200 MΩ to 2 GΩ
1,4•10-3 •R + 0,93 MΩDCL LaboratoryActive
  • AC voltage Calibration of sources
  • DM-DCLD-SOP-MEL-IM E04 Fluke 8508A Reference Multimeter U = measured voltage value
  • 0,0 mV to 200 mV 20 Hz to 40 Hz
0,13•10-3 •U + 4,4 µVDCL LaboratoryActive
  • 0,0 mV to 200 mV >40 Hz to 100 Hz
0,11•10-3 •U + 4,4 µVDCL LaboratoryActive
  • 0,0 mV to 200 mV >100 Hz to 2 kHz
0,11•10-3 •U + 2,7 µVDCL LaboratoryActive
  • 0,0 mV to 200 mV >2 kHz to 10 kHz
0,13•10-3 •U + 4,4 µVDCL LaboratoryActive
  • >200 mV to 2 V 20 Hz to 40 Hz
0,11•10-3 •U + 18 µVDCL LaboratoryActive
  • >200 mV to 2 V >40 Hz to 100 Hz
87•10-6 •U +19 µVDCL LaboratoryActive
  • >200 mV to 2 V >100 Hz to 2 kHz
73•10-6 •U + 18 µVDCL LaboratoryActive
  • >200 mV to 2 V >2 kHz to 10 kHz
0,11•10-3 •U + 19 µVDCL LaboratoryActive
  • DM-DCLD-SOP-MEL-IM E04 Fluke 8508A Reference Multimeter U = measured voltage value
  • >2 V to 20 V 20 Hz to 40 Hz
0,11•10-3 •U + 0,18 mVDCL LaboratoryActive
  • DM-DCLD-SOP-MEL-IM E04 Fluke 8508A Reference Multimeter U = measured voltage value
  • >2 V to 20 V >40 Hz to 100 Hz
89•10-6 •U + 0,18 mVDCL LaboratoryActive
  • >2 V to 20 V >100 Hz to 2 kHz
74•10-6 •U + 0,18 mVDCL LaboratoryActive
  • >2 V to 20 V >2 kHz to 10 kHz
0,11•10-3 •U + 0,18 mVDCL LaboratoryActive
  • >20 V to 200 V 20 Hz to 40 Hz
0,11•10-3 •U + 1,8 mVDCL LaboratoryActive
  • >20 V to 200 V >40 Hz to 100 Hz
90•10-6 •U + 1,8 mVDCL LaboratoryActive
  • >20 V to 200 V >100 Hz to 2 kHz
78•10-6 •U + 1,8 mVDCL LaboratoryActive
  • >20 V to 200 V >2 kHz to 10 kHz
0,11•10-3 •U + 1,8 mVDCL LaboratoryActive
  • >200 V to 1050 V 55 Hz to 10 kHz
0,11•10-3 •U + 19 mVDCL LaboratoryActive
  • AC Current calibration of sources
  • DM-DCLD-SOP-MEL-IM E04 Fluke 8508A Reference Multimeter I = measured current value
  • 0,0 µA to 200 µA 55 Hz to 1 kHz
1,0•10-3 •I + 82 nADCL LaboratoryActive
  • 0,0 µA to 200 µA >1 kHz to 5 kHz
2,4•10-3 •I + 0,12 µADCL LaboratoryActive
  • 0,0 µA to 200 µA >5 kHz to 10 kHz
6,4•10-3 •I + 0,16 µADCL LaboratoryActive
  • >200 µA to 2 mA 55 Hz to 1 kHz
0,84•10-3 •I + 0,20 µADCL LaboratoryActive
  • >200 µA to 2 mA >1 kHz to 5 kHz
1,6•10-3 •I + 0,22 µADCL LaboratoryActive
  • >200 µA to 2 mA >5 kHz to 10 kHz
4,0•10-3 •I + 0,27 µADCL LaboratoryActive
  • >2 mA to 20 mA 55 Hz to 1 kHz
0,43•10-3 •I + 2,5 µADCL LaboratoryActive
  • >2 mA to 20 mA >1 kHz to 5 kHz
0,70•10-3 •I + 2,4 µADCL LaboratoryActive
  • >2 mA to 20 mA >5 kHz to 10 kHz
1,6•10-3 •I + 2,9 µADCL LaboratoryActive
  • >20 mA to 200 mA 55 Hz to 1 kHz
0,43•10-3 •I + 25 µADCL LaboratoryActive
  • >20 mA to 200 mA >1 kHz to 5 kHz
0,85•10-3 •I + 44 µADCL LaboratoryActive
  • >20 mA to 200 mA >5 kHz to 10 kHz
1,6•10-3 •I + 82 µADCL LaboratoryActive
  • >200 mA to 2 A 55 Hz to 1 kHz
0,74•10-3 •I + 0,20 mADCL LaboratoryActive
  • >200 mA to 2 A >1 kHz to 5 kHz
4,8•10-3 •I + 0,82 mADCL LaboratoryActive
  • >200 mA to 2 A >5 kHz to 10 kHz
20•10-3 •I + 4,0 mADCL LaboratoryActive
  • >2 A to 20 A 55 Hz to 1 kHz
1,4•10-3 •I + 4,4 mADCL LaboratoryActive
  • >2 A to 20 A 1 kHz to 5 kHz
24•10-3 •I + 4,2 mADCL LaboratoryActive
  • Temperature Indicators & Simulators by Electrical Simulation and Measurements
  • DM-DCLD-SOP-MEL- IM E05 Comparison method using multi-product calibrator (Fluke 5520A MPC)
  • RTD -200 °C up to 800 °C
0.2 °CDCL LaboratoryActive
  • Thermocouple Type K -200 °C up to 1372 °C
0.4 °CDCL LaboratoryActive
  • Thermocouple Type J -210 °C up to 1200 °C
0.3 °CDCL LaboratoryActive
  • Thermocouple Type R 0 °C up to 1767 °C
0.5 °CDCL LaboratoryActive
  • Thermocouple Type S 0 °C up to 1767 °C
0.5 °CDCL LaboratoryActive
  • Thermocouple Type T -250 °C up to 400 °C
0.5 °CDCL LaboratoryActive
  • Thermocouple Type N -200 °C up to 1300 °C
0.4 °CDCL LaboratoryActive
  • Thermocouple Type B 600 °C up to 1820 °C
0.5 °CDCL LaboratoryActive
  • Thermocouple Type E -250 °C up to 1000 °C
0.5 °CDCL LaboratoryActive
  • Thermocouple Type C 0 °C up to 2316 °C
0.7 °CDCL LaboratoryActive
Dimensions
  • Gauge blocks: millimetre (Steel, Tungsten Carbide and Ceramic)
  • DMS 2015 Measurement of central length by mechanical comparison. ISO 3650
  • 0.5 mm to 10 mm
0.08 µmDCL LaboratoryActive
  • >10 mm to 25 mm
0.10 µmDCL LaboratoryActive
  • >25 mm to 50 mm
0.12 µmDCL LaboratoryActive
  • >50 mm to 75 mm
0.15 µmDCL LaboratoryActive
  • >75 mm to 100 mm
0.18 µmDCL LaboratoryActive
  • Micrometer (External)
  • DMS 2014 For determining error of indicated size Comparison with reference gauge blocks
  • 0 mm to 50 mm
2.5 μmDCL LaboratoryActive
  • 50 mm to 100 mm
3 μmDCL LaboratoryActive
  • Micrometer (Internal) (Calliper Type)
  • DMS 2014 For determining error of indicated size Comparison with reference gauge blocks using gauge block accessories
  • 5 mm to 50 mm
4 μmDCL LaboratoryActive
  • Vernier Callipers (Including Electronic)
  • DMS 2012 For determining error of indicated size Comparison with gauge blocks or calliper checker
  • 0 mm to 300 mm
20 μmDCL LaboratoryActive
  • 300 mm to 600 mm
25 μmDCL LaboratoryActive
  • Dial Gauge Calibrators / dial-indicator testers
  • DMS 2001 For determining error of indicated displacement Mechanical comparison to calibrated gauge blocks using precise digital height gauge
  • 0 mm to 50 mm
2.2 μmDCL LaboratoryActive
  • Dial Gauges
  • DMS 2013 For determining error of indicated displacement, Comparison with dial gauge calibrator
  • 0 mm to 100 mm
4 μmDCL LaboratoryActive
  • Precise Height Gauge (Height measurement system)
  • DM-DCLD-SOP-MEL-IM D10 For determining error of indicated vertical size Comparison with calibrated gauge block
  • 0 mm to 600 mm
6 μmDCL LaboratoryActive
  • Surface Plate (Table) (Granite/steel)
  • DM-DCLD-SOP-MEL-IM D13 Determining flatness using Inclination Measuring System
  • Up to 2500 x 1600 mm
9 µmCustomer PremisesActive
  • Calliper Checkers
  • DMS 2002 For determining face spacing Comparison to calibrated gauge blocks using precise digital height gauge
  • Up to 650 mm
8 μmDCL LaboratoryActive
  • LVDT Transducers
  • DMS 2011 For determining error of indicated displacement Mechanical comparison to calibrated gauge blocks
  • 0 mm to 100 mm
4 + (0.05 x L) μm, L: mmCustomer PremisesActive
  • Feeler Gauges
  • DMS 2024 For determining thickness Comparison method using calibrated thickness gauge
  • Up to 1 mm
3.5 μmDCL LaboratoryActive
  • I-checker (Precise dial gauge calibrator / dial-test indicator)
  • DM-DCLD-SOP-MEL-IM D16 For determining error of indicated displacement Mechanical comparison to calibrated gauge blocks using precise digital height gauge
  • 0 mm to 100 mm
2.2 μmDCL LaboratoryActive
  • Depth Micrometer
  • DM-DCLD-SOP-MEL-IM D17 For determining error of indicated depth Comparison with calibrated gauge blocks
  • 0 mm to 25 mm
2.5 μmDCL LaboratoryActive
  • Digimatic Thickness Gauge
  • DM-DCLD-SOP-MEL-IM D18 For determining error of indicated size Comparison with calibrated gauge blocks
  • 0 mm to 12.7 mm
2.5 μmDCL LaboratoryActive
  • Gauge Block Comparator
  • DMS 2023 For determining error of indicated displacement Mechanical comparison to gauge block couples EURAMET cg-2 (Previously EA-10/02)
  • 0 mm to 100 mm
0.040 μmDCL Laboratory / Customer PremisesActive
  • Metal Ruler
  • DM-DCLD-SOP-MEL-IM D20 Measurement of line spacing using length measuring machine equipped with optical magnification unit.
  • 0 mm to 1000 mm
0.18 mmDCL LaboratoryActive
  • Measuring Tape
  • DM-DCLD-SOP-MEL-IM D21 Measurement of line spacing using length measuring machine equipped with optical magnification unit. Stitching is carried out after length of 1000 mm.
  • 0 mm to 5000 mm
0.26 mmDCL LaboratoryActive
  • DM-DCLD-SOP-MEL-IM D21 Measurement of line spacing using length measuring machine equipped with optical magnification unit. Stitching is carried out after length of 1 m
  • 0 m to 50 m
0.25 + (0.07 x L) mm, L: mDCL LaboratoryActive
  • Tape and Scale Calibrator
  • DM-DCLD-SOP-MEL-IM D27 For determining error of indicated displacement, Comparison with reference measuring tape
  • 0 mm to 1000 mm
0.050 mmDCL Laboratory/ Customer PremisesActive
Mass and Balance
  • Balance
  • DM-DCLD-SOP-MEL-IM-M02 In Accordance to Euramet cg 18 & ASTM
  • Up to 3000 kg
2 x 10 -6 of the balance rangeCustomer PremisesActive
  • Mass Conventional Mass
  • DMS 2005 & DMS 2006 Substitution weighing with air buoyancy corr.
  • 50 kg
25 mgDCL laboratoryActive
  • 20 kg
10 mgDCL laboratoryActive
  • 10 kg
5.0 mgDCL laboratoryActive
  • 5 kg
2.5 mgDCL laboratoryActive
  • 2 kg
1.0 mgDCL laboratoryActive
  • 1 kg
0.5 mgDCL laboratoryActive
  • 500 g
0.25 mgDCL laboratoryActive
  • 200 g
0.10 mgDCL laboratoryActive
  • 100 g
0.05 mgDCL laboratoryActive
  • 50 g
0.03 mgDCL laboratoryActive
  • 20 g
0.025 mgDCL laboratoryActive
  • 10 g
0.020 mgDCL laboratoryActive
  • 5 g
0.016 mgDCL laboratoryActive
  • 2 g
0.012 mgDCL laboratoryActive
  • 1 g
0.010 mgDCL laboratoryActive
  • 500 mg
0.008 mgDCL laboratoryActive
  • 200 mg
0.006 mgDCL laboratoryActive
  • 100 mg
0.005 mgDCL laboratoryActive
  • 50 mg
0.004 mgDCL laboratoryActive
  • 20 mg
0.003 mgDCL laboratoryActive
  • 10 mg
0.003 mgDCL laboratoryActive
  • 5 mg
0.003 mgDCL laboratoryActive
  • 2 mg
0.003 mgDCL laboratoryActive
  • 1 mg
0.003 mgDCL laboratoryActive
  • DM-DCLD-SOP-MEL-IM-M01 Mass Scale and Substitution Weighing with air buoyancy corr.
  • E1: 1 mg to 20 kg
E1 and E2, F1, F2, M1DCL LaboratoryActive
  • E2: 1 mg to 50 kg
E1 and E2, F1, F2, M1DCL LaboratoryActive
Volume
  • Pipette and Micropipette (piston operated / digital)
  • DM-DCLD-SOP-MEL-IM V01 Comparison method using gravimetric method for single channel pipet
  • 0.2 to 2 µl
0.03 µlDCL LaboratoryActive
  • 0.5 to 10 µl
0.04 µlDCL LaboratoryActive
  • 2 to 20 µl
0.05 µlDCL LaboratoryActive
  • 10 to 100 μl
0.06 µlDCL LaboratoryActive
  • 20 to 200 μl
0.07 µlDCL LaboratoryActive
  • 10 to 300 μl
0.08 µlDCL LaboratoryActive
  • 50 to 500 μl
0.10 µlDCL LaboratoryActive
  • 0.1 to 1 ml
0.20 µlDCL LaboratoryActive
  • 1.0 to 5 mL
0.3 µlDCL LaboratoryActive
  • 1.0 to 10 mL
0.4 µlDCL LaboratoryActive
  • DM-DCLD-SOP-MEL-IM V01 Comparison method using gravimetric method for multi-channel pipette
  • 5 to 50 µl
0.05 µlDCL LaboratoryActive
  • 10 to 100 µl
0.06 µlDCL LaboratoryActive
  • 20 to 200 µl
0.07 µlDCL LaboratoryActive
  • 30 to 300 μl
0.15 µlDCL LaboratoryActive
  • 50 to 500 μl
0.20 µlDCL LaboratoryActive
  • 0.1 to 1 ml
0.50 µlDCL LaboratoryActive
  • Burette
  • DM-DCLD-SOP-MEL-IM V02 Comparison method using gravimetric method based on ISO 4787
  • 10 ml
0.012 mlDCL LaboratoryActive
  • 25 ml
0.030 mlDCL LaboratoryActive
  • 50 ml
0.058 mlDCL LaboratoryActive
  • 100 ml
0.116 mlDCL LaboratoryActive
  • Beaker
  • 10 ml
0.012 mlDCL LaboratoryActive
  • 20 ml
0.030 mlDCL LaboratoryActive
  • 50 ml
0.058 mlDCL LaboratoryActive
  • 100 ml
0.116 mlDCL LaboratoryActive
  • 200 ml
0.116 mlDCL LaboratoryActive
  • 500 ml
0.120 mlDCL LaboratoryActive
  • 1000 ml
0.126 mlDCL LaboratoryActive
  • 2000 ml
0.156 mlDCL LaboratoryActive
  • Measuring Cylinders
  • 10 ml
0.058 mlDCL LaboratoryActive
  • 20 ml
0.058 mlDCL LaboratoryActive
  • 50 ml
0.058 mlDCL LaboratoryActive
  • 100 ml
0.058 mlDCL LaboratoryActive
  • 200 ml
0.058 mlDCL LaboratoryActive
  • 500 ml
0.068 mlDCL LaboratoryActive
  • 1000 ml
0.078 mlDCL LaboratoryActive
  • 2000 ml
0.120 mlDCL LaboratoryActive
  • One-Mark Flask
  • 10 ml
0.004 mlDCL LaboratoryActive
  • 20 ml
0.004 mlDCL LaboratoryActive
  • 50 ml
0.006 mlDCL LaboratoryActive
  • 100 ml
0.01 mlDCL LaboratoryActive
  • 200 ml
0.018 mlDCL LaboratoryActive
  • 500 ml
0.044 mlDCL LaboratoryActive
  • 1000 ml
0.086 mlDCL LaboratoryActive
  • 2000 ml
0.17 mlDCL LaboratoryActive
Force
  • Universal Testing Machines (Tension)
  • DMS 2019 Comparison method using force proving instruments based on BS EN ISO 7500-1
  • 100 N up to 1200 kN (Tension)
0.0012Customer PremisesActive
  • Universal Testing Machines (Compression)
  • 100 N up to 3000 kN (Compression)
0.0012Customer PremisesActive
  • Force Measuring Systems- Tension, Compression
  • DMS 2019 Comparison method using calibrated deadweights based on BS EN ISO 7500-1
  • 0.01 N up to 235 N
0.0008Customer PremisesActive
  • Calibration of Force Testing Machines (Tension, Compression)
  • DMS 2019 Comparison method using Load Cells based on BS EN ISO 7500-1, BS EN 12390-4
  • 100 N up to 3000 kN (Compression) 100 N up to 1200 kN (Tension)
0.0012Customer PremisesActive
  • Force Verification/ Calibration of Testing Machines
  • DM-DCLD-SOP-MEL-IM F01 Comparison method using calibrated deadweights /load cell based on ASTM E4
  • 0.01 N up to 3000 kN (Compression)
0.002Customer PremisesActive
  • 0.01 N up to 1200 kN (Tension)
0.002Customer PremisesActive
  • Compression Testing Machines for Concrete - Performance Test (Stability Verification)
  • DMS 2022 Comparison method using Strain Gauge Load Cells based on BS EN 12390-4
  • Alignment of Upper Machine Platen Alignment of Machine Components, Restraint on movement of Upper Platten
N/ACustomer PremisesActive
  • Force measuring devices using soil testing apparatus
  • DMS 2020 Comparison method using Load Cells based on BS 1377 part 1
  • 25 N up to 50 kN
0.0022DCL LaboratoryActive
  • 25 N up to 200 kN
0.007Customer PremisesActive
  • Extensometers
  • DMS 2021 Comparison method using calibrated extensometer calibration rigs / calibrators based on BS EN ISO 9513
  • L ³ 0.02 mm to 58.5 mm
(1+0.3*L) µm L distance in mmCustomer PremisesActive
  • DM-DCLD-SOP-MEL- IM D25 Comparison method using calibrated extensometer calibration rigs/ calibrators based on ASTM E83
  • L ³ 0.02 mm up to ≤ 5 mm
1 μmCustomer PremisesActive
  • L > 5 mm up to ≤ 58.5 mm
2.5 μmCustomer PremisesActive
Time & Speed
  • TIME (Stopwatch)
  • DM-DCLD-SOP-MEL-IM S02 Comparison method using calibrate High Resolution Counter / Timer / Analyzer
  • 1 min to 30 min
1.0 ms / minDCL LaboratoryActive
  • 30 min to 2 hours
9.7 ms / minDCL LaboratoryActive
  • Rotational Speed ( Centrifuge , Motors , etc )
  • DM-DCLD-SOP-MEL-IM S01 Comparison method using calibrated Digital Tachometer
  • 60 - 10000 rpm
0.9 rpmCustomer PremisesActive
  • 10000 - 50000 rpm
2.0 rpmCustomer PremisesActive
  • Timer
  • DM-DCLD-SOP-MEL-IM S02 Comparison method using calibrated Stopwatch
  • 1 hour
0.86 secCustomer PremisesActive
  • Digital Tachometer ( Non - contact mode )
  • DM-DCLD-SOP-MEL-IM S03 Comparison method using calibrated waveform generator
  • 60 - 10000 rpm
0.1 rpmDCL LaboratoryActive
  • 10000 - 99000 rpm
0.6 rpmDCL LaboratoryActive